Assessment of Bridging Stent Grafts in Branched Endovascular Aortic Repair (EVAR) Procedures Using Intravascular Ultrasound (IVUS).

Marco V Usai1, Alexander Oberhuber1, Giuseppe Asciutto1.   

Abstract

Entities:  

Keywords:  BEVAR; EVAR; IVUS; Stentgraft

Year:  2020        PMID: 33937892      PMCID: PMC8077033          DOI: 10.1016/j.ejvsvf.2020.04.003

Source DB:  PubMed          Journal:  EJVES Vasc Forum        ISSN: 2666-688X


× No keyword cloud information.

Introduction

Complex endovascular aortic repair (EVAR) procedures such as branched EVAR (BEVAR) require significant amounts of contrast media or radiation exposure. In addition, the use of intra-operative cone beam computed tomography (CT) with its additional radiation exposure is increasing because of its reliability in investigating the anatomy of the implanted stent grafts. In particular, cone beam CT has been shown to deliver high intra-operative endoleak detection rates and technical issues requiring direct intervention. Intravascular ultrasound (IVUS) has become a useful diagnostic tool in endovascular procedures. In the case of EVAR, it has been shown to be helpful in assessing the proximal and distal stent graft landing zones, reducing the need for contrast media in patients with compromised renal function. In particular, the IVUS technology can characterise the presence of thrombus and calcifications in a reliable way, also providing real time vessel diameter measurements. Its usefulness in dissection cases to discriminate between the true and false lumen has already been described. Therefore, the present authors developed a diagnostic protocol integrating the use of IVUS technology during BEVAR procedures. The IVUS technology was chosen to prove its applicability in detecting geometric anomalies of the bridging grafts during BEVAR.

Surgical technique

After full deployment of the bridging stent graft through the brachial artery, and after angiographic assessment of its patency, a 0.018 High Torque Command Wire (Abbott Laboratories, Abbott Park, IL, USA) was advanced to the distal end of the bridging stent graft. The Glidecath catheter was then removed and a Vision PV 0.018 IVUS catheter (Volcano, Philips, Amsterdam, The Netherlands) was advanced through the bridging stent graft. After activation of the B mode modality, the IVUS catheter was pulled gently back through the bridging stent graft up to the entry of the branch (Photo 1). The IVUS technique gave a 360° image of the vessel, detecting the presence of mismatch or bridging stent graft kinking using both B mode and Cromaflo (Supplementary material).
Photo 1

Intravascular ultrasound catheter (§) during pull back from the right renal artery and bridging stent.

Intravascular ultrasound catheter (§) during pull back from the right renal artery and bridging stent. All branches were investigated sequentially with the IVUS catheter at the end of each bridging stent implantation. Since the introduction of the IVUS technology at the study clinic, 33 target vessels have been cannulated with the Volcano IVUS system. In two cases a haemodynamic non-relevant kinking of the bridging stent graft (<50% reduction of the diameter) at the level of the ostium of a superior mesenteric artery (Figure 1, Figure 2, Figure 3) and of a left renal artery (Figure 4, Figure 5, Figure 6) was found. No relevant stenosis, dissection, or kinking of the bridging stent grafts and/or at the target vessel were observed (Figure 2, Figure 3, Figure 4). Furthermore, no procedure related complications from using the IVUS technology were detected. This was confirmed in all cases, using the first post-operative CT scan 30 days after the BEVAR procedure.
Figure 1

Intravascular ultrasound image of the stent graft at the level of the ostium of the superior mesenteric artery showing an elliptical shape.

Figure 2

Intra-arterial angiography demonstrates minimal (<50%) reduction of the diameter at the level of the ostium of the superior mesenteric artery (blue arrow).

Figure 3

Plain radiographic demonstration of the minimal kinking of the superior mesenteric artery (yellow arrow) bridging stent.

Figure 4

Intravascular ultrasound image of the stent graft at the level of the ostium of the left renal artery showing an elliptical shape.

Figure 5

Intra-arterial angiography demonstrates minimal (<50%) diameter reduction at the level of the left renal artery (blue arrow).

Figure 6

Plain radiographic demonstration of the minimal kinking of the bridging stent graft of the left renal artery (blue arrow).

Intravascular ultrasound image of the stent graft at the level of the ostium of the superior mesenteric artery showing an elliptical shape. Intra-arterial angiography demonstrates minimal (<50%) reduction of the diameter at the level of the ostium of the superior mesenteric artery (blue arrow). Plain radiographic demonstration of the minimal kinking of the superior mesenteric artery (yellow arrow) bridging stent. Intravascular ultrasound image of the stent graft at the level of the ostium of the left renal artery showing an elliptical shape. Intra-arterial angiography demonstrates minimal (<50%) diameter reduction at the level of the left renal artery (blue arrow). Plain radiographic demonstration of the minimal kinking of the bridging stent graft of the left renal artery (blue arrow).

Discussion

The use of IVUS in conventional endovascular aortic procedures and its potential in reducing radiation exposure and contrast medium use have been described already. In peripheral arterial and venous interventions, the application of IVUS has been associated with lower post-procedural complication and amputation rates with a non-significant increase in hospitalisation costs. Furthermore, the post-procedural luminal area, measured by IVUS, has been shown to be predictive of short and long term patency, highlighting the clinical utility of adjunctive imaging when performing lower extremity endovascular interventions. IVUS has also been evaluated for access vessel assessment prior to transarterial valve repair, with comparable if not more reliable image quality when it comes to the choice of the access, as well as potential radiation and contrast volume reduction. At the beginning of this experience, the present authors aimed to apply the IVUS technology to all complex aortic reconstructions. However, difficulties were encountered when advancing the IVUS catheter through the bridging stent graft via the femoral route, which is the preferred route for fenestrated EVAR (FEVAR). Therefore, it was decided to restrict use of the IVUS technique to BEVAR cases only via brachial access. In this preliminary experience, the IVUS technology proved to be applicable in studying the anatomy of the bridging stent grafts during BEVAR procedures. The technical success rate was 100%. No relevant stenosis, dissection, or kinking of the bridging stent grafts were observed. The additional procedural time (3.5 mins per vessel in the present experience) and costs should be considered the main drawbacks of this technology. However, the information gained about the anatomy of the bridging stent graft with the IVUS technology may be useful during BEVAR procedures.

Conclusions

IVUS has become an essential tool in the present authors' practice for assessing bridging stent grafts geometry in BEVAR procedures. At this time, the IVUS technique should be considered a complementary diagnostic tool in complex EVAR. This technique could potentially reduce radiation exposure and procedure costs and thus replace the cone-beam CT.
  8 in total

1.  15-Year Patency and Life Expectancy After Primary Stenting Guided by Intravascular Ultrasound for Iliac Artery Lesions in Peripheral Arterial Disease.

Authors:  Hisao Kumakura; Hiroyoshi Kanai; Yoshihiro Araki; Yoshiaki Hojo; Toshiya Iwasaki; Shuichi Ichikawa
Journal:  JACC Cardiovasc Interv       Date:  2015-11-18       Impact factor: 11.195

2.  Intravascular Ultrasound in Lower Extremity Peripheral Vascular Interventions: Variation in Utilization and Impact on In-Hospital Outcomes From the Nationwide Inpatient Sample (2006-2011).

Authors:  Sidakpal S Panaich; Shilpkumar Arora; Nilay Patel; Nileshkumar J Patel; Chirag Savani; Achint Patel; Badal Thakkar; Vikas Singh; Samir Patel; Nish Patel; Kanishk Agnihotri; Parth Bhatt; Abhishek Deshmukh; Vishal Gupta; Ramak R Attaran; Carlos I Mena; Cindy L Grines; Michael Cleman; John K Forrest; Apurva O Badheka
Journal:  J Endovasc Ther       Date:  2015-12-04       Impact factor: 3.487

3.  Single-Center Experience and Preliminary Results of Intravascular Ultrasound in Endovascular Aneurysm Repair.

Authors:  Felice Pecoraro; Umberto M Bracale; Arduino Farina; Giovanni Badalamenti; Francesca Ferlito; Mario Lachat; Ettore Dinoto; Vincenzo Asti; Guido Bajardi
Journal:  Ann Vasc Surg       Date:  2018-11-27       Impact factor: 1.466

Review 4.  Using IVUS during EVAR and TEVAR: improving patient outcomes.

Authors:  Benjamin J Pearce; William D Jordan
Journal:  Semin Vasc Surg       Date:  2009-09       Impact factor: 1.000

5.  Vascular Assessment for Transcatheter Aortic Valve Replacement: Intravascular Ultrasound Compared With Computed Tomography.

Authors:  Essa Essa; Nader Makki; Peter Bittenbender; Quinn Capers; Barry George; Gregory Rushing; Juan Crestanello; Konstantinos Dean Boudoulas; Scott M Lilly
Journal:  J Invasive Cardiol       Date:  2016-12       Impact factor: 2.022

6.  A retrospective study of intravascular ultrasound use in patients undergoing endovascular aneurysm repair: its usefulness and a description of the procedure.

Authors:  K Hoshina; M Kato; T Miyahara; A Mikuriya; N Ohkubo; T Miyata
Journal:  Eur J Vasc Endovasc Surg       Date:  2010-08-23       Impact factor: 7.069

7.  Impact of onlay fusion and cone beam computed tomography on radiation exposure and technical assessment of fenestrated-branched endovascular aortic repair.

Authors:  Emanuel R Tenorio; Gustavo S Oderich; Giuliano A Sandri; Pinar Ozbek; Jussi M Kärkkäinen; Thanila A Macedo; Terri Vrtiska; Stephen Cha
Journal:  J Vasc Surg       Date:  2018-12-04       Impact factor: 4.268

8.  Intravascular ultrasound assisted sizing in thoracic endovascular aortic repair improves aortic remodeling in Type B aortic dissection.

Authors:  Julia Lortz; Konstantinos Tsagakis; Christos Rammos; Michael Horacek; Thomas Schlosser; Heinz Jakob; Tienush Rassaf; Rolf Alexander Jánosi
Journal:  PLoS One       Date:  2018-04-19       Impact factor: 3.240

  8 in total
  2 in total

Review 1.  [Innovations in the endovascular treatment of peripheral arterial disease].

Authors:  Jörg Teßarek; Alexander Oberhuber
Journal:  Gefasschirurgie       Date:  2021-08-13

Review 2.  Peri-Operative Management of Patients Undergoing Fenestrated-Branched Endovascular Repair for Juxtarenal, Pararenal and Thoracoabdominal Aortic Aneurysms: Preventing, Recognizing and Treating Complications to Improve Clinical Outcomes.

Authors:  Andrea Xodo; Mario D'Oria; Bernardo Mendes; Luca Bertoglio; Kevin Mani; Mauro Gargiulo; Jacob Budtz-Lilly; Michele Antonello; Gian Franco Veraldi; Fabio Pilon; Domenico Milite; Cristiano Calvagna; Filippo Griselli; Jacopo Taglialavoro; Silvia Bassini; Anders Wanhainen; David Lindstrom; Enrico Gallitto; Luca Mezzetto; Davide Mastrorilli; Sandro Lepidi; Randall DeMartino
Journal:  J Pers Med       Date:  2022-06-21
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.