| Literature DB >> 33912643 |
Chunbao Li1, Yaping Huang1, Yubin Xu1.
Abstract
OBJECTIVE: Procalcitonin (PCT) testing adds value in the early detection of infection and sepsis, as well as in management of antibiotic therapy. We determined the analytical and diagnostic performance of four PCT assays at POC.Entities:
Keywords: AQT90 FLEX; Finecare; Getein 1100; Method comparison; PCT; Point-of-care; Procalcitonin; mLabs
Year: 2021 PMID: 33912643 PMCID: PMC8065277 DOI: 10.1016/j.plabm.2021.e00214
Source DB: PubMed Journal: Pract Lab Med ISSN: 2352-5517
Assay specifications of the 5 PCT assays in this study.
| TAT (min) | 21 | <8 | 15 | 15 | 18 |
| Assay stability in months (storage temperature) | 8 (2–8 °C) | 12 (2–8 °C) | 24 (RT) | 24 (4–30 °C) | NA |
| Sample | WB or plasma | WB, plasma, serum, or finger tip | WB, plasma, or serum | WB, plasma, or serum | plasma or serum |
| Pipetting required | No | Yes | Yes | Yes | No |
| Sample volume | 2 mL (for multi-testing) | 250 μL | Pl, S: 50 μL | 3-4 drops | collection tube (for multi-testing) |
| B. Analytical characteristics | |||||
| Reportable range | Pl: 0.08–100 ng/mL | 0.02–100 ng/mL | 0.1–100 ng/mL | 0.1–50 ng/mL | 0.02–100 ng/mL |
| LOD (Functional sensitivity) | Pl: 0.059 ng/mL | 0.02 ng/mL | 0.1 ng/mL | 0.1 ng/mL | 0.06 ng/mL |
| Precision (within run); | <8% | <10% | <10% | <10% | <5% |
All information drawn from current product labeling or manufacturer web-sites. WB: Whole Blood, Pl: Plasma, NA: Not Applicable.
Precision of the 4 POC assays in plasma or whole blood, at low and high levels of PCT.
| A. Plasma | ||||
|---|---|---|---|---|
| Low PCT Level | AQT90 FLEX | mLabs | Finecare | GeTein 1100 |
| Mean ng/mL | 0.66 | 0.67 | 0.59 | 0.42 |
| N | 20 | 20 | 20 | 20 |
| SD | 0.02 | 0.07 | 0.05 | 0.05 |
| CV % | 3.38 | 9.98 | 8.7 | 12.8 |
| High PCT Level | AQT90 FLEX | mLabs | Finecare | GeTein 1100 |
| Mean ng/mL | 2.45 | 2.41 | 1.62 | 1.57 |
| N | 19 | 20 | 20 | 20 |
| SD | 0.05 | 0.47 | 0.18 | 0.07 |
| CV % | 2.1 | 19.64 | 11.21 | 4.6 |
| B. Whole blood | ||||
| Low PCT Level | AQT90 FLEX | mLabs | Finecare | GeTein 1100 |
| Mean ng/mL | 0.24 | 0.77 | 0.35 | 0.35 |
| N | 9 | 10 | 9 | 10 |
| SD | 0.01 | 0.21 | 0.02 | 0.08 |
| CV % | 4.66 | 27.96 | 5.25 | 23.9 |
| High PCT Level | AQT90 FLEX | mLabs | Finecare | GeTein 1100 |
| Mean ng/mL | 1.96 | 2.01 | 4.38 | 12.9 |
| N | 9 | 9 | 10 | 10 |
| SD | 0.05 | 0.59 | 0.44 | 1.6 |
| CV % | 2.7 | 29.51 | 9.96 | 12.51 |
Fig. 1Linearity of the four POC assays in plasma at low or high PCT levels.
Concordance (Cohen’s Kappa) and Agreement between POC assays and reference method.
Fig. 2Bland-Altman analysis of PCT results in plasma for all 5 assays (n = 110). Results with the Cobas assay were compared to the AQT Assay (A), mLabs assay (B), Finecare assay (C), and Getein assay (D.) In order to minimize the effect of outliers we calculated the differences at certain mean PCT concentration as percentages of that mean. To enable a visual comparison, scaling of both axes were kept consistent, causing a few values with means > 10 ng/mL not being depicted, 1 for AQT90, 4 for mLabs, and 3 for both Finecare and Getein.
Fig. 3Passing-Bablok Linear Regression. In the upper diagrams, red lines indicate the best linear fit between the assay and the PCT assay on the Cobas system. The light grey lines indicate the 95% CI around this line of agreement. The bold dotted line indicates total equivalence between the two assays. The stars indicate results outside of the scale of the diagram. Logarithmic scaling was applied at lower PCT levels in to better illustrate the analytical differences between the four POC assays (lower diagrams). (For interpretation of the references to colour in this figure legend, the reader is referred to the Web version of this article.)
Passing-Bablok Linear Regression with “Best Fit” expression.
| Y-Assay | X-Assay | Linear expression | Linearity Test (2-sided, α = 0.05) | Correlation | ||
|---|---|---|---|---|---|---|
| intercept (95% CI) | slope (95% CI) | p = | Pearson’s P-value | R2-value | ||
| AQT90 | Cobas | −0.044 (−0.063 to −0.029) | 1.184 (1.154–1.216) | 0.056 | 0.96 | 0.92 |
| mLabs | Cobas | −0.231 (−0.231 to −0.163) | 1.550 (1.384–1.762) | <0.001 | 0.62 | 0.38 |
| Finecare | Cobas | −0.021 (−0.053 to 0.015) | 0.919 (0.864–0.972) | <0.001 | 0.64 | 0.41 |
| Getein | Cobas | −0.129 (−0.176 to −0.079) | 1.255 (1.173–1.338) | 0.01 | 0.94 | 0.89 |
All expressed in PCT concentrations (ng/mL).