Daniel Geisler1, Megan Culler Freeman2, Glenn J Rapsinski2,3, Sarah E Wheeler4.
Abstract
OBJECTIVES: Serologic assay performance studies for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in pediatric populations are lacking, and few seroprevalence studies have routinely incorporated orthogonal testing to improve accuracy.Entities:
Keywords: Orthogonal testing; Pediatric population; SARS-CoV-2; Seroprevalence
Mesh:
Substances:
Year: 2021 PMID: 33899091 PMCID: PMC8130881 DOI: 10.1093/ajcp/aqab033
Source DB: PubMed Journal: Am J Clin Pathol ISSN: 0002-9173 Impact factor: 2.493
Orthogonal Testing of Pediatric Samples
| First Assay | Second Assay | Third Assay | |||
|---|---|---|---|---|---|
| Pos (Neg) | Assay | Pos (Neg) | Assay | Pos (Neg) | Assay |
| 20 (0) | EuroIGG | 3 (17) | Beckman | 2 (18) | Siemens-C |
| 9 (0) | EuroIGG | 0 (9) | Beckman | 0 (9) | Siemens-V |
| 44 (0) | EuroIGG | 11 (33) | Beckman | — | — |
| Total: 73 | 14 (59) | 2 (27) | |||
Beckman, Beckman Coulter Access SARS-CoV-2 IgG test on the AU5800 analyzer; ELISA, enzyme-linked immunosorbent assay; EuroIGG, EUROIMMUN Anti-SARS-CoV-2 ELISA IgG assay; Neg, negative; Pos, positive; Siemens-C, ADVIA Centaur XP SARS-CoV-2 Total Antibody assay; Siemens-V, Dimension Vista 1500 SARS-CoV-2 Total Antibody assay.