| Literature DB >> 33968444 |
Ángel López-Vázquez1, Inés Contreras2, Sergio Martin-Prieto3, Ángel López-Castro3.
Abstract
PURPOSE: To evaluate the weight of intraocular lenses (IOLs) depending on their material, dioptric power, toricity, focality, and haptic design.Entities:
Year: 2021 PMID: 33968444 PMCID: PMC8081614 DOI: 10.1155/2021/6686700
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Figure 1The balance used to weigh the intraocular lenses.
Figure 2An intraocular lens, set on absorbent paper to remove surplus fluid prior to being weighed.
Weights and characteristics of all IOLs included.
| Company | Model | Se (D) | Hydrated weight (Mg) | Hydrophobic | Focality | Toric | Design |
|---|---|---|---|---|---|---|---|
| Zeiss | CT LUCIA 211P | 9.50 | 15 | Yes | Mono | No | CL |
| Zeiss | CT LUCIA 611P | 28 | 19 | Yes | Mono | No | CL |
| AJL | AIALA DRY | 21 | 22 | Yes | Mono | No | CL |
| Alcon | AU00T0 | 27 | 14 | Yes | Mono | No | CL |
| Alcon | CNA0T0 | 18 | 15 | Yes | Mono | No | CL |
| Alcon | SN60AT | 7 | 17 | Yes | Mono | No | CL |
| Alcon | SN60WF | 18 | 16 | Yes | Mono | No | CL |
| Alcon | SN60AT | 18 | 18 | Yes | Mono | No | CL |
| Aaren scientific | AARIS | 22.50 | 20 | Yes | Mono | No | CL |
| J&J | PCB00 | 19 | 17 | Yes | Mono | No | CL |
| Topcon | LS-312 | 18.5 | 24 | No | Mono | No | CL |
| Alcon | MN60AC | 22.5 | 16 | Yes | Mono | No | 3P |
| Alcon | PANOPTIX | 20 | 16 | Yes | Multi | No | CL |
| J&J | ZA9003 | 11.5 | 20 | Yes | Mono | No | 3P |
| Zeiss | CT SPHERIS 204 | 21 | 32 | No | Mono | No | P |
| Zeiss | CT ASPHINA 409 MP | 22 | 31 | No | Mono | No | P |
| Zeiss | CT ASPHINA 404 | 20 | 30 | No | Mono | No | P |
| Topcon | LS313 | 20 | 30 | No | Mono | No | P |
| Topcon | LS313MF30 | 24.5 | 31 | No | Mono | Yes | P |
| Topcon | LS313 T3 | 10.5 | 29 | No | Mono | Yes | P |
| Topcon | LS313T1 | 27.25 | 34 | No | Mono | Yes | P |
| Zeiss | AT LISA 839 MP | 12.5 | 26 | No | Multi | No | P |
| Zeiss | AT LISA 939 MP | 21.5 | 33 | No | Multi | No | P |
| Topcon | LU313MFT | 13.5 | 33 | No | Multi | Yes | P |
| Medical mix | MICRO +123 | 17 | 25 | No | Mono | No | 4H |
| Medical mix | MICRO +A123 | 25.5 | 26 | No | Mono | No | 4H |
| Medical mix | MICROPURE 123 | 27.5 | 18 | No | Mono | No | 4H |
| Medical mix | MICROPURE 123 | 12.5 | 16 | No | Mono | No | 4H |
| Medical mix | FINEVISION PO | 15 | 23 | No | Multi | No | 4H |
| Medical mix | ANKORIS TORIC | 20.25 | 24 | No | Mono | Yes | 4H |
| Medical mix | ANKORIS TORIC | 24.75 | 25 | No | Mono | Yes | 4H |
| Medicontur | BI-FLEX | 19.5 | 28 | No | Mono | No | CL |
| Medicontur | BI-FLEX | 18.5 | 26 | No | Mono | No | CL |
| Medicontur | BI-FLEX | 19.5 | 28 | No | Mono | No | CL |
| B&L | MX60T | 20.62 | 16 | 1Yes | Mono | Yes | CL |
| B&L | MX60T | 20.62 | 16 | 1Yes | Mono | Yes | CL |
| B&L | MX60T | 21 | 16 | 1Yes | Mono | Yes | CL |
| B&L | MX60T | 22 | 17 | 1Yes | Mono | No | CL |
| Freedom ring | CTR 13-11 | 0 | <1 |
Mono, monofocal; Multi, multifocal; CL, C-loop; 4H, four haptics; P, plate; 3P, 3-piece.
Average values according to the different characteristics evaluated.
| Group | n | Average weight (mg) | Standard deviation |
|---|---|---|---|
| Hydrophobic | 17 | 17.06 | 2.11 |
| Hydrophilic | 21 | 27.24 | 4.73 |
| Monofocal | 33 | 22.15 | 6.16 |
| Multifocal | 5 | 26.20 | 7.19 |
| Toric | 8 | 26.00 | 7.09 |
| Non-toric | 30 | 21.80 | 5.96 |
| <15 D | 7 | 22.29 | 7.06 |
| ≥15 D | 31 | 22.77 | 6.30 |
| ≥24 D | 7 | 23.86 | 7.24 |
| <24 D | 31 | 22.42 | 6.23 |
| C-loop | 19 | 18.95 | 4.48 |
| 4-haptic | 7 | 22.43 | 3.87 |
| Plate | 10 | 30.90 | 2.33 |
| 3-piece | 2 | 18 | 2.83 |