Performance evaluation of the QIAstat-Dx® Respiratory SARS-CoV-2 Panel.

Samuel Lebourgeois1, Alexandre Storto1, Bernard Gout2, Quentin Le Hingrat1, Gustave Ardila Tjader3, Maria Del Carmen Cerdan4, Alistair English5, Josep Pareja4, Joanna Love4, Nadhira Houhou-Fidouh3, Davide Manissero5, Diane Descamps1, Benoit Visseaux6.   

Abstract

OBJECTIVE: The aim of this study was to evaluate the QIAstat-Dx® Respiratory SARS-CoV-2 Panel (QIAstat-SARS-CoV-2), which is a closed, fully automated, multiplex polymerase chain reaction (PCR) assay that detects severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) and 21 other pathogens that cause respiratory disease.
METHODS: Nasopharyngeal swabs from patients with or suspected of having coronavirus disease 2019 were collected and tested at Bichat-Claude Bernard Hospital, Paris, France. Using the World Health Organisation-approved real-time-PCR assay developed by the Charité Institute of Virology as the reference, positive percent agreement (PPA) and negative percent agreement (NPA) were calculated.
RESULTS: In total, 189 negative and 88 positive samples were analyzed. QIAstat-SARS-CoV-2 had an NPA of 90.48% (95% confidence interval (CI), 85.37%, 94.26%) and a PPA of 94.32% (95% CI, 87.24%, 98.13%). Co-infections were detected by QIAstat-SARS-CoV-2 in 4/277 specimens. The methods exhibited comparable failure rates (23/307 [7.5%] vs. 6/298 [2.0%] for QIAstat-SARS-CoV-2 and reference methods, respectively). The turnaround time was shorter for QIAstat-SARS-CoV-2 compared with the reference method (difference in mean -14:30 h [standard error, 0:03:23; 95% CI, -14:37, -14:24]; P < 0.001).
CONCLUSIONS: QIAstat-SARS-CoV-2 shows good agreement with the reference assay, providing faster and accurate results for detecting SARS-CoV-2.
Copyright © 2021 The Authors. Published by Elsevier Ltd.. All rights reserved.

Entities:  

Keywords:  COVID-19; Diagnostic testing; Multiplex; Real time-PCR; SARS-CoV-2

Year:  2021        PMID: 33901651      PMCID: PMC8064816          DOI: 10.1016/j.ijid.2021.04.066

Source DB:  PubMed          Journal:  Int J Infect Dis        ISSN: 1201-9712            Impact factor:   3.623


Introduction

Early identification of patients with severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infection enables rapid isolation to prevent transmission (European Centre for Disease Prevention and Control, 2020, National Institutes of Health, 2020, The World Health Organization, 2020). Symptoms of coronavirus disease 2019 (COVID-19) are common to a range of respiratory pathogens; therefore, accurate diagnostics are required for differential diagnosis. The QIAstat-Dx® Respiratory SARS-CoV-2 Panel (QIAstat-SARS-CoV-2) is a closed, fully automated, multiplex assay that detects SARS-CoV-2 and 21 other respiratory pathogens. Prior independent validation studies of the SARS-CoV-2 assay demonstrated comparable performance with a WHO-recommended reverse-transcription polymerase chain reaction (RT-PCR) assay (Visseaux et al., 2020). The remaining 21 targets in the panel have been previously validated (Boers et al., 2020). This study aimed to evaluate the SARS-CoV-2 assay performance characteristics in QIAstat-SARS-CoV-2 against the WHO-recommended reference method (WHO-Charité) (Corman et al., 2020).

Methods

This was an observational, retrospective study of specimens from patients with or suspected of having COVID-19. The primary objective was to compare the performance of QIAstat-SARS-CoV-2 with the WHO-Charité reference method (Corman et al., 2020), which was the standard-of-care test at the investigation site (acceptance criteria: positive percent agreement [PPA] and negative percent agreement [NPA] ≥90%). Secondary objectives were to evaluate failure rates and differences between methods in mean cycle threshold (Ct) values and time-to-results. Collection of nasopharyngeal swab specimens and analysis was performed at Bichat–Claude Bernard Hospital, Paris, France. De-identified, residual samples were tested, and only transport medium liquid samples were included. Further methodological details are provided in the Supplementary data. The study was conducted following the ethical principles of the Declaration of Helsinki and regulations regarding Good Clinical Practice.

Results

The analysis took place between March 23 and June 04, 2020. 293 specimens were collected, and 16 samples were excluded from the analysis (Figure 1 ). The remaining 277 samples were tested using both QIAstat-SARS-CoV-2 and the reference method. An additional 25 results were excluded based on a failed QIAstat-SARS-CoV-2 (n = 13) or reference method test (n = 12; Figure 1). Patient age at hospital admission was available for 237 subjects (mean 58.2 years [range, 1–97 years; standard deviation, 20.24 years]).
Figure 1

Study flow diagram.

Study flow diagram. In total, 18/189 samples that were negative according to the reference method were positive using QIAstat-SARS-CoV-2 (NPA: 90.48% [95% confidence interval (CI), 85.37%, 94.26%]; Table 1 ). According to the reference method, five of 88 samples that were positive were negative according to QIAstat-SARS-CoV-2 (PPA: 94.32% [95% CI, 87.24%, 98.13%]). These 23 discordant results were investigated by reviewing patient medical source data. An alternative RT-PCR (Cobas or Altona) method on different samples obtained during the patients’ follow-up confirmed one of five QIAstat-SARS-CoV-2-negative reference assay-positive results to be positive. Seven of 18 QIAstat-SARS-CoV-2-positive reference assay-negative results were confirmed as positive. Both methods exhibited comparable failure rates; 23/307 (7.5%) QIAstat-SARS-CoV-2 tests failed compared with 6/298 (2.0%) reference method assays.
Table 1

Performance of QIAstat-Dx® Respiratory SARS-CoV-2 Panel system in comparison with the reference method.

QiaStat-Dx®Reference method
PositiveNegativeTotal
Positive8318101
Negative5171176
Total88189277
PPA, % (95% CI)94.32 (87.24, 98.13)
NPA, % (95% CI)90.48 (85.37, 94.26)

CI, confidence interval; NPA, negative percent agreement; PPA, positive percent agreement; SARS-CoV-2; severe acute respiratory syndrome coronavirus 2.

Performance of QIAstat-Dx® Respiratory SARS-CoV-2 Panel system in comparison with the reference method. CI, confidence interval; NPA, negative percent agreement; PPA, positive percent agreement; SARS-CoV-2; severe acute respiratory syndrome coronavirus 2. The mean Ct value for QIAstat-SARS-CoV-2 was significantly lower than for each component of the reference method (difference in mean: −6.747 [standard error (SE), 0.329; 95% CI, –7.398, –6.097]; P < 0.001 and –4.979 [SE, 0.271; 95% CI, –5.516, –4.443]; P < 0.001, for RNA-dependent RNA polymerase- and E gene-components, respectively). When plotting individual Ct values obtained with QIAstat-SARS-CoV-2 against those for each component of the reference method, a proportional shift was observed over the whole range of experimental Ct values (Supplementary Figure 1). Of the 277 QIAstat-SARS-CoV-2 results analyzed, four specimens were positive for multiple pathogens, including two co-infections of SARS-CoV-2 and another respiratory pathogen (Supplementary Table 1). The mean time to result was significantly shorter for QIAstat-SARS-CoV-2 than the reference method (Δ –14:30 h [SE, 03:23; 95% CI, –14:37, –14:24]; P < 0.001).

Discussion

The QIAstat-SARS-CoV-2 panel demonstrated PPA and NPA with the WHO-recommended assay greater than 90%, both in this study and in another previous smaller study (Visseaux et al., 2020). Rapid testing for SARS-CoV-2 enables quick triage and the identification of patients requiring isolation (Brendish et al., 2020). In this study, QIAstat-SARS-CoV-2 was 14.5 h quicker than the reference method; this rapid testing with QIAstat-SARS-CoV-2 was consistent with previous results (Brendish et al., 2020, Visseaux et al., 2020). Near-patient multiplex tests allow testing multiple pathogens in a single assay, simplifying testing workflows and assisting in the timely differential diagnosis of infectious diseases. The utility of near-patient multiplex testing using the QIAstat-Dx® Respiratory Panel to reduce time to diagnosis and improve patient management has previously been demonstrated (Bouzid et al., 2020). In this study, co-infections were identified in four samples, including two samples positive for SARS-CoV-2. Co-infections with SARS-CoV-2, at rates in-line with this study, have previously been reported in the literature (Lai et al., 2020, Lansbury et al., 2020). In conclusion, QIAstat-SARS-CoV-2 produces concordant results with the WHO-Charité reference method, but in a significantly shorter time and in a near-patient setting.

Funding

This study has been funded by QIAGEN Manchester Ltd., Manchester, UK.

Conflict of interest

BG is a contractor for QIAGEN. MCC, AE, JP, JL, and DM are employees of QIAGEN. DD has received personal fees from Viiv Healthcare, , and Janssen Cilag. BV has received grants from QIAGEN, personal fees from QIAGEN, BioMérieux, Hologic and , and non-financial support from QIAGEN and BioMérieux. SL, AS, QH, GAT, and NHF do not declare any conflicts of interest.
  3 in total

1.  Evaluation of the QIAstat-Dx Respiratory SARS-CoV-2 panel, a rapid multiplex PCR method for the diagnosis of COVID-19.

Authors:  Masahiro Ishikane; Hiroyuki Unoki-Kubota; Ataru Moriya; Satoshi Kutsuna; Honami Ando; Yasushi Kaburagi; Tetsuya Suzuki; Noriko Iwamoto; Moto Kimura; Norio Ohmagari
Journal:  J Infect Chemother       Date:  2022-02-09       Impact factor: 2.065

2.  Evaluation of the QIAstat-Dx RP2.0 and the BioFire FilmArray RP2.1 for the Rapid Detection of Respiratory Pathogens Including SARS-CoV-2.

Authors:  Hayley Cassidy; Mart van Genne; Erley Lizarazo-Forero; Hubert G M Niesters; Lilli Gard
Journal:  Front Microbiol       Date:  2022-03-24       Impact factor: 5.640

3.  Pathogen Profile of Children Hospitalised with Severe Acute Respiratory Infections during COVID-19 Pandemic in the Free State Province, South Africa.

Authors:  Ayodeji E Ogunbayo; Milton T Mogotsi; Hlengiwe Sondlane; Kelebogile R Nkwadipo; Saheed Sabiu; Martin M Nyaga
Journal:  Int J Environ Res Public Health       Date:  2022-08-21       Impact factor: 4.614

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.