| Literature DB >> 33959456 |
Elizabeth L Covington1, Richard A Popple1.
Abstract
A procedure is presented to assess performance at non-zero couch angles and perform routine quality assurance (QA) on surface-guided radiotherapy (SGRT) imaging systems used for stereotactic radiosurgery (SRS). A low-cost anthropomorphic phantom was used to assess the system under patient-like conditions. The phantom is embedded with a tungsten ball bearing (BB) to facilitate the use of surface imaging (SI) with concurrent megavoltage (MV) imaging to cross-compare and validate SI-reported offsets. Data analysis is done via in-house software that utilized the SGRT system's log files for automated analysis. This procedure enables users to assess and inter-compare MV-reported offsets with their SGRT system. The analysis provides SGRT system residual error so that users are aware of inherent offsets present in addition to increases in translational offsets due to couch walkout. The procedure was validated with two commercial SGRT systems. The procedure can be used with any surface imaging system and linear accelerator system.Entities:
Keywords: radiation therapy; sgrt; stereotactic radiosurgery; surface imaging; surface-guided radiotherapy
Year: 2021 PMID: 33959456 PMCID: PMC8093097 DOI: 10.7759/cureus.14278
Source DB: PubMed Journal: Cureus ISSN: 2168-8184
Figure 1(Left) Polystyrene phantom with embedded tungsten carbide ball bearing for radiographic imaging. (Right) Representation of BB depths used for testing. All BBs were places midline with different phantoms used for each depth. A representative ROI for fSRS is shown in pink.
BB: Ball bearing; ROI: Region of interest; fSRS: Frameless stereotactic radiosurgery.
Figure 2The SGRT system residual error compared to that of EPID images in the lateral (x) and longitudinal (y) direction for System A. The BB placement was at midplane depth (10 cm).
SGRT: Surface guided radiotherapy; EPID: Electronic portal imaging device; BB: Ball bearing.
Figure 3The magnitude (mm) of the SGRT system residual error as a function of couch angle (IEC61217) for each system tested on an Edge linear accelerator with the gantry at zero. BB placement was at midplane depth (10 cm) for each system.
SGRT: Surface guided radiotherapy; BB: Ball bearing.
Example worksheet for calculating SGRT residual error. BB location is determined via MV imaging while phantom offsets are recorded from the surface imaging system.
SGRT: Surface guided radiotherapy; BB: Ball bearing; MV: Megavoltage.
| Couch Angle (°) | BB (Lat., Long.) (mm) | Phantom (Lat., Long.) (mm) | SGRT Residual Error (Lat., Long) (mm) | SGRT Residual Error Magnitude (mm) |
| 0 | 0.00, 0.00 | -0.05, -0.04 | 0.05, -0.04 | 0.06 |
| 90 | 0.10, 0.20 | 0.20, 0.40 | -0.10, -0.20 | 0.22 |
| 270 | 0.30, 0.20 | 0.10, 0.60 | 0.20, -0.40 | 0.45 |