Stephanie Toigo1, Michel Jacques2, Tarek Razek1,3, Ewa Rajda1,4, Sidney Omelon5, Frederic Dankoff6, Rami Tohme7, Patricia Lefebvre8, Dan L Deckelbaum1,3.
Abstract
Objective: Bottlenecks in the personal protective equipment (PPE) supply chain have contributed to shortages of PPE during the COVID-19 pandemic, resulting in fractures in the functionality of healthcare systems. This study was conducted with the aim of determining the effectiveness of retrofitted commercial snorkel masks as an alternative respirator for healthcare workers during infectious disease outbreaks.Entities:
Keywords: COVID-19; fit test; personal protective equipment; respirator; snorkel mask
Year: 2021 PMID: 33926606 PMCID: PMC8209433 DOI: 10.1017/dmp.2021.133
Source DB: PubMed Journal: Disaster Med Public Health Prep ISSN: 1935-7893 Impact factor: 1.385
Figure 1.Retrofitted Aria Ocean Reef® full-face snorkeling mask, for use as a personal protective equipment respirator outside of water.
Fit test results using the Aria Ocean Reef® retrofitted full-face respirator and historical fit test results using medical-grade respirators
| Fit test results of the Aria Ocean Reef® | |||
|---|---|---|---|
| Participants | Pass | Fail | |
| ————————————————— | |||
| Fit test results: Aria Ocean Reef® | 71 (100) | 59 (83.1) | 12 (16.9) |
| Fit test procedure: Aria Ocean Reef® | |||
| Bitrex® | 67 (94.4) | 55 (82.1) | 12 (17.9) |
| PortaCount® | 4 (5.6) | 4 (100) | 0 (0) |
| Model of Aria Ocean Reef® tested | |||
| Small | 38 (53.5) | 36 (94.7) | 2 (5.3) |
| Large | 18 (25.4) | 17 (94.4) | 1(5.6) |
| Both | 1 (1.4) | 0 (0) | 1 (100) |
| Not indicated | 14 (19.7) | 6 (42.9) | 8 (57.1) |
| Fit test results: Medical-grade respirators ( | 65 (90.3) | 53 (81.5) | 12 (18.5) |
| Medical-grade: all failed | 5 (7.7) | 4 (80.0) | 1 (20.0) |
| Medical-grade: passed with one respirator | 25 (38.5) | 19 (76.0) | 6 (24.0) |
| Medical-grade: passed with >1 respirator | 35 (53.8) | 30 (85.7) | 5 (14.3) |
| Mean ± SD (min-max) | 2.2 ± 0.43 (2.0-3.0) | 2.6 ± 0.55 (2.0-3.0) | |
Before testing, participants were able to try both sizes of the retrofitted respirator to select the one that provided the most comfort and then underwent testing with that size of respirator.
A subsample of 65 participants had previously undergone fit testing using medical-grade respirators. The fit test results of these respirators have been presented with those same participants’ fit test results using the Aria Ocean Reef® respirator.
Types of disposable N-95 respirators and reusable half face piece respirators previously tested: 3M™ 1860/ 1860S/ 1870+/ 8200/ 82100, Moldex® 1510/ 1511/ 1512/ 1517, 3M 7501/ 7502/ 7503 and Champak© F550.