INTRODUCTION: Physical human-robot interaction offers a compelling platform for assessing recovery from neurological injury; however, robots currently used for assessment have typically been designed for the requirements of rehabilitation, not assessment. In this work, we present the design, control, and experimental validation of the SE-AssessWrist, which extends the capabilities of prior robotic devices to include complete wrist range of motion assessment in addition to stiffness evaluation. METHODS: The SE-AssessWrist uses a Bowden cable-based transmission in conjunction with series elastic actuation to increase device range of motion while not sacrificing torque output. Experimental validation of robot-aided wrist range of motion and stiffness assessment was carried out with five able-bodied individuals. RESULTS: The SE-AssessWrist achieves the desired maximum wrist range of motion, while having sufficient position and zero force control performance for wrist biomechanical assessment. Measurements of two-degree-of-freedom wrist range of motion and stiffness envelopes revealed that the axis of greatest range of motion and least stiffness were oblique to the conventional anatomical axes, and approximately parallel to each other. CONCLUSIONS: Such an assessment could be beneficial in the clinic, where standard clinical measures of recovery after neurological injury are subjective, labor intensive, and graded on an ordinal scale.
INTRODUCTION: Physical human-robot interaction offers a compelling platform for assessing recovery from neurological injury; however, robots currently used for assessment have typically been designed for the requirements of rehabilitation, not assessment. In this work, we present the design, control, and experimental validation of the SE-AssessWrist, which extends the capabilities of prior robotic devices to include complete wrist range of motion assessment in addition to stiffness evaluation. METHODS: The SE-AssessWrist uses a Bowden cable-based transmission in conjunction with series elastic actuation to increase device range of motion while not sacrificing torque output. Experimental validation of robot-aided wrist range of motion and stiffness assessment was carried out with five able-bodied individuals. RESULTS: The SE-AssessWrist achieves the desired maximum wrist range of motion, while having sufficient position and zero force control performance for wrist biomechanical assessment. Measurements of two-degree-of-freedom wrist range of motion and stiffness envelopes revealed that the axis of greatest range of motion and least stiffness were oblique to the conventional anatomical axes, and approximately parallel to each other. CONCLUSIONS: Such an assessment could be beneficial in the clinic, where standard clinical measures of recovery after neurological injury are subjective, labor intensive, and graded on an ordinal scale.
Rehabilitation robots have become prominent in the clinical research setting for
applications in motor recovery after neurological injury.[1] Many research trials have been carried out with these devices, and some
improvements in motor scores related to performing activities of daily living (ADLs)
have been observed;[2] however, the best way to perform neurorehabilitation is still unclear. This
is in part due to a limited understanding of the mechanisms of recovery from
neurological injury.[3] In an attempt to increase our understanding of recovery throughout
neurorehabilitation, accurate, descriptive, and repeatable assessments are beginning
to be investigated.[4-8] The only assessments regularly
incorporated into neurorehabilitation though are clinical measures,[9-11] which are typically
subjective, labor intensive, and graded on an ordinal scale.[12] In contrast, robotic measures offer the possibility for objective, efficient,
and descriptive assessments.[13-15]One compelling robot-aided assessment is the evaluation of biomechanical joint properties,[16] such as stiffness and range of motion (ROM). The wrist in particular is an
important joint due to its necessity in performing ADLs[17] and its unique anatomical structure.[18] In a study with cadavers,[18] wrist stiffness and ROM envelopes in the flexion/extension (FE) and
radial/ulnar deviation (RUD) plane were discovered to be ellipses obliquely oriented
to the conventional anatomical axes. The axis of these ellipses for minimum wrist
stiffness and maximum ROM was found to be in the direction of radial-extension to
ulnar-flexion, a direction often termed the dart thrower’s motion. Prior studies
have estimated the dart thrower’s motion axis, through the axis of least stiffness,
to be oriented at an angle of 20–30∘ from extension towards radial
deviation.[18-20]Studying how stiffness and ROM change over the course of neurorehabilitation could be
a promising functional examination of recovery. To perform such an evaluation, an
accurate and reliable method for estimating both stiffness and ROM is needed.
Multi-degree-of-freedom (DOF) wrist stiffness and ROM measurements have only been
measured in a study with cadavers,[18] and the only in vivo 2-DOF wrist stiffness studies have used the InMotion
Wrist rehabilitation robot. As a result of limited device ROM, studies with the
InMotion Wrist have been confined to evaluating wrist stiffness within 70% of ADL
wrist ROM. Additionally, these studies noted limitations on continuous torque
output.[19,21]In contrast with passive stiffness, maximum wrist ROM has received relatively little
attention—though many studies have evaluated ADL ROM through a variety of
means.[17,20,22] Traditionally, ROM has been estimated with electrical goniometers,[22] but this method can have issues with cross coupling when measuring multi-DOF movements.[23] Another method is to use motion capture,[20] which can facilitate ROM assessment of several joints. A drawback of motion
capture is its large physical footprint, non-trivial setup time, and issues with
marker occlusion and slipping. On the other hand, a robot is the only platform
capable of measuring both stiffness and ROM, simplifying any need for accurate
alignment between setups. Such a robot has not yet been realized, possibly related
to the fact that the majority of robotic assessments are performed with devices
designed for robotic rehabilitation, which has its own design considerations such as
inherent backdrivability and generally the goal of supporting ADL ROM.While wrist stiffness and ROM envelopes could provide new insights into understanding
recovery throughout neurorehabilitation, the wrist rehabilitation robots developed
thus far do not have sufficient ROM. In this work, the development of the series
elastic (SE)-AssessWrist, a novel 2-DOF wrist exoskeleton for biomechanical wrist
stiffness and ROM assessment, is described and validated (see Figure 1). The SE-AssessWrist overcomes
limitations in using a traditional rehabilitation robot for wrist stiffness or ROM
assessment by employing series elastic actuation in conjunction with a laterally
flexible Bowden cable transmission and remotely located geared DC motors. The
essential features of the SE-AssessWrist are characterized in the context of the
intended wrist stiffness and ROM assessments. The paper concludes with a validation
study (n = 5) with able-bodied participants, illustrating the
potential for wrist biomechanical assessment in the clinic.
Figure 1.
(Left) SE-AssessWrist isometric view. (1) Hand attachment point, (2) series
elastic FE joint, (3) FE motors, (4) series elastic RUD joint, and (5) RUD
motors. The FE and RUD motors actuate their series elastic joints through a
Bowden cable transmission. (Center) Device DOFs: (6) FE, (7) RUD, and (8) a
passive linear joint for wrist alignment. Wrist FE and RUD are measured
through the encoders at the output of the series elastic elements placed
on-board the device. (Right) A user interacting with the SE-AssessWrist: (9)
open hand attachment through a Velcro strap, (10) elastic element connected
to the input pulley, (11) RUD encoder cables, (12), wrist support, and (13)
elbow support.
(Left) SE-AssessWrist isometric view. (1) Hand attachment point, (2) series
elastic FE joint, (3) FE motors, (4) series elastic RUD joint, and (5) RUD
motors. The FE and RUD motors actuate their series elastic joints through a
Bowden cable transmission. (Center) Device DOFs: (6) FE, (7) RUD, and (8) a
passive linear joint for wrist alignment. Wrist FE and RUD are measured
through the encoders at the output of the series elastic elements placed
on-board the device. (Right) A user interacting with the SE-AssessWrist: (9)
open hand attachment through a Velcro strap, (10) elastic element connected
to the input pulley, (11) RUD encoder cables, (12), wrist support, and (13)
elbow support.
Methods
The following section provides details on the implementation of the mechanical design
necessary for performing wrist biomechanical assessment using the SE-AssessWrist.
Details on the mechanical design highlight the considerations for meeting the ROM
and torque requirements. Additionally, the methodology for experimental validation
of the robotic device with n = 5 participants is described.
SE-AssessWrist design
Mechanical and control requirements
To measure user stiffness, a key requirement for the SE-AssessWrist is to be
able to move the user’s joint slowly, while also having sufficient torque
output and resolution. In previous studies, passive wrist stiffness
estimates were carried out by having the robot move at a velocity of
approximately 0.1–0.2 rad/s, typically to the limits of the device ROM.
Additionally, these studies noted limitations based on the device torque
output of 1.95 N ⋅ m.[19,21] As such, requirements
for the SE-AssessWrist DOFs were set to provide a lower limit on velocity of
0.1 rad/s and 3 N ⋅ m of continuous torque output. The requirement on torque
resolution was set to 10 N ⋅ mm to provide sufficient stiffness estimation
resolution.To measure wrist ROM, the robotic device must have a device ROM on par with
the human wrist. We estimate an upper bound on maximum human wrist ROM to be
185∘ in FE and 100∘ in RUD.[18,24] Note that designing
for maximum wrist ROM is a different challenge than designing for ADL ROM,
which is the target for most wrist rehabilitation robots.[17,25,26]
Additionally, during ROM assessment, the device needs to be sufficiently
transparent so that a user can comfortably operate the device to explore
their own ROM limits. Most wrist rehabilitation robots are inherently
backdrivable, but as a result of the cables and motors used for this
backdrivability, such devices typically have static friction on the order of
0.1–0.2 N ⋅ m.[25,27,28] The maximum target
torque to backdrive the SE-AssessWrist in either joint was set to
0.2 N ⋅ m.
Kinematic structure
Since the standard in wrist modeling, and also exoskeleton design, is to
place wrist FE as the first axis and RUD second,[25,32] the SE-AssessWrist
follows this convention with a serial revolute-revolute exoskeletal
structure. With this serial structure, the user’s wrist FE and RUD joint
angles are measured directly by the output encoders placed at the series
elastic actuated joints (see Figures 1 and 2). A passive, and unmeasured,
prismatic joint at the handle is also included for wrist alignment purposes.
The user interfaces with the device through an open hand configuration,
facilitating a natural and relaxed position beneficial to passive stiffness estimation.[21]
Figure 2.
(a) Custom-designed double Archimedes spiral spring with an
integrated hub (1) for mating to its output shaft. (b) The spring is
connected to an input pulley, which includes an indent (2) so that
the two surfaces only touch on a small area where spring deflection
is minimal. (c) Exploded CAD rendering of the spring and pulley,
including the six 6–32 screws (3) and four dowel pins (4) used to
mate the two components. (d) A CAD rendering of the connected spring
and pulley. A cable (not shown) runs in a race on the pulley and is
anchored at the pulley via a set screw that is tightened through the
thru-hole on the spring (5). (e) Spring sub-assembly including the
encoder hubdisks (6) and spring output shaft (7).
(a) Custom-designed double Archimedes spiral spring with an
integrated hub (1) for mating to its output shaft. (b) The spring is
connected to an input pulley, which includes an indent (2) so that
the two surfaces only touch on a small area where spring deflection
is minimal. (c) Exploded CAD rendering of the spring and pulley,
including the six 6–32 screws (3) and four dowel pins (4) used to
mate the two components. (d) A CAD rendering of the connected spring
and pulley. A cable (not shown) runs in a race on the pulley and is
anchored at the pulley via a set screw that is tightened through the
thru-hole on the spring (5). (e) Spring sub-assembly including the
encoder hubdisks (6) and spring output shaft (7).
Bowden cable transmission
To achieve the desired torque and ROM, a Bowden cable transmission was
selected. A Bowden cable transmission is advantageous for attaining large
ROM while maintaining necessary torque output since a high torque actuator
can be housed off-board without affecting joint compactness or
mass.[33-35] A design trade-off for using a Bowden cable
transmission is that it adds friction to the system. This friction can be
alleviated partly through careful consideration of materials and Bowden
cable construction. Characterization of Bowden cable force transmission has
shown that adding a Teflon liner between the cable and conduit increases
force transmissivity.[36] As a result, the SE-AssessWrist Bowden cables are constructed from a
nylon coated 7 × 19 stainless steel braided cable and a slick lubed inner
tubing conduit. To minimize the pretension required on the cables, two
motors per DOF are used since pretension increases static friction in the
transmission, which in turn negatively impacts control performance.
Series elastic actuation
To estimate torque accurately and to enable backdrivability despite the
Bowden cable transmission, the SE-AssessWrist adopts a series elastic
actuation architecture.[37,38] Similar to Stienen et al.,[39] a rotational elastic element was realized with a double Archimedes
spiral made from aluminum 7075-T651. The spring was manufactured with
computer numerical control machining, which enabled a monolithic design for
the spring and hub. The elastic elements (see Figure 2) were incorporated on the
exoskeleton frame to provide accurate torque estimation at the output.To measure spring deflection, US Digital’s EM2 10,000 counts per revolution
transmissive optical encoder module with a 50.8 mm diameter transmissive
rotary hubdisk was selected and placed at both ends of the spring. This
encoder was chosen for its low cost, compactness and ease of use. The
encoder leads to a wrist position resolution of 1.57 rad. The spring rate of the elastic elements were
characterized for 3 N ⋅ m peak torque and found to be 75.96 and
77.23 N ⋅ m/rad, corresponding to a torque resolution of 12 N mm. The
springs showed linearity within this region with (see Figure 3). Further information regarding the mechanical design
of the spring, including details on finite element analysis, can be found in
our previous work.[40]
Figure 3.
Quasi-static characterization of the custom rotary spring with a
ground-truth torque sensor. The spring rate was estimated to be
75.96 N ⋅ m/rad, which closely matched the finite element analysis
prediction of 73 N ⋅ m/rad. Only one spring characterization plot
has been included for visualization. Finite element analysis of the
spring was performed through a static simulation in Solidworks by
placing a fixed constraint at the spring and applying a torque to
the pulley.
Quasi-static characterization of the custom rotary spring with a
ground-truth torque sensor. The spring rate was estimated to be
75.96 N ⋅ m/rad, which closely matched the finite element analysis
prediction of 73 N ⋅ m/rad. Only one spring characterization plot
has been included for visualization. Finite element analysis of the
spring was performed through a static simulation in Solidworks by
placing a fixed constraint at the spring and applying a torque to
the pulley.
Motor selection
To achieve the desired torque requirements, Maxon Motor’s RE40-148877 was
chosen. Since this motor can output 0.187 N ⋅ m of continuous torque, torque
amplification was required to meet the 3 N ⋅ m requirement. To accomplish
this, a planetary gearhead with a 43:1 gear ratio was selected. This gearbox
can sustain a continuous torque output of 15 N ⋅ m and a peak torque of
22.5 N ⋅ m.
SE-AssessWrist mechanical properties
Device ROM is defined as when the device either makes contact with itself or
reaches some physical constraint (such as available cable length). In
particular, any serial device that follows the wrist convention of FE carrying
the RUD axis will inevitably make contact with the user or itself, which will
determine the ROM limits. For the SE-AssessWrist, device ROM is given as two
rotational directions (wrist extension and ulnar deviation) in which the device
makes contact with itself and in the other two directions (wrist flexion and
radial deviation), by the maximum rotation of the human wrist. The device was
designed to contact itself in 65° of extension and 60° of ulnar deviation while
maximum allowable limits of the human wrist are estimated to be 120° in flexion
and 40° in radial deviation.Continuous torque output was estimated by considering the motor torque constant
and gear ratio, as well as inefficiencies in the gearbox (72%) and Bowden cable
transmission (85%).[36] The resulting continuous torque estimate is 4.9 N ⋅ m, while the maximum
torque is limited by the predicted 5.1 N ⋅ m spring yield torque.Table 1 compares
wrist ROM and torque specifications of the SE-AssessWrist with other wrist
robots, all of which have been designed for robot-aided rehabilitation as
opposed to robot-aided assessment. An image of a user connected to the
SE-AssessWrist is shown in Figure 1. To see the device operating, see our supplementary video
attachment.
Table 1.
Range of motion and torque output of the SE-AssessWrist and other
prominent wrist robots.
Wrist flexion/extension
Wrist radial/ulnar deviation
ROM (deg)
Torque (Nm)
ROM (deg)
Torque (Nm)
SE-AssessWrist
185
4.9
100
4.9
ARMin-II[29]
85
N.A.
–
–
CADEN-7[17]
120
N.A.
60
N.A.
IIT Wrist[30]
144
1.53
72
1.63
InMotion Wrist[27]
120
1.43
75
1.43
MAHI Exo-II[31]
72
1.67
72
1.93
RiceWrist-S[25]
130
3.37
75
2.11
OpenWrist[28]
135
3.60
75
2.30
WRES[26]
75
1.62
40
1.62
ROM: range of motion.
Range of motion and torque output of the SE-AssessWrist and other
prominent wrist robots.ROM: range of motion.
Control strategy
Stiffness assessment
Wrist stiffness is evaluated by using position control on the robot’s joints
to slowly move the user’s wrist while recording the user’s joint position
through the output encoders and interaction torque from spring deflection.
Due to the use of two motors for a single-DOF, each robot DOF is
over-actuated (see Figure
4); however, since the cables only produce rotational motion of
the output pulley when under tension, each motor controls a single
direction. As such, for the constant velocity needed in the stiffness
assessment experiments, each joint consisted of a leader and follower motor.
The lead motor was commanded to follow a desired trajectory through
proportional-derivative (PD) control while the follower motor was sent a
constant negative torque command. This control approach was implemented as
follows: where τ is the motor torque,
desired load (or equivalently joint or output) position,
θ load position, load velocity, k proportional
gain, k derivative gain,
τ feed-forward torque, and
τ follow motor torque.
Figure 4.
Block diagram illustrating the SE-AssessWrist’s series elastic
actuated, Bowden cable transmission scheme.
Block diagram illustrating the SE-AssessWrist’s series elastic
actuated, Bowden cable transmission scheme.The feed-forward torque was found experimentally by slowly increasing the
motor torque and recording the torque required to initiate movement through
the Bowden cable transmission at the output pulley. Feed-forward
compensation of friction was found to improve the initial portion of the
position control trajectory, as oscillations introduced from backlash and
friction were reduced. On the other hand, the follower torque was found
experimentally such that it provided sufficient slack in the cable so that
the lead motor would not have to overcome friction present in the follower’s
transmission.
Range of motion assessment
Wrist ROM is evaluated by measuring a user’s motion with the robot’s joint
encoders placed at the series elastic joints while the device is operated
through zero force control. Since the SE-AssessWrist is not backdrivable in
the traditional sense, the purpose of the zero force controller is to allow
the user to more freely explore their wrist ROM while reducing resistance
from the device. With a series elastic actuator, zero force control is
equivalent to specifying the desired spring deflection to be zero
(). To achieve zero force control despite the static
friction in the transmission, a controller that leverages the capabilities
of the device to perform accurate position control was chosen.[41] To regulate torque in this force control approach, the motor attempts
to maintain .As in all series elastic devices, a given DOF cannot regulate lower torque
than that of its sensing resolution. Additionally, to overcome backlash, the
device’s default state in this control mode is to provide tension on both
sides of the spring such that the user can create a torque to inform the
controller to perform active zero force control. Once a deadzone limit
τ is exceeded, the zero force
controller is implemented. This control approach is given as where if τ is exceed, the zero
force control is implemented on the lead motor while the follower motor is
sent a constant torque command as in the position controller. A block
diagram of the force control approach is shown in Figure 5.
Figure 5.
Zero force control block diagram for the lead motor used for
backdrivability in the range of motion assessment.
Zero force control block diagram for the lead motor used for
backdrivability in the range of motion assessment.
Experimental validation
A validation study was conducted using the SE-AssessWrist to measure 2-DOF active
wrist ROM and passive stiffness of able-bodied individuals. Five participants (1
female, 4 male) with an age range of 22-32 years old (mean = 26.4 yr,
SD = 4.16 yr) participated in the experiment. All participants were right hand
dominant with no current injury or known history of neuromuscular injury in
their wrist. Approval for the experiment was obtained through the Rice
University Institutional Review Board (IRB-FY2018-331).
Measuring wrist muscle activity
Muscle activity was measured during the stiffness trials to determine if the
user’s muscles were passive. To measure activity relating to the wrist, a
surface electromyography (sEMG) electrode was placed on each of the four
wrist muscles: flexor carpi radialis, extensor carpi ulnaris, flexor carpi
ulnaris, and extensor carpi radialis longus. Prior to donning the
exoskeleton, participants were equipped with the sEMG electrodes and
performed three repetitions of maximum voluntary contraction for both FE and
RUD. The maximum signal for each muscle was recorded and used for estimating
user passivity.
Wrist alignment
During the experiment, participants sat in a chair with a posture consisting
of moderate shoulder flexion, shoulder abduction, elbow flexion, and a
neutral forearm orientation. To isolate wrist motion, a cuff was comfortably
compressed around the forearm near the elbow. The neutral orientation of the
forearm was defined visually with the top of the radial styloid in line with
the device’s FE rotational axis. The neutral orientation of the wrist was
defined with respect to having an open grasp. As in Crisco et al.[18] and similar to the definitions in Wu et al.,[32] neutral wrist orientation was defined visually by aligning the dorsal
surfaces of the forearm and hand until flush (FE neutral), and then aligning
the third metacarpal’s long axis to be parallel to the forearm’s long axis
(RUD neutral).
Range of motion measurement
To find a user’s maximum active wrist ROM, each participant was asked to
actively move the farthest they could in a set of 24 directions while the
device was operated under the zero force control scheme. Angles chosen
aligned with the traditional anatomical FE/RUD axes, and directions spaced
15° apart. Participants repeated the 24 movements 3 times with the
directions being presented in 3 blocks. Within each block, the movement
directions were randomized to balance learning effects. To assist with
finding ROM, participants were presented with a virtual display that
contained a cursor identifying the user’s 2D position (]), and a line with angle in the direction of the desired movement (see Figure 6).
Figure 6.
Graphical user interface displayed to participants to facilitate
measuring active ROM. The axes and text have been added for
visualization.
Graphical user interface displayed to participants to facilitate
measuring active ROM. The axes and text have been added for
visualization.A user interacting with the SE-AssessWrist during the study.
The image highlights the arm cuffs, human-robot interface, and
off-board actuators. (1) Forearm support, (2) visual display, (3)
distal wrist cuff, (4) exoskeleton interface, (5) SE-AssessWrist,
(6) FE motors, and (7) RUD motors.For ease of visual interpretation, positive was defined as being in the wrist
extension and radial deviation directions. Participants were asked to move
in the direction of the line as far as they comfortably could and to then
return to the origin prior to the next movement. So as to provide
directional information while not biasing participants’ ROM movement
attempts, the extreme point of the line was a value (150°) that could not be
achieved by any participant. Prior to data collection, participants were
allowed to practice the task for a few trials to familiarize themselves with
the experiment.
Passive stiffness measurement
To measure a user’s passive stiffness, the robot moved the participant’s
wrist through position control while the user attempted to remain passive,
i.e., to not interact or invoke muscle activity. The wrist was moved by the
SE-AssessWrist in 24 equally spaced directions 3 times starting from wrist
extension and moving counterclockwise as is standard in the
literature.[18,19,21] The robot used the user-specific ROM found in the
prior experiment as the position limits for stiffness estimation. Since it
can be counterintuitive to remain passive while a limb is being moved,
participants were given a familiarization trial (approximately 2–3 minutes)
prior to data collection. To move the wrist, the robot was commanded to
follow a ramp position trajectory (constant velocity) with a velocity
magnitude of 0.2 rad/s so as to assist with avoiding muscle activity due to
stretch reflex.To analyze the stiffness data, each movement was segmented. Segments started
once the position magnitude was greater than 3° to eliminate any short-range
stiffness effects, or oscillations introduced at startup. To avoid using
data in the estimate after the robot stopped moving, segments were
terminated once the reference position was the final position for a given
movement. The linear stiffness of the wrist in that direction, which is the
magnitude of the 2-DOF directional stiffness, was then calculated using
multiple-linear regression for outbound movements. The position and torque
signals were not down sampled or filtered.Position dependent torques, due to device gravity, encoder misalignment, or
conduit flex, were subtracted from the measured joint torques. These torques
were characterized prior to the experiment by recording joint torques while
the device swept the FE and RUD workspace with a representative mass
attached at the handle. Additionally, since the focus of this work is on
validation of the SE-AssessWrist, and not biomechanics of the general
population, we present results from a subset of the passive wrist stiffness
experiment. This subset consisted of trials where the participant viewed a
display with their processed sEMG amplitude for biofeedback. They were
instructed to try to minimize their sEMG amplitude. Prior to the biofeedback
sEMG condition, participants underwent the stiffness experiment without any
biofeedback.
Results
Control performance
Characterization of the SE-AssessWrist control schemes was performed through
real-time software implemented in a Matlab-Simulink environment communicating
with Quanser’s Q8-USB data acquisition board sampled at 1000 Hz. Velocity
estimates of encoder positions were obtained through the Q8’s built-in
instantaneous velocity estimator. Analog voltage commands from the Q8-USB were
sent to servo amplifiers (Advanced Motion Controls AMC 12A8), which converted
the voltage commands to current control the brushed DC motors.
Position control
The position control law in equation (1) was evaluated
in an experiment without a user. During the experiment, the device made
movements to 12 targets spaced evenly in the FE and RUD space. Gains of
N ⋅ m/rad and N ⋅ ms/rad were selected (note that vectors are given as
[FE RUD]). Low PD gains were used for the benefit of reducing oscillations
in the velocity output. The feed-forward and follower torques were set to be
τ = 0.3 N ⋅ m and
τ = –0.1 N ⋅ m. As can be seen in Figure 8, the device
is able to track the desired position trajectory well, but with some steady
state error, . The average absolute error over the experiment was
rad with a maximum absolute error of rad, while regulating a velocity with a standard deviation
of rad/s about the desired velocity.
Figure 8.
Position control performance with low PD control gains to various
ramp position commands for the FE (left) and RUD (right). Only three
movement portions of the experiment are shown for visualization.
While the low gains result in some steady state error, which is
acceptable since stiffness estimates are based off of measured joint
positions, they more importantly produced the desired constant
velocity with lower variance than with higher PD control gains.
Position control performance with low PD control gains to various
ramp position commands for the FE (left) and RUD (right). Only three
movement portions of the experiment are shown for visualization.
While the low gains result in some steady state error, which is
acceptable since stiffness estimates are based off of measured joint
positions, they more importantly produced the desired constant
velocity with lower variance than with higher PD control gains.
Zero force control
The zero force control approach described in equation (2) was implemented
while an experimenter moved the device to 12 equally spaced targets in the
FE/RUD space. The experimenter moved their wrist at a pace expected during
the ROM portion of the validation study. The zero force controller used PD
gains of N ⋅ m/rad and N ⋅ ms/rad, as well as a deadzone of = [0.15 0.15] N ⋅ m. The results of this experiment are
shown in Figure 9.
The user was able to actively backdrive the device to find their ROM limits,
which were within those of the device’s limits. Additionally, the spring
torque, , during the experiment was low with N ⋅ m and a maximum absolute torque of = N ⋅ m.
Figure 9.
(Left) The ROM of the SE-AssessWrist (blue, dashed line) was
evaluated by manually moving the device within its limits. Plotted
for comparison are the ROM capabilities of the InMotion Wrist robot[19] (red, dash-dot line) and the maximum ROM found in the cadaver
studies in Crisco et al.[18] (black, small-width dashed line). The SE-AssessWrist ROM
exceeds prior rehabilitation robot designs, and is capable of
measuring the maximum ROM found in the work of Crisco et al. A user
(solid, green line) explored their ROM using the SE-AssessWrist
while it was operated under zero force control so that the user
could backdrive the device. Forces felt by the user in FE (left) and
RUD (right) resulted in a perceived virtual friction due to the
deadzone in the zero force controller. The magnitude of this
perceived friction is on par with the 0.1–0.2 N⋅m of static friction
present in existing wrist rehabilitation robots.
(Left) The ROM of the SE-AssessWrist (blue, dashed line) was
evaluated by manually moving the device within its limits. Plotted
for comparison are the ROM capabilities of the InMotion Wrist robot[19] (red, dash-dot line) and the maximum ROM found in the cadaver
studies in Crisco et al.[18] (black, small-width dashed line). The SE-AssessWrist ROM
exceeds prior rehabilitation robot designs, and is capable of
measuring the maximum ROM found in the work of Crisco et al. A user
(solid, green line) explored their ROM using the SE-AssessWrist
while it was operated under zero force control so that the user
could backdrive the device. Forces felt by the user in FE (left) and
RUD (right) resulted in a perceived virtual friction due to the
deadzone in the zero force controller. The magnitude of this
perceived friction is on par with the 0.1–0.2 N⋅m of static friction
present in existing wrist rehabilitation robots.
Robotic assessment experimental validation
Active range of motion measurement
From the active ROM measurements, participants’ maximum values were
calculated for each movement direction. These values are plotted as a 2-DOF
wrist ROM envelope in Figure 10. The mean active ROM measurements in this work are
69.8° in flexion, 53.6° in extension, 48.0° in ulnar deviation, and 30.1° in
radial deviation. The axis of greatest ROM found in this study was in the
direction of radial-extension to ulnar-flexion, making an angle of 30∘ from
wrist extension in the direction of radial-extension. The average completion
time for the active ROM experiment was 5 minutes and 56 s.
Figure 10.
Results from the (left) active range of motion and (right) passive
stiffness experiments. For the range of motion experiment, the dots
represent each participant’s maximum directional value. The solid
black line is the mean of these maximum values for a given
direction. In the wrist stiffness plot, the dots correspond to the
average directional stiffness for a given participant over the three
trials. The solid black line is the mean of these stiffness values
across participants for a given direction. As found in a study with cadavers,[18] the stiffness and range of motion plots are ellipsoidal in
shape and oriented obliquely to the conventional anatomical axes.
The dashed red lines indicate the axes of greatest range of motion
and least stiffness, which are both in the wrist radial-extension to
ulnar-flexion direction.
Results from the (left) active range of motion and (right) passive
stiffness experiments. For the range of motion experiment, the dots
represent each participant’s maximum directional value. The solid
black line is the mean of these maximum values for a given
direction. In the wrist stiffness plot, the dots correspond to the
average directional stiffness for a given participant over the three
trials. The solid black line is the mean of these stiffness values
across participants for a given direction. As found in a study with cadavers,[18] the stiffness and range of motion plots are ellipsoidal in
shape and oriented obliquely to the conventional anatomical axes.
The dashed red lines indicate the axes of greatest range of motion
and least stiffness, which are both in the wrist radial-extension to
ulnar-flexion direction.Participants’ stiffness for each of the 24 directions was calculated as the
average of the 3 measurement repetitions. Stiffness was calculated using
only the torque contribution in line with the movement, thus representing
the restoring stiffness. The 2-DOF stiffness envelope from these
measurements is shown in Figure 10. The axis of least stiffness found in this study was
in the direction of radial-extension to ulnar-flexion, making an angle of
22.5° from wrist extension in the direction of radial-extension.The maximum torque for each participant across stiffness measurements ranged
from 0.8 N ⋅ m to 1.5 N ⋅ m (0.3 N ⋅ m standard deviation), while the
average wrist torque ranged from 0.19 to 0.31 N ⋅ m (0.05 N ⋅ m standard
deviation). Mean sEMG activity across all muscles for all participants was
less than 0.67% of maximum voluntary contraction. This sEMG activity is
similar to the approximately 1–3% of maximum voluntary contraction found
during passive wrist stiffness measurements in Pando et al.,[19] providing evidence of user compliance with the instruction to relax
their wrist muscles during this portion of the experiment.
Discussion
Both the stiffness and ROM envelopes were observed to be oblique to the conventional
anatomical axes. We found that the directions of maximum ROM and least stiffness
were aligned with a direction from wrist extension/radial deviation to flexion/ulnar
deviation, a direction often termed the “dart thrower’s motion”, and that these axes
were nearly parallel to each other. The results are consistent with observations
made in a cadaver study,[18] which also identified the obliqueness of the wrist stiffness and ROM
envelopes, giving some support to the validity of the SE-AssessWrist for evaluating
2-DOF wrist biomechanics. While this validation study demonstrates the capabilities
of the SE-AssessWrist, due to the small sample size used in this work, it does not
allow us to make conclusions about the wrist biomechanical properties of the general
population. Compared with previous studies that investigated wrist stiffness in vivo
with devices designed for ADL ROM,[18,19,21] this work includes both wrist
stiffness and maximum ROM assessment.Further merits of our approach include being able to investigate passive stiffness
over the user-specific ROM. Previously, ROM estimates have been carried out based on
limitations in device ROM.[19,21] Being able to investigate stiffness over the user-specific ROM
could enable more complete evaluations of wrist end point stiffness, and ensure that
the user’s ROM is not exceeded. Additionally, in contrast with the open-loop torque
control employed in rehabilitation robots, leading to an estimate of wrist stiffness
from motor current, the SE-AssessWrist has direct torque estimation for assessing
wrist stiffness. Having a device that can assess active wrist ROM, could be
beneficial for providing a ground truth measurement of passive ROM, which could be
beneficial for situations in which the user cannot voluntarily move to their
limits.In addition to the experimental validation study, demonstration of device performance
for the assessment application was presented in two characterization experiments.
Position control experiments highlighted accurate trajectory following despite the
Bowden cable transmission. Additionally, since the device is non-backdrivable, an
active transparent mode was implemented through a zero force controller to enable
wrist ROM assessment. In this way, the device was able to accurately regulate the
spring torque to within a small deadzone of 0.15 N ⋅ m, similar to the friction of
0.1–0.2 N ⋅ m found in other wrist exoskeletons.[25,27,28] The PD control gains used in
the robotic assessment experimental validation were similar to those used in the
control performance characterization experiments. Prior to clinical use, control
performance should be increased through automated adaptive PD control gains for each
user, as well as incorporating a model of the human wrist.Robot-aided assessment is not the only platform for measuring wrist ROM. Motion
capture provides a means for not only measuring wrist ROM, but potentially ROM of
the entire body. While it might be possible to study ROM with a passive device or
through motion capture, neither solution would offer the possibility of assessing
both stiffness and ROM through the same mechanism, which could be beneficial for
repeatability, accuracy, and efficiency. By assessing both properties with the same
device, wrist alignment can be maintained, allowing for a direct comparison between
the two. Additionally, motion capture suffers from issues such as requiring a large
setup area, a substantial setup time, and can suffer from marker occlusion and slip.
While it might be useful to “calibrate” a robot using motion capture, given the
prior limitations, it might be challenging to use ROM from motion capture as a
ground truth measurement. Limitations of the developed robot are its larger
footprint compared with other wrist rehabilitation robots due to the Bowden cable
transmission, as well as the increase in actuators required for each joint.Although in this work we have focused on the SE-AssessWrist’s potential applicability
for wrist assessment, such a platform could enable other research areas. For
example, the SE-AssessWrist might be used for biologically-inspired actuation and
control paradigms. In particular, the mechanical actuation architecture has
similarities to the three-element Hill muscle model, including series elasticity
from the Bowden cable flexibility, as well as parallel actuation (cables) and
stiffness (series elastic element). The device could also serve as a nonlinear
controls platform, which would benefit from a mathematical model of the system. A
complete mathematical model of the system is left for future investigations since
the focus in this work was on the viability of the assessment approach and not to
optimize control performance. Additionally, while the device was designed for the
slow movements for estimating passive wrist stiffness, the device is modular, and
readily adaptable for other applications—such as a low-inertia yet high-powered
system.
Conclusions
In this work, we presented the development, control, and experimental validation of
the SE-AssessWrist: a serial 2-DOF series elastic actuated Bowden cable-based
exoskeleton. As found through experiments, the SE-AssessWrist has both the necessary
mechanical properties and control performance to measure complete wrist ROM and
stiffness. Additionally, through an experiment with five right-hand dominant,
able-bodied individuals, we confirmed a finding originally reported in a cadaver
study—that wrist stiffness and ROM envelopes are oblique to the conventional
anatomical axes along a direction of least stiffness and maximum ROM termed the
“dart thrower’s motion”. Prior to clinical studies, future investigations should
determine the reliability of these measurements through repeated trials comparing
estimates from robotic measurements to those made by ground truth sensors.In the future, this device could be used in experiments with able-bodied individuals
to develop a database of nominal wrist stiffness and ROM, which could serve as a
reference for wrist biomechanical assessment of neurologically impaired individuals.
Such reference data could be appropriately scaled to individuals with wrist
sensorimotor control impairments after neurological injury. After the end of an
assessment session, the clinician could present the comparison of the magnitude and
orientation (e.g. the angle of the “dart thrower’s motion”) of the user’s stiffness
and range of motion profiles with the reference set. Using the device to study wrist
biomechanical impairment after neurological injury, such as stroke, could reveal
insights into the evolution of wrist stiffness and ROM envelopes throughout
rehabilitation, which might lead to important insights into the recovery of wrist
function.
Authors: Hermano Igo Krebs; Bruce T Volpe; Dustin Williams; James Celestino; Steven K Charles; Daniel Lynch; Neville Hogan Journal: IEEE Trans Neural Syst Rehabil Eng Date: 2007-09 Impact factor: 3.802
Authors: Arno H A Stienenw; Edsko E G Hekman; Huub ter Braak; Arthur M M Aalsma; Frans C T van der Helm; Herman van der Kooij Journal: IEEE Trans Biomed Eng Date: 2009-04-07 Impact factor: 4.538
Authors: Zachary A Wright; Emily Lazzaro; Kelly O Thielbar; James L Patton; Felix C Huang Journal: IEEE Trans Neural Syst Rehabil Eng Date: 2017-10-16 Impact factor: 3.802