Literature DB >> 33937128

A β-Amyloid(1-42) Biosensor Based on Molecularly Imprinted Poly-Pyrrole for Early Diagnosis of Alzheimer's Disease.

Rezvan Dehdari Vais1, Hossein Yadegari2, Hossein Heli3, Naghmeh Sattarahmady3,4.   

Abstract

BACKGROUND: Alzheimer's disease (AD) is a common form of dementia, characterized by production and deposition of β-amyloid peptide in the brain. Thus, β-amyloid peptide is a potentially promising biomarker used to diagnose and monitor the progression of AD.
OBJECTIVE: The study aims to develop a biosensor based on a molecularly imprinted poly-pyrrole for detection of β-amyloid.
MATERIAL AND METHODS: In this experimental study, an imprinted poly-pyrrole was employed as an artificial receptor synthesized by electro-polymerization of pyrrole on screen-printed carbon electrodes in the presence of β-amyloid. β-amyloid acts as a molecular template within the polymer. The biosensor was evaluated by cyclic voltammetry using ferro/ferricyanide marker. The parameters influencing the biosensor performance, including electro-polymerization cycle umbers and β-amyloid binding time were optimized to achieve the best biosensor sensitivity.
RESULTS: The β-amyloid binding affinity with the biosensor surface was evaluated by the Freundlich isotherm, and Freundlich constant and exponent were obtained as 0.22 ng mL-1 and 10.60, respectively. The biosensor demonstrated a detection limit of 1.2 pg mL-1. The biosensor was applied for β-amyloid determination in artificial cerebrospinal fluid.
CONCLUSION: The biosensor is applicable for early Alzheimer's disease detection. Copyright: © Journal of Biomedical Physics and Engineering.

Entities:  

Keywords:  Artificial Antibody; Beta-Amyloid; Bioprobe; Biosensing Technique; Molecular Imprinting; Neurodegenerative Disease; Sensory Aids; β-Amyloid (1-40)

Year:  2021        PMID: 33937128      PMCID: PMC8064131          DOI: 10.31661/jbpe.v0i0.1070

Source DB:  PubMed          Journal:  J Biomed Phys Eng        ISSN: 2251-7200


Introduction

Alzheimer’s disease (AD) is a neurodegenerative disorder characterized by memory loss and cognitive dysfunction; it is a common form of dementia. It is investigated that about 80 million people will be afflicted with AD in 2050 [ 1 ]. AD is caused due to the production and deposition in the brain of β-amyloid peptides that included 39 to 42 amino acid residues [ 2 ]. Although positron emission tomography imaging shows cortical amyloid pathology with cognitive criteria [ 3 , 4 ], early diagnosis of AD stage has still been of great significance. Disease biomarkers are detectable indicators of the disease states, which can be specific molecules, genes, antigens, proteins, cells, or hormones. Detection of biomarkers is the new route of the disease state and its diagnosis. Along this line, different categories of biosensors have been introduced for the biomarker detection, including aptasensors [ 5 ], genosensors [ 6 , 7 ], non-enzymatic biosensors [ 8 , 9 ], and immunosensors [ 10 ]. For fabrication of the biosensors, nanomaterials with special size and shape have been applied as transducers [ 11 - 13 ]. Among these nanobiosensors, electrochemical transduction is sensitive, simple, robust, low-cost, and miniaturizable, and works with small sized samples [ 13 ]. Nanomaterials, as components of the electrochemical nanobiosensors (transducer or else), make them more sensitive and selective due to the unique redox properties, chemical stability and tuneability of covering the substrates [ 5 , 6 , 8 , 9 , 11 - 14 ]. As an indicator of AD, β-amyloid peptides, and specially β-amyloid(1-42), are promising biomarkers used to diagnose and monitor the progression of AD [ 15 ]. Up to now, the techniques of ELISA [ 16 ], different modes of mass spectrometry (MS) [ 17 - 22 ], immunosensors [ 23 ], scanning tunneling electron microscopy [ 24 ], magnetic resonance imaging [ 25 ], fluorimetry [ 26 ], in vivo optical imaging [ 27 , 28 ], positron emission tomography [ 29 , 30 ] and surface plasmon resonance [ 31 ] have been reported for detection of β-amyloid for AD diagnosis. However, the ELISA method suffers from high limits of quantitation, high costs of enzyme-linked antibodies and consumption time. On the other hand, new modes of MS, including time-of-flight secondary ion MS [ 17 ], on-line immunoaffinity-liquid chromatography-MS [ 18 ], hydrogen exchange with top-down electron capture dissociation MS [ 19 ], matrix-assisted laser desorption ionization-time-of-flight MS [ 20 ], column-switching liquid chromatography-tandem MS [ 21 ], and covalent chiral derivatized ultraperformance liquid chromatography-tandem MS [ 22 ] have the disadvantages of relative insensitivity, complexicity, high cost, time-consuming and difficulty for β-amyloid determination. The immunosensors have disadvantages arising from the instability of antibodies. Therefore, there is a need to fabricate novel sensing devices for β-amyloid determination. Molecular imprinting technology is a tailor-made recognition capability for (template) molecules at the molecular level using molecularly imprinted polymers (MIPs). The template molecule can be memorized with a high affinity with specific binding sites in the cross-linked polymer matrix. Subsequent removal of the template reveals binding sites that are complementary in size and shape to the original imprinted template molecule [ 32 ]. Upon rebinding, MIPs recognize the template molecule with a high selectivity. In MIPs, a retention mechanism based on molecular recognition is dominated like immunosorbents. As a result, MIPs are often named as synthetic antibodies, while they offer better handling, stability (toward acids, bases, ions and organic solvents), lower costs, high pressure resistance, wide working temperature range and ease of preparation [ 32 ], compared to antibody-based routes making them attractive for numerous applications. MIPs have been successfully applied as an alternative tool over the biological entities in several medicinal fields such as drug delivery, cell recognition, and protein recognition [ 33 - 35 ]. Until now, MIPs-based biosensors have been developed for the detection of biomarkers such as troponin I [ 36 ], troponin T [ 37 ], prostate-specific antigen [ 38 ], myoglobin [ 39 ], microRNA [ 40 ] and neopterin [ 41 ]. However, β-amyloid detection using MIPs has not been approached. In the present study, a MIP-based biosensor for determination of β-amyloid was prepared to apply for early detection of AD.

Material and Methods

In this experimental study, all chemicals used throughout this study were of analytical grade from Scharlau (Spain) or Merck (Germany) and employed without further purification. Double distilled water was used to prepare all aqueous solutions. Lyophilized β-amyloid(1-42) peptide was purchased from Sigma (USA). Deionized water was used for solutions preparation. Electrochemical experiments were carried out using a µ-Autolab potentiostat/galvanostat (the Netherlands) interfaced to a PC with GPES 4.9 software. Screen printed carbon electrodes (SPCEs) were purchased from DropSens (Spain) and employed for all measurements. In the optimized preparation conditions of the biosensor, an electrochemical pretreatment of SPCEs was firstly done in 100 mmol L-1 Tris-HCl buffer, pH= 7.4 (Tris) at 1.2 V for 60 s, followed by rinsing the electrode surface with Tris. Afterward, 30 μL of the polymerization solution containing 100 mmol L-1 pyrrole monomer and 1.29 μg mL-1 β-amyloid(1-42) (as a template) prepared in Tris was placed on the pretreated SPCE surface. Electro-polymerization was then carried out by cyclic voltammetry (CV) in a potential range between -0.1 to 0.8 V with a potential sweep rate of 50 mV s-1 for five consecutive cycles. After electro-polymerization process, the removal of template molecule from the polymeric layer was performed by immersing the electrode into 5.0 mmol L-1 aqueous solution of oxalic acid at room temperature for 24 h. The resultant MIP film was then washed with Tris and applied as the working electrode for all electrochemical measurements described below. A NIPE was prepared under the same experimental conditions without the presence of β-amyloid(1-42) template in the electro-polymerization solution. The stepwise fabrication process of the biosensors is schematically presented in Figure 1.
Figure 1

The stepwise fabrication process of the biosensors.

The stepwise fabrication process of the biosensors. CV was employed to investigate the analytical performance of the biosensor at different concentrations of β-amyloid(1-42) in Tris and artificial cerebrospinal fluid (CSF). Initially, the biosensor was incubated in solutions of β-amyloid(1-42) at 37 °C for 20 min. After that, the biosensor was washed with Tris to remove non-bonded β-amyloid(1-42), and the CV measurements were carried out from -0.35 to 0.35 V with a potential sweep rate of 50 mV s-1 in 0.5 mmol L-1 K3[Fe(CN)6]/K4[Fe(CN)6] 1:1 mixture prepared in Tris. An artificial CSF was made with a composition of NaCl 126 + NaH2PO4 1.24 + NaHCO3 26 + KCl 2.5 + MgSO4 1.2 + CaCl2 2.0 + D-glucose 10 mmol L-1, pH=7.35, and with a stream of 95% air + 5% CO2 was saturated. β-amyloid(1-42) of different concentrations were prepared in the artificial CSF, diluted with Tris of the final dilution ratio of 1:1, and found by the biosensor.

Results

Electro-polymerization of pyrrole was performed in the synthesis solution containing β-amyloid(1-42) as a molecular template, as described in section 2.3. β-amyloid(1-42)binding into the active sites of the biosensor surface was electrochemically monitored through repression of the redox kinetic of the K3[Fe(CN)6]/K4[Fe(CN)6] marker. To optimize the working conditions of the biosensor, anodic peak currents in cyclic voltammograms of the biosensor were measured upon re-binding by β-amyloid(1-42) of 3.0 ng mL-1. Because the thickness of PPy film has a significant influence on the biosensing of β-amyloid(1-42), it was changed by alterations in the potential cycle number during the electro-polymerization. For this purpose, PPy was synthesized using different potential cycles, and evaluated before and after capturing β-amyloid(1-42) of 3.0 ng mL-1; the results are presented in Figure 2A. To judge about the best potential cycle number for the synthesis of PPy, the anodic peak currents were measured and their differences before and after capturing the target are plotted versus cycle number in Figure 2B. Selection of the anodic peak was due to the more sharpness of this peak, compared to the cathodic counterpart. Based on the results, the highest peak currents, and more importantly, the highest peak current difference before and after β-amyloid(1-42) capturing was attained for PPy synthesis with five potential cycles. These results indicated that up to five cycles, the MIP layer thickened leading to better formation of the β-amyloid(1-42)-resembled holes in the bulk of the polymer. However, more cycles of potential lead to formation of a very thick polymeric layer and inhibition of the marker accessibility to the underlying current collector (vie infra, Figure 3). Therefore, the best sensitivity was attained with five potential cycles of PPy synthesis, and also employed for further studies.
Figure 2

(A) Cyclic voltammograms recorded before and after binding of the biosensor with β-amyloid(1-42) for different potential cycles of 1, 3, 5, 10 and 15. (B) Dependency of the anodic peak current difference (difference of before β-amyloid(1-42) binding and after that) on the cycle numbers of the molecularly imprinted polymer (MIP) synthesis.

Figure 3

The fabrication processes and signal generation of the biosensor.

(A) Cyclic voltammograms recorded before and after binding of the biosensor with β-amyloid(1-42) for different potential cycles of 1, 3, 5, 10 and 15. (B) Dependency of the anodic peak current difference (difference of before β-amyloid(1-42) binding and after that) on the cycle numbers of the molecularly imprinted polymer (MIP) synthesis. The fabrication processes and signal generation of the biosensor. Another parameter that should be optimized for a MIP-base biosensor is the biosensor surface-target binding time. The biosensor was incubated with β-amyloid(1-42) of 3.0 ng mL-1 at room temperature for different binding times of 3, 5, 10, 20, 30 and 40 min; cyclic voltammograms were recorded, as shown in Figure 4. According to these results, the peak current decrement continued upon time prolonging up to 20 min and then remained almost constant. Therefore, a binding time of 20 min was selected for β-amyloid(1-42)) targeting.
Figure 4

(A) Cyclic voltammograms recorded using the biosensor before and after capturing β-amyloid(1-42) at different binding times of 3, 5, 10, 20, 30 and 40 min. (B) Dependency of the anodic peak current on the binding time of β-amyloid(1-42).

(A) Cyclic voltammograms recorded using the biosensor before and after capturing β-amyloid(1-42) at different binding times of 3, 5, 10, 20, 30 and 40 min. (B) Dependency of the anodic peak current on the binding time of β-amyloid(1-42). For the elution of the bond β-amyloid(1-42)from the biosensor surface after fabrication of a new biosensor or after already binding for reusing, alkaline and organic compounds-containing solvents were not suitable because the screen-printed electrodes were pulled up in these solutions. Our results also showed that mineral acids and acetic acid/surfactant mixtures could not elute β-amyloid(1-42) from the biosensor, while a 5.0 mmol L-1 oxalic acid solution at room temperature overnight successfully eluted the target, and this later eluent was employed thereafter. In order to quantitate the β-amyloid(1-42)by the biosensor, cyclic voltammograms were recorded after binding with different concentrations of β-amyloid(1-42), and are shown in Figure 5A. Upon increment in the β-amyloid(1-42) concentration, the peak current was lowered due to blockage of the pinholes at the MIP surface by the target and prevention of accessibility of the redox marker for electron exchange. At the higher concentrations, therefore, a surface saturation happened, and the rate of the current decrement was slow. Despite the fact that PPy is a conducting polymer, it has a limited conductivity in Tris-buffer of 100 mmol L-1, pH=7.4 (working conditions of the biosensor) and represses the kinetics of charge transfer of the redox marker. Thus, a limited current is passed, which is important and appropriate when NIPE is evaluated (see below). On the other side, the subsequent elution of the β-amyloid(1-42) template during the biosensor preparation formed lots of pores with a protein-resembled shape. This porous structure supplies a network for the marker to attain the underneath carbon substrate, and a signal for the marker is recordable. Capturing β-amyloid(1-42) by the biosensor blocks the holes and limits the redox marker accessibility to the carbon substrate for electron transfer. Therefore, the current depends on the β-amyloid(1-42) blocking level of the biosensor. The fabrication processes and signal generation of the biosensor are schematically depicted in Figure 3.
Figure 5

(A) Cyclic voltammograms recorded using the biosensor before and after capturing different concentrations of β-amyloid(1-42). (B) The dependency of the peak currents on the β-amyloid(1-42) concentration (calibration curve).

(A) Cyclic voltammograms recorded using the biosensor before and after capturing different concentrations of β-amyloid(1-42). (B) The dependency of the peak currents on the β-amyloid(1-42) concentration (calibration curve). On the basis of the data presented in Figure 5A, the dependency of the peak currents on the β-amyloid(1-42) concentration can be derived, as shown in Figure 5B. The peak current dependency was linear in a concentration range of 0.003-7 ng mL-1, and the trend line had a regression equation of y = -(1.39 ± 0.045)x + (26.96 ± 0.062). The use of the standard deviation of the anodic peak current in the absence of the target (the blank signal), and the slope of the calibration curve are presented in Figure 5B, the limit of detection (LOD) of β-amyloid(1-42) by the biosensor was calculated. In order to evaluate the importance of the surface imprinting on biosensor function, we fabricated a NIPE, and recorded the cyclic voltammograms in the presence of different concentrations of β-amyloid(1-42), as depicted in Figure 6. The data indicated that NIPE did not generate significant signals upon varying the β-amyloid(1-42) concentration.
Figure 6

Cyclic voltammogram recorded using the biosensor (voltammogram with the highest peak currents), and cyclic voltammograms recorded using Non-molecularly imprinted polymer electrode (NIPE) before and after capturing different concentrations of β-amyloid(1-42).

Cyclic voltammogram recorded using the biosensor (voltammogram with the highest peak currents), and cyclic voltammograms recorded using Non-molecularly imprinted polymer electrode (NIPE) before and after capturing different concentrations of β-amyloid(1-42). An isotherm for estimation of β-amyloid(1-42) binding affinity into the biosensor at different concentrations can be plotted, using the data presented in Figure 5A. The models of Langmuir [ 42 ], Freundlich [ 43 ] and two-type simultaneous binging [ 44 ] were examined. The best fitting for the peak current-β-amyloid(1-42) concentration was attained based on the Freundlich adsorption isotherm, using the following equation: (Ipa0 - Ip) / Ipa0 = K × C1/n (1) where (Ipa0 - Ip) / Ip0 indicates that the amount of β-amyloid(1-42) captured at a concentration of C, and Ipa0 is the anodic peak current in the absence of β-amyloid(1-42); Ip is the anodic peak current in the presence of β-amyloid(1-42). In addition, K is the Freundlich constant, and n is the Freundlich exponent. Figure 7 shows the corresponding isotherm based on the Freundlich adsorption isotherm for the biosensor. The results indicated that the amount of the captured β-amyloid(1-42) gradually decreased upon increment in the β-amyloid(1-42) concentration, and then reached an adsorption plateau at 5 ng mL-1. The adsorption plateau is the result of saturation of binding sites at the biosensor surface by β-amyloid(1-42). Using the regression equation in Figure 7, we obtained K and n values as 0.22 ng mL-1 and 10.60, respectively. The high value obtained for n confirmed that the amount of the bound β-amyloid(1-42) greatly increased with a small change in the concentration (at its low concentrations), while more imprinted sites were rebounded, and the adsorption intensity weakened and eventually remained constant (at its high concentrations); later, saturation occurred and the current ratio remained constant.
Figure 7

Dependency of (Ipa0 - Ip) / Ipa0 on the concentration of β-amyloid(1-42) and the fitting curve based on the Freundlich adsorption isotherm.

Dependency of (Ipa0 - Ip) / Ipa0 on the concentration of β-amyloid(1-42) and the fitting curve based on the Freundlich adsorption isotherm. For evaluation of the biosensor capability for β-amyloid(1-42) detection in real samples, β-amyloid(1-42) of different concentrations was supplemented into CSF and cyclic voltammograms were recorded, as shown in Figure 8A. Upon increment in the β-amyloid(1-42) concentration, the peak current was lowered similarly as observed for the β-amyloid(1-42)quantitation in Tris. A calibration curve was also plotted for the data, which shown in Figure 8B. Based on this calibration curve, β-amyloid(1-42) could be determined in CSF by a sensitivity of 1.09 µA mL ng-1 and a LOD of 2.3 pg mL-1.
Figure 8

A) Cyclic voltammograms recorded using the biosensor before and after capturing different concentrations of β-amyloid(1-42) in artificial cerebrospinal fluid (CSF). (B) The dependency of the peak currents on the β-amyloid(1-42) concentration (calibration curve).

A) Cyclic voltammograms recorded using the biosensor before and after capturing different concentrations of β-amyloid(1-42) in artificial cerebrospinal fluid (CSF). (B) The dependency of the peak currents on the β-amyloid(1-42) concentration (calibration curve).

Discussion

MIPs-based biosensors, as artificial receptors, provide an alternative method for selective recognition of biomolecules. The process of molecular imprinting includes the creation of a polymeric film with selective recognition cavities based on a biomolecular template [ 32 ]. In the present study, poly-pyrrole (PPy) polymer which synthesized in the presence of β-amyloid(1-42) template, was employed to fabricate the biosensor because it can be formed electrochemically and stably at neutral (physiological) pHs that are important for the biosensor applications in biological sample analyses [ 45 ]. For removing of template from the biosensor surface, it should be mentioned that different solvents might be employed. A literature review on the solvents employed for protein elution from MIPs is summarized in Table 1. The reported solvents include mineral acids and bases, and organic solutes or solvents dissolved in aqueous solutions, accompanied with or without employment of a surfactant. In this research, the oxalic acid solution can successfully eluted β-amyloid(1-42) template. The designed biosensor could successfully detect β-amyloid(1-42) with a LOD of as 1.2 pg mL-1 which is comparable with different methods of amyloid detection (Table 2). In order to evaluate the actual performance of biosensor, it was applied for β-amyloid(1-42) determination in artificial CSF with high sensitivity and selectivity, confirming the biosensor applicability for AD diagnosis.
Table 1

Different solvents reported for protein elution from molecularly imprinted polymers (MIPs).

ProteinEluentReference
AvidinChloroform[46]
Ferritin5 mmol L-1 NaOH[47]
Troponin TEthanol:water (2:1 V/V) + 0.25 mol L-1 NaOH[48]
MyoglobinProteinase K[49]
Cytochrome C1 mol L-1 H2SO4[50]
Immunoglobulin G3 mol L-1 NaCl + 0.1% sodium dodecyl sulfate[51]
Bovine hemoglobin1 mol L-1 oxalic acid[52]
Bovine hemoglobin10% Acetic acid + 10% sodium dodecyl sulfate[53]
AcetylcholinesteraseGlycine-HCl[54]
Bovine serum albumin5% Acetic acid + 10% sodium dodecyl sulfate[55]
Bovine serum albumin5% Acetic acid + 10% sodium dodecyl sulfate[56]
Human serum albumin30% NaOH [57]
Ovarian cancer antigen1% Acetic acid + 3% sodium dodecyl sulfate[58]
Bovine leukemia virus glycoproteins1 mol L-1 H2SO4[59]
Table 2

Comparison of different methods of amyloid detection.

Amyloid typeMethodDetecting elementLinear rangeLODReference
β-amyloid(1-42)ElectrochemistryAnti mAβ/protein G/SAM/AuNPs 0.01-100 nmol L-10.57 nmol L-1[60]
β-amyloid(1-40)ElectrochemistryAnti mAβ/pyrenyl groups1-200 µmol L-12.04 µmol L-1[60]
β-amyloid(1-40)ElectrochemistryAnti mAβ/SAM/AuNPs 1-1000 nmol L-12.65 nmol L-1[60]
β-amyloid(1-40)ElectrochemistryAntibody0.01-100 nmol L-110 pmol L-1[61]
β-amyloid(1-40/1-42)ElectrochemistryAntibody0.2-40 nmol L-150 pmol L-1[62]
β-amyloid(1-42/1-16)ElectrochemistryAntibody0-0.5 nmol L-15 pmol L-1[63]
β-amyloid(1-40)FluorescenceAptamer0-70 nmol L-13.57 nmol L-1[64]
β-Amyloid(1-42)ElectrochemistryBiotinylated PrPC (95-110) peptide10-6-1 µmol L-10.5 pmol L-1[65]
β-amyloid(1-42)FluorescenceAntibody-magnetic nanoparticle10-3000 pg mL-120 pg mL-1[66]
β-amyloid(1-40)FluorescenceAntibody-magnetic nanoparticle10-4000 pg mL-120 pg mL-1[66]
β-amyloid(1-40)SPRAntibody-3.3 pmol L-1[67]
β-amyloid(1-42)SPRAntibody-3.5 pmol L-1[67]
β-amyloid(1-42)ElectrochemistryAntibody/AuNPs0.001-10 ng mL-11 pg mL-1[68]
β-amyloid(1-42)ECLAntibody0.73-3000 pg mL-11.24 pg mL-1[69]
β-amyloid(1-40/1-42/1-16)ElectrochemistryHeme β-amyloid(1-16)-AuNPs/ mAb0.02-1.50 nmol L-110 pmol L-1[70]
β-amyloid(1-42)LSPRAntibody<10 pmol L-1<100 fmol L-1[71]
β-amyloid(1-42)LSPRSingle gold nanoparticle10-6-100 µmol L-11.5 pmol L-1[72]
β-amyloidElectrochemistryAntibody-Aptamer 0.5-30 nmol L-1100 pmol L-1[73]
β-amyloid(1-42)ElectrochemistryAntibody-AuNPs 10-1000 pg mL-15.2 pg mL-1[74]
β-amyloid(1-42)ElectrochemistryAntibody0.5-500 ng mL-10.1 ng mL-1[75]
β-amyloid(1-42)ElectrochemistryAntibody--[76]
β-amyloid(1-42)Bio-barcode detectionAntibody0.1-100 fmol L-1100 amol L-1[77]
β-amyloid(1-42)MicroarrayMatched antibody pair0-5 ng mL-173.07 pg mL-1[78]
β-Amyloid(1-42)ElectrochemistryOligopeptide0.48-12 nmol L-1240 pmol L-1[79]
β-amyloid(1-42)ElectrochemistryMIP0.003-7 ng mL-11.2 pg mL-1, 266 fmol L-1Present study

Anti mAβ: Monoclonal amyloid beta antibody, SAM: Self-assembled monolayer, AuNPs: Gold nanoparticles, PrPC: Cellular prion protein, ECL: Electrochemiluminescence, mAb: Monoclonal antibody, LSPR: Localized surface plasmon resonance, ApoE4: Apolipoprotein E4, MIP: Molecularly imprinted polymer, SPR: Surface plasmon resonance, LOD: Limit of detection

Different solvents reported for protein elution from molecularly imprinted polymers (MIPs). Comparison of different methods of amyloid detection. Anti mAβ: Monoclonal amyloid beta antibody, SAM: Self-assembled monolayer, AuNPs: Gold nanoparticles, PrPC: Cellular prion protein, ECL: Electrochemiluminescence, mAb: Monoclonal antibody, LSPR: Localized surface plasmon resonance, ApoE4: Apolipoprotein E4, MIP: Molecularly imprinted polymer, SPR: Surface plasmon resonance, LOD: Limit of detection

Conclusion

A synthetic β-amyloid(1-42) receptor was designed using molecular imprinting approach to fabricate a biosensor. The biosensor represented a high affinity and sensitivity. The receptor was fabricated by electro-polymerization of pyrrole on carbon screen-printed electrodes in the presence of β-amyloid(1-42) as a template. Formation of a thin MIP layer was critical to improve the signals as well as the performance of the biosensor towards the target β-amyloid(1-42). The resultant β-amyloid(1-42) receptor was examined by recording electrochemical signals of the ferro/ferricyanide redox probe after the β-amyloid(1-42) binding process. At the same time, successful elution of β-amyloid(1-42) was achieved from the MIP surface. A Freundlich isotherm fitting revealed a high affinity of the receptor toward the imprinted β-amyloid(1-42). β-amyloid(1-42) binding onto the biosensor imposed a significant decrement in the peak current of the redox marker and provided a calibration curve with a linear dependency of the peak currents on the β-amyloid(1-42) concentration. A very low LOD value was attained, compared to other reported techniques, confirming the applicability of the biosensor for β-amyloid(1-42) detection at a clinically relevant level.
  58 in total

1.  Binding characteristics of radiofluorinated 6-dialkylamino-2-naphthylethylidene derivatives as positron emission tomography imaging probes for beta-amyloid plaques in Alzheimer's disease.

Authors:  E D Agdeppa; V Kepe; J Liu; S Flores-Torres; N Satyamurthy; A Petric; G M Cole; G W Small; S C Huang; J R Barrio
Journal:  J Neurosci       Date:  2001-12-15       Impact factor: 6.167

Review 2.  Adsorption--from theory to practice.

Authors:  A Dabrowski
Journal:  Adv Colloid Interface Sci       Date:  2001-10-08       Impact factor: 12.984

3.  Electrochemical aptamer/antibody based sandwich immunosensor for the detection of EGFR, a cancer biomarker, using gold nanoparticles as a signaling probe.

Authors:  Hoda Ilkhani; Morteza Sarparast; Abolhassan Noori; S Zahra Bathaie; Mir F Mousavi
Journal:  Biosens Bioelectron       Date:  2015-07-03       Impact factor: 10.618

4.  Conformer-specific hydrogen exchange analysis of Aβ(1-42) oligomers by top-down electron capture dissociation mass spectrometry.

Authors:  Jingxi Pan; Jun Han; Christoph H Borchers; Lars Konermann
Journal:  Anal Chem       Date:  2011-06-14       Impact factor: 6.986

5.  A new methodology for simultaneous quantification of total-Aβ, Aβx-38, Aβx-40, and Aβx-42 by column-switching LC/MS/MS.

Authors:  Ken-ichi Watanabe; Chihiro Ishikawa; Hiroshi Kuwahara; Kimihiko Sato; Setsuko Komuro; Tetsuya Nakagawa; Naruaki Nomura; Shiro Watanabe; Masashi Yabuki
Journal:  Anal Bioanal Chem       Date:  2011-12-27       Impact factor: 4.142

Review 6.  Immunosensors in Early Cancer Diagnostics: From Individual to Multiple Biomarker Assays.

Authors:  Lucie Korecká; Karel Vytřas; Zuzana Bílková
Journal:  Curr Med Chem       Date:  2018       Impact factor: 4.530

7.  Competitive electrochemical immunoassay for detection of β-amyloid (1-42) and total β-amyloid peptides using p-aminophenol redox cycling.

Authors:  Lin Liu; Qige He; Feng Zhao; Ning Xia; Huijing Liu; Sujuan Li; Ruili Liu; Hui Zhang
Journal:  Biosens Bioelectron       Date:  2013-08-01       Impact factor: 10.618

8.  A method for evaluating the level of soluble β-amyloid(1-40/1-42) in Alzheimer's disease based on the binding of gelsolin to β-amyloid peptides.

Authors:  Yanyan Yu; Lin Zhang; Chenglin Li; Xiaoyu Sun; Daoquan Tang; Guoyue Shi
Journal:  Angew Chem Int Ed Engl       Date:  2014-09-22       Impact factor: 15.336

9.  An ultrasensitive human cardiac troponin T graphene screen-printed electrode based on electropolymerized-molecularly imprinted conducting polymer.

Authors:  Bárbara V M Silva; Blanca A G Rodríguez; Goreti F Sales; Maria Del Pilar T Sotomayor; Rosa F Dutra
Journal:  Biosens Bioelectron       Date:  2015-10-28       Impact factor: 10.618

10.  A First Tetraplex Assay for the Simultaneous Quantification of Total α-Synuclein, Tau, β-Amyloid42 and DJ-1 in Human Cerebrospinal Fluid.

Authors:  Niels Kruse; Michael G Schlossmacher; Walter J Schulz-Schaeffer; Eugeen Vanmechelen; Hugo Vanderstichele; Omar M El-Agnaf; Brit Mollenhauer
Journal:  PLoS One       Date:  2016-04-26       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.