RF electrode-tissue coverage significantly influences steam pop incidence and lesion size.

Felix Bourier1, Miruna Popa1, Marc Kottmaier1, Susanne Maurer1, Fabian Bahlke1, Marta Telishevska1, Sarah Lengauer1, Katharina Koch-Büttner1, Marielouise Kornmayer1, Elena Risse1, Amir Brkic1, Tilko Reents1, Gabriele Hessling1, Isabel Deisenhofer1.   

Abstract

BACKGROUND: Steam pops are a rare complication associated with radiofrequency (RF) ablation and are hard to predict. The aim of this study was to assess the influence of coverage between the RF ablation electrode and cardiac tissue on steam pop incidence and lesion size. METHODS AND
RESULTS: An ex vivo model using porcine cardiac preparations and contact force sensing catheters was designed to perform RF ablations at different coverage levels between the RF electrode and cardiac tissue. During coverage level I, only the distal part of the ablation electrode was in contact with tissue. During coverage level II half of the ablation electrode, and during coverage level III the entire ablation electrode was embedded in tissue. RF applications (n = 60) at different coverage levels I-III were systematically performed using the same standardized ablation protocol. Ablations during coverage level III resulted in a significantly higher rate of steam pops (100%) when compared to ablations during coverage level II (10%) and coverage level I (0%), log rank p < .001. Coverage level I ablations resulted in significantly smaller lesion depths, diameters, and impedance drops when compared to higher coverage level ablations, p < .001. In the controlled ex vivo model, there was no difference in applied contact force or energy between different coverage levels.
CONCLUSIONS: The level of coverage between RF electrode, cardiac tissue, and the surrounding fluid significantly influenced the incidence of steam pops in an ex vivo setup. Larger coverage between RF electrode and tissue resulted in significantly larger lesion dimensions.
© 2021 Wiley Periodicals LLC.

Keywords:  lesion size; radiofrequency ablation; steam pop

Year:  2021        PMID: 33928696     DOI: 10.1111/jce.15063

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  3 in total

1.  Perpendicular catheter orientation during papillary muscle ablation results in larger, deeper lesions.

Authors:  Udi Nussinovitch; Paul Wang; Sanjiv Narayan; Mohan Viswanathan; Nitish Badhwar; Lijun Zheng; William H Sauer; Duy T Nguyen
Journal:  J Cardiovasc Electrophysiol       Date:  2022-02-15       Impact factor: 2.942

2.  Optimal Ablation Settings Predicting Durable Scar Detected Using LGE-MRI after Modified Left Atrial Anterior Line Ablation.

Authors:  Mathias Forkmann; Christian Mahnkopf; Marcel Mitlacher; Marc Wolff; Beatriz Tose Costa Paiva; Sonia Busch
Journal:  J Clin Med       Date:  2022-02-04       Impact factor: 4.241

3.  In-Silico Modeling to Compare Radiofrequency-Induced Thermal Lesions Created on Myocardium and Thigh Muscle.

Authors:  Juan J Pérez; Enrique Berjano; Ana González-Suárez
Journal:  Bioengineering (Basel)       Date:  2022-07-19