Dosimetric calculations and measurements of gold plaque ophthalmic irradiators using iridium-192 and iodine-125 seeds.

G Luxton1, M A Astrahan, P E Liggett, D L Neblett, D M Cohen, Z Petrovich.   

Abstract

The dosimetry of ophthalmic plaques designed to hold iridium-192 or iodine-125 seeds is investigated experimentally and by means of a computer model. A phantom for thermoluminescent dosimetry (TLD) which permits measurements to within 2 mm of the plaque surface is described. TLD data are compared with model calculations that take into account the active length of the seeds, anisotropy of dose distribution from single seeds, and scatter within the phantom. An isotropic point source calculational model is accurate for clinical calculations, particularly at depths greater than 5 mm. Relative central axis dose measurements for 125I in a gold plaque are also in agreement with the model. Comparisons of 192Ir, 125I and 60Co plaques are presented. The relative advantages of using these isotopes in eye plaques are discussed.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3391814     DOI: 10.1016/0360-3016(88)90362-8

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  2 in total

1.  Gamma irradiation of ocular melanoma and lymphoma cells in the presence of gold nanoparticles: in vitro study.

Authors:  Mozhgan Rezaei Kanavi; Somayeh Asadi; Sahar Balagholi; Fatemeh Alikarami; Hassan Nosrati; Hamid Ahmadieh
Journal:  J Appl Clin Med Phys       Date:  2018-04-29       Impact factor: 2.102

2.  The retina dose-area histogram: a metric for quantitatively comparing rival eye plaque treatment options.

Authors:  Melvin A Astrahan
Journal:  J Contemp Brachytherapy       Date:  2013-03-29
  2 in total