Multiplex Ventilation: Solutions for Four Main Safety Problems.

Morgan E Sorg1, Richard D Branson2, Umur Hatipoğlu2, Robert L Chatburn2.   

Abstract

BACKGROUND: The COVID-19 pandemic has led to an increased demand for mechanical ventilators and concerns of a ventilator shortage. Several groups have advocated for 1 ventilator to ventilate 2 or more patients in the event of such a shortage. However, differences in patient lung mechanics could make sharing a ventilator detrimental to both patients. Our previous study indicated failure to ventilate in 67% of simulations. The safety problems that must be solved include individual control of tidal volume (VT), individual measurement of VT, individualization of PEEP settings, and individual PEEP measurement. The purpose of this study was to evaluate potential solutions developed at our institution.
METHODS: Two separate lung simulators were ventilated with a modified multiplex circuit using pressure control ventilation. Parameters of the lung models used for simulations (resistance and compliance) were evidence-based from published studies. Individual circuit-modification devices were first evaluated for accuracy. Devices were an adjustable flow diverter valve, a prototype dual volume display, a PEEP valve, and a disposable PEEP display. Then the full modified multiplex circuit was assessed by ventilating 6 pairs of simulated patients with different lung models and attempting to equalize ventilation. Ventilation was considered equalized when VT and end-expiratory lung volume were within 10% for each simulation.
RESULTS: The adjustable flow diverter valve allowed volume adjustment to 1 patient without affecting the other. The average error of the dual volume display was -17%. The PEEP valves individualized PEEP, but the PEEP gauge error ranged from 17% to 41%. Using the multiplex circuit, ventilation was equalized regardless of differences in resistance or compliance, reversing the "failure modes" of our previous study.
CONCLUSIONS: The results of this simulation-based study indicate that devices for individual control and display of VT and PEEP are effective in extending the usability and potential patient safety of multiplex ventilation.
Copyright © 2021 by Daedalus Enterprises.

Entities:  

Keywords:  COVID-19; differential ventilation; dual ventilation; emergency ventilation; lung simulator; mechanical ventilation; national stockpile; simultaneous ventilation; split ventilation; surge ventilation; ventilator sharing; ventilator shortage

Year:  2021        PMID: 33906955     DOI: 10.4187/respcare.08749

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  1 in total

1.  Translational design for limited resource settings as demonstrated by Vent-Lock, a 3D-printed ventilator multiplexer.

Authors:  Helen Xun; Christopher Shallal; Justin Unger; Runhan Tao; Alberto Torres; Michael Vladimirov; Jenna Frye; Mohit Singhala; Brockett Horne; Bo Soo Kim; Broc Burke; Michael Montana; Michael Talcott; Bradford Winters; Margaret Frisella; Bradley S Kushner; Justin M Sacks; James K Guest; Sung Hoon Kang; Julie Caffrey
Journal:  3D Print Med       Date:  2022-09-14
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.