Effect of macrogeometry and bone type on insertion torque, primary stability, surface topography damage and titanium release of dental implants during surgical insertion into artificial bone.

Gabrielle Alencar Ferreira Silva1, Fernanda Faot2, Anna Paula da Rosa Possebon3, Wander José da Silva4, Altair Antoninha Del Bel Cury5.   

Abstract

This study investigated the influence of implant macrogeometry and bone type on insertion torque (IT), primary stability (ISQ), surface topography damage, and the amount of titanium (Ti) released during insertion. Forty implants with different macrogeometries (Facility - Cylindrical with spiral-shaped threads; Alvim - Tapered with buttress-shaped threads) were inserted into artificial bone types I-II and III-IV. Surface morphology was evaluated by Scanning Electron Microscope (SEM) and roughness parameters with Laser Scanning Confocal Microscopy (LSCM) before and after insertion (AI). Implant macrogeometry was characterized by LSCM. The chemical composition of bone beds was determined by SEM associated with Energy Dispersive X-Ray Spectroscopy. The amount of Ti released was analyzed with Energy Dispersive X-Ray Fluorescence. Alvim had greater IT and ISQ than Facility. Bone types I-II require higher IT of implants. Alvim also had greater internal threads angle, higher initial roughness, and significant reduction of roughness AI, compared to Facility. The functional surface height reduced AI, especially in flank and valley of threads. Height of surface roughness of Alvim and Facility implants was similar AI. Implants surface morphology changes and metallic particles on bone beds were observed after implant insertion, mainly into bone types III-IV. Implants inserted into bone types I-II showed less surface damage. Alvim implants released more Ti (37.52 ± 25.03 ppm) than Facility (11.66 ± 28.55 ppm) on bone types III-IV. The implant macrogeometry and bone types affect IT, ISQ, surface damage, and Ti amount released during insertion. Alvim implants were more wear susceptible, releasing higher Ti concentration during insertion into bone types III-IV.
Copyright © 2021 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Bone density; Dental implants; Implant design; Surface topography; Titanium

Year:  2021        PMID: 33932754     DOI: 10.1016/j.jmbbm.2021.104515

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  4 in total

Review 1.  Relationship between dental implant macro-design and osseointegration: a systematic review.

Authors:  Simone Kreve; Izabela Ferreira; Mariana Lima da Costa Valente; Andréa Cândido Dos Reis
Journal:  Oral Maxillofac Surg       Date:  2022-09-29

2.  Effects of insertion torque values on the marginal bone loss of dental implants installed in sheep mandibles.

Authors:  Sergio Alexandre Gehrke; Jaime Aramburú Júnior; Tiago Luis Eirles Treichel; Tales Dias do Prado; Berenice Anina Dedavid; Piedad N de Aza
Journal:  Sci Rep       Date:  2022-01-11       Impact factor: 4.996

3.  Host inflammatory response and clinical parameters around implants in a rat model using systemic alendronate and zoledronate acid drug administrations.

Authors:  Kristian Kniha; Lothar Rink; Jana Wolf; Stephan Christian Möhlhenrich; Florian Peters; Marius Heitzer; Frank Hölzle; Ali Modabber
Journal:  Sci Rep       Date:  2022-03-15       Impact factor: 4.379

4.  The Role of Titanium Particles and Ions in the Pathogenesis of Peri-Implantitis.

Authors:  Farah Asa'ad; Peter Thomsen; Marcel F Kunrath
Journal:  J Bone Metab       Date:  2022-08-31
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.