| Literature DB >> 33898687 |
Rosanna Inzitari1, Sebastian Vencken1, David Daghfal2, Karl McAuley1, Anthony McDermott1, Marie Galligan1, Peter Doran1.
Abstract
BACKGROUND: Troponin is a widely used cardiac protein biomarker for acute coronary syndrome. Its increasing importance drives an increasing need to assess, in real-world conditions, the performance of the tests to measure it. We evaluated the performance characteristics of high-sensitivity troponin I assay reagents and ancillary agents on the Abbott ARCHITECT ci4100, ARCHITECT i2000SR and Alinity ci using historical quality control data spanning 5 years. DESIGN ANDEntities:
Keywords: Assay; Cardiovascular disease; Diagnostics; Immunoassay; Troponin
Year: 2021 PMID: 33898687 PMCID: PMC8058562 DOI: 10.1016/j.plabm.2021.e00217
Source DB: PubMed Journal: Pract Lab Med ISSN: 2352-5517
Descriptive statistics for hs-TnI reagent lots and control lots on ARCHITECT ci4100, ARCHITECT i2000SR and Alinity ci.
| ARCHITECT ci4100 | |||
|---|---|---|---|
| Statistic | Low (20 ng/L) | Medium (200 ng/L) | High (15,000 ng/L) |
| NReagent | 11 | 11 | 11 |
| NControl | 7 | 7 | 7 |
| NTotal | 248 | 248 | 248 |
| Mean | 20.80 | 195.0 | 15,528 |
| BiasTarget | 3.98 | −2.49 | 3.52 |
| BiasFRT | 2.60 | 1.14 | −0.08 |
| CVIntrareagents | 5.25 | 3.84 | 4.35 |
| CVInterreagents | 7.16 | 5.83 | 4.25 |
| CVIntracontrol | 6.88 | 5.30 | 5.11 |
| CVIntercontrol | 3.17 | 3.63 | 1.95 |
| CVTotal | 7.53 | 6.37 | 5.28 |
| NReagent | 16 | 16 | 16 |
| NControl | 6 | 6 | 6 |
| NTotal | 394 | 395 | 395 |
| Mean | 19.61 | 196.6 | 15,150 |
| BiasTarget | −1.95 | −1.72 | 1.00 |
| BiasFRT | 0.37 | −0.81 | −2.03 |
| CVIntrareagents | 5.45 | 4.09 | 3.09 |
| CVInterreagents | 6.50 | 3.86 | 5.80 |
| CVIntracontrol | 7.13 | 5.25 | 5.47 |
| CVIntercontrol | 6.19 | 2.73 | 5.67 |
| CVTotal | 8.80 | 5.79 | 6.25 |
| NReagent | 3 | 3 | 3 |
| NControl | 3 | 3 | 3 |
| NTotal | 107 | 108 | 108 |
| Mean | 19.87 | 198.7 | 15,383 |
| BiasTarget | −0.64 | −0.67 | 2.55 |
| BiasFRT | −4.37 | −0.94 | 2.37 |
| CVIntrareagents | 5.44 | 4.59 | 4.02 |
| CVInterreagents | 3.24 | 3.34 | 4.22 |
| CVIntracontrol | 5.33 | 4.55 | 3.94 |
| CVIntercontrol | 3.61 | 3.36 | 4.27 |
| CVTotal | 5.91 | 5.39 | 5.26 |
Fig. 1Lot-specific mean hs-TnI concentrations over lot manufacturing date on the ARCHITECT ci4100. Points indicate lot-specific means Each point is a unique lot. Error bars represent lot-specific standard deviations. Horizontal dashed line represents the ideal target hs-TnI concentration.
Fig. 2Lot-specific mean hs-TnI concentrations over lot manufacturing date on the ARCHITECT i2000SR. Points indicate lot-specific means. Error bars represent lot-specific standard deviations. Horizontal dashed line represents the target-release hs-TnI concentration.
Fig. 3hs-TnI concentration at three levels over the age of the controls and reagents on the ARCHITECT ci4000. Regression lines represent marginal mean hs-TnI concentration against the margin of control age (left column) or reagent age (right column). Shaded bands represent 95% confidence intervals. Coloration of data points represent individual reagent lots.
Fig. 4hs-TnI concentration at three levels over the age of the controls and reagents on the ARCHITECT i2000SR. Regression lines represent marginal mean hs-TnI concentration against the margin of control age (left column) or reagent age (right column). Shaded bands represent 95% confidence intervals. Coloration of data points represent individual lots.
Fig. 5hs-TnI concentration at three levels over the time from last calibration on the ARCHITECT ci4100 and ARCHITECT i2000SR. Regression lines represent marginal mean hs-TnI concentration against the margin of time from last calibration. Shaded bands represent 95% confidence intervals. Coloration of data points represent individual reagent lots.