Jianxi Qu1, Mathieu Chenier2, Yushan Zhang1, Chang-Qing Xu1,2.
Abstract
A rapid, sensitive and simple microflow cytometry-based agglutination immunoassay (MCIA) was developed for point-of-care (POC) quantitative detection of SARS-CoV-2 IgM and IgG antibodies. The antibody concentration was determined by using the transit time of beads aggregates. A linear relationship was established between the average transit time and the concentration of SARS-CoV-2 IgM and IgG, respectively. The limit of detection (LOD) of SARS-CoV-2 IgM and IgG by the MCIA measurement are 0.06 mg/L and 0.10 mg/L, respectively. The 10 µL sample consumption, 30 min assay time and the compact setup make this technique suitable for POC quantitative detection of SARS-CoV-2 antibodies.Entities:
Keywords: SARS-CoV-2 IgG; SARS-CoV-2 IgM; agglutination immunoassay; microflow cytometry; point-of-care
Year: 2021 PMID: 33919836 PMCID: PMC8070841 DOI: 10.3390/mi12040433
Source DB: PubMed Journal: Micromachines (Basel) ISSN: 2072-666X Impact factor: 2.891
Figure 1A diagram of the microfluidic platform for SARS-CoV-2 antibody detection.
Figure 2Schematic illustration of the MCIA for the quantitative detection of SARS-CoV-2 IgM/IgG.
Figure 3(a) One second of raw data showing the side scattered light signal obtained from a 0.6 mg/L SARS-CoV-2 IgM sample. (b) A detailed view of a side scattered light signal showing the definition of the transit time.
Figure 4Transit time distribution achieved from the detection of (a) negative sample (PBS) and (b) 1 mg/L SARS-CoV-2 IgM.
Figure 5The measured relationship between the average transit time and the concentration of SARS-CoV-2 IgM detected at different days of secondary antibody conjugated MB storage. Each symbol is the mean value of three detections from duplicated samples. Error bars are the standard deviation of the three detections.
Figure 6The measured relationship between the average transit time and the concentration of SARS-CoV-2 IgG detected at different days of secondary antibody conjugated MB storage. Each symbol is the mean value of three detections from duplicated samples. Error bars are the standard deviation of the three detections.