| Literature DB >> 33880407 |
Marco Antônio de Oliveira Monteiro1, Alberto Nogueira da Gama Antunes2, Roberta Tarkany Basting1.
Abstract
BACKGROUND: In anesthetic techniques, touching bones can cause needle bending. Theoretically, a needle should support such deflection without fracturing. However, it is possible that a needle may fracture depending on the quality and type of needle used. This study evaluated the physical, chemical, and micromorphological characteristics of long and short dental anesthetic needles, as well as the mechanical properties of flexural load and bending resistance when needles are subjected to different bending angles.Entities:
Keywords: Dental Anesthesia; Flexural Strength; Local Anesthesia; Needles
Year: 2021 PMID: 33880407 PMCID: PMC8039161 DOI: 10.17245/jdapm.2021.21.2.139
Source DB: PubMed Journal: J Dent Anesth Pain Med ISSN: 2383-9309
Specifications of dental needles used in this study according to their manufacturers
| Trademark | Country | Gauge | Type / Classification | Length (in millimeters) | Diameter (in millimeters) | Lot |
|---|---|---|---|---|---|---|
| Dentojet | Italy | 27G | Long | 38 | 0.4 | 17080578 |
| Injex | Brazil | 27G | Long | 30 | 0.4 | 515/18 |
| Jets | Korea | 30G | Extra short | 12 | 0.3 | 3091098 |
| Jets | Korea | 27G | Long | 30 | 0.4 | 3091098 |
| Misawa | Japan | 30G | Extra short | 12 | (*) | 170120 |
| Misawa | Japan | 27G | Long | 42 | (*) | 130705 |
| Procare | China | 27G | Long | (*) | (*) | 280817L |
| Selekto | Germany | 30G | Short | 21 | 0.3 | 8H170403 |
| Septoject XL | France | 27G | Long | 30 | 0.4 | F05202AA |
| Terumo | Japan | 30G | (*) | 21 | 0.3 | 171226 |
| Terumo | Japan | 27G | (*) | 35 | 0.4 | 151015 |
| Unoject | Japan | 30G | Short | 25 | 0.3 | 150317 |
(*) Manufacturer does not indicate information on packaging.
Fig. 1Needle measurement locations. C1, anterior length; C2, posterior length; D1, diameter closest to the hub (cannon); D2, central diameter of the anterior part; D3, diameter closest to the bevel; DP, posterior diameter.
Fig. 2NFMD. (A) Carpule syringe positioned and locked in the NFMD. The chisel coupled to the testing machine is positioned to induce the first bend in the gingival needle. (B) Flexion of the needle with a 60° inclination. (C) After bending, the syringe is rotated by 180° such that the needle is facing upward and the chisel positioned to induce the second bend of 60°.
Average anterior and posterior lengths, and average diameter of extrashort, short and long dental needles
| Variable | Needle type | Needle trademark | Mean | (standard deviation) |
|---|---|---|---|---|
| Anterior needle length (C1) | Extra short | Jets 30G | 11.28 (0.18) a | |
| Misawa 30G | 10.99 (0.25) b | |||
| P-value | 1P = 0.0083 | |||
| Short | Selekto 30G | 21.23 (0.64) b | ||
| Terumo 30G | 20.36 (0.26) c | |||
| Unoject 30G | 24.31 (0.12) a | |||
| P-value | 2P < 0.0001 | |||
| Long | Dencojet 27G | 37.83 (0.30) b | ||
| Injex 27G | 28.78 (0.60) f | |||
| Jets 27G | 29.44 (0.23) e | |||
| Misawa 27G | 42.44 (0.18) a | |||
| Procare 27G | 35.94 (0.41) c | |||
| Septoject XL 27G | 29.33 (0.23) e | |||
| Terumo 27G | 35.16 (0.22) d | |||
| P-value | 2P < 0.0001 | |||
| Posterior needle length (C2) | Extra short | Jets 30G | 11.58 (0.13) a | |
| Misawa 30G | 11.01 (0.10) b | |||
| P-value | 1P < 0.0001 | |||
| Short | Selekto 30G | 11.37 (0.22) a | ||
| Terumo 30G | 10.53 (0.20) b | |||
| Unoject 30G | 9.53 (3.24) b | |||
| P-valor | 2P < 0.0001 | |||
| Long | Dencojet | 10.92 (0.21) d | ||
| Injex 27G | 11.39 (0.44) c | |||
| Jets 27G | 11.39 (0.15) c | |||
| Misawa 27G | 10.44 (0.20) e | |||
| Procare 27G | 12.08 (0.42) b | |||
| Septoject XL 27G | 13.29 (0.15) a | |||
| Terumo 27G | 10.54 (0.14) e | |||
| P-value | 2P < 0.0001 | |||
| Average needle diameter | Extra short | Jets 30G | 0.31 (0.01) a | |
| Misawa 30G | 0.30 (0.01) a | |||
| P-value | 1P = 0.2163 | |||
| Short | Selekto 30G | 0.30 (0.01) b | ||
| Terumo 30G | 0.31 (0.00) a | |||
| Unoject 30G | 0.31 (0.00) a | |||
| P-value | 3P = 0.0013 | |||
| Long | Dencojet 27G | 0.40 (0.01) ab | ||
| Injex 27G | 0.39 (0.01) c | |||
| Jets 27G | 0.40 (0.01) ab | |||
| Misawa 27G | 0.40 (0.01) ab | |||
| Procare 27G | 0.41 (0.01) a | |||
| Septoject XL 27G | 0.40 (0.01) ab | |||
| Terumo 27G | 0.40 (0.00) bc | |||
| P-value | 3P = 0.0232 |
Different letters within each classification represent statistically significant differences (P ≤ 0.05). 1T-test, 2one-way ANOVA, and 3generalized linear models.
Fig. 3Micromorphologies of the needles. (A) Needle bevel with 60-fold magnification. (B) Tip of the needle with 430-fold magnification. (C) Needle fracture site with 200-fold magnification.
Percentage of chemical element each needle according to EDS analysis
| Needle trademark | Chemical element | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Fe | C | Cr | Ni | O | Si | Mn | Al | Ca | |
| Dencojet 27G | 51 | 15.8 | 14.6 | 6.6 | 5.3 | 5.1 | 1.3 | 0.2 | - |
| Injex 27G | 55.8 | 12.2 | 15.2 | 6.3 | 3.6 | 4.3 | 1 | 1.5 | 0.2 |
| Jets 30G | 58.1 | 16.4 | 8.9 | 7.1 | 2.9 | 2.9 | 1.5 | 1.7 | 0.4 |
| Jets 27G | 46.1 | 16.7 | 13.3 | 5.5 | 5.7 | 11.2 | 1.2 | 0.3 | - |
| Misawa 30G | 65 | 4.7 | 18.1 | 8.3 | - | 1.6 | 1.6 | 0.7 | - |
| Misawa 27G | 66.6 | 2.8 | 18.5 | 8.6 | - | 1.1 | 1.7 | 0.7 | - |
| Procare 27G | 61.3 | 8.3 | 17 | 7 | 2 | 2.4 | 1.1 | 1 | - |
| Selekto 30G | 63.8 | 5.1 | 17.8 | 8 | 1.6 | 1.5 | 1.6 | 0.6 | - |
| Septoject XL 27G | 63.5 | 5.5 | 18 | 8.5 | - | 1.5 | 1.7 | 1.2 | - |
| Terumo 27G | 64.7 | 4.6 | 18.6 | 8.1 | - | 1.4 | 1.7 | 1 | - |
| Terumo 30G | 57 | 11 | 16 | 7 | 2.7 | 3.8 | 1.5 | 1 | - |
| Unoject 30G | 67.2 | 1.9 | 18.5 | 8.2 | - | 0.9 | 1.8 | 1.4 | - |
Median (minimum; maximum) loads (in Newtons) for folding anesthetic needles (27G and 30G) from different manufacturers according to angles over two bends
| Flexion | Needle trademark | Needle flexion angle | ||
|---|---|---|---|---|
| 30° | 60° | 90° | ||
| First | Jets 30G | 3.51 (2.69; 3.69) Bf | 4.39 (4.01; 4.94) Aef | 4.27 (3.65; 5.06) Af |
| Misawa 30G | 8.02 (5.20; 8.30) Ac | 7.27 (6.71; 7.47) Ac | 7.17 (5.96; 9.93) Ad | |
| Selekto 30 G | 6.57 (4.23; 6.75) Ad | 5.38 (3.24; 6.04) Ae | 4.33 (3.45; 4.59) Bf | |
| Terumo 30G | 5.44 (3.40; 6.51) Bd | 6.43 (5.40; 6.59) Bd | 7.64 (6.33; 8.48) Ad | |
| Unoject 30G | 4.05 (3.20; 4.25) Ae | 4.03 (3.69; 4.73) Af | 4.05 (3.65; 5.66) Af | |
| Dencojet 27G | 8.82 (6.75; 9.97) Bbc | 10.46 (7.39; 13.48) ABb | 11.38 (9.25; 13.48) Ab | |
| Injex 27G | 9.45 (7.64; 10.50) Ab | 8.00 (5.60; 11.50) ABbc | 7.72 (6.53; 8.10) Bd | |
| Jets 27G | 6.08 (5.22; 6.81) Ad | 6.04 (5.60; 7.43) Ad | 5.56 (4.98; 6.39) Ae | |
| Misawa 27G | 8.16 (7.80; 8.60) Ac | 9.09 (7.37; 9.87) Ab | 6.25 (5.90; 8.22) Bd | |
| Procare 27G | 13.72 (10.81; 14.51) Ba | 14.75 (12.79; 19.66) Aa | 14.49 (13.32; 15.80) ABa | |
| Septoject XL 27G | 11.16 (9.01; 13.66) Aa | 9.61 (6.33; 13.22) Ab | 6.99 (5.72; 14.51) Ad | |
| Terumo 27G | 9.23 (4.86; 11.42) Abc | 8.58 (7.51; 11.12) Ab | 8.10 (7.62; 8.93) Ac | |
| Second | Jets 30G | 3.63 (3.00; 3.95) Bd | 4.63 (3.97; 5.06) Afg | *4.73 (3.95; 5.20) Af |
| Misawa 30G | *7.64 (5.18; 8.20) Ac | *6.99 (6.73; 7.33) Ad | *7.05 (5.84; 9.05) Ad | |
| Selekto 30 G | 6.14 (4.61; 7.86) Ac | 5.52 (3.26; 5.76) ABf | 4.45 (3.65; 4.84) Bf | |
| Terumo 30G | 6.02 (5.06; 7.86) Bc | 6.53 (5.46; 6.57) Be | 7.72 (6.49; 7.96) Ad | |
| Unoject 30G | *3.45 (3.04; 3.77) Cd | 4.35 (4.15; 4.53) Bg | *5.14 (5.00; 6.00) Ae | |
| Dencojet 27G | 8.86 (7.21; 10.11) Bb | 13.34 (8.84; 15.92) Aab | *12.67 (11.40; 14.51) Ab | |
| Injex 27G | 10.40 (8.24; 10.96) Bb | *11.65 (7.37; 14.45) ABbc | *11.87 (9.73; 13.16) Abc | |
| Jets 27G | 6.99 (5.78; 7.74) Bc | *8.30 (5.98; 12.99) Ac | *6.97 (6.29; 9.33) ABd | |
| Misawa 27G | *10.07 (7.64; 10.60) Ab | *9.83 (8.62; 13.40) Abc | *10.60 (6.21; 11.71) Ac | |
| Procare 27G | 13.66 (12.49; 14.20) Ba | *14.12 (12.91; 18.64) Aa | *14.16 (13.14; 15.31) ABa | |
| Septoject XL 27G | *11.77 (9.65; 15.76) Aa | 11.44 (7.33; 13.10) Abc | *11.95 (7.60; 15.47) Abc | |
| Terumo 27G | 8.72 (8.14; 10.98) Bb | *10.34 (9.13; 12.93) Abc | *8.72 (8.32; 9.45) Bc | |
(*) Significant difference compared to the first flexion for the same manufacturer and angle (P ≤ 0.05). Different letters (uppercase horizontally and lowercase vertically when comparing brands for each angle) indicate statistically significant differences (P ≤ 0.05). P (angle) = 0.0010; P (brand) < 0.0001; P (angle × brand) = 0.0018; P (flexion) < 0.0001; P (angle × flexion) < 0.0016; P (brand × flexion) < 0.0001; P (angle × brand × flexion) = 0.0092.
Median (minimum; maximum) numbers of bends before the fracturing of anesthetic needles (27G and 30G) from different manufacturers under different bending angles
| Needle trademark | Number of fracturing | ||
|---|---|---|---|
| 30° | 60° | 90° | |
| Jets 30G | - | 7.00 (7.00; 9.00) Aab | 8.00 (7.00; 9.00) Aa |
| Misawa 30G | - | 5.00 (5.00; 5.00) Aef | 5.00 (4.00; 5.00) Acd |
| Selekto 30 G | - | 6.00 (6.00; 9.00) Aabc | 5.00 (4.00; 9.00) Abc |
| Terumo 30G | - | 6.00 (6.00; 8.00) Abcd | 5.00 (4.00; 5.00) Bde |
| Unoject 30G | - | 8.00 (7.00; 9.00) Aa | 6.00 (4.00; 6.00) Bbc |
| Dencojet 27G | - | 6.00 (5.00; 10.00) Aabc | 4.00 (4.00; 5.00) Bde |
| Injex 27G | - | 6.00 (5.00; 9.00) Abcd | 5.00 (5.00; 5.00) Bde |
| Jets 27G | - | 7.00 (6.00; 9.00) Aab | 6.00 (5.00; 9.00) Aab |
| Misawa 27G | - | 6.00 (5.00; 6.00) Acde | 5.00 (5.00; 7.00) Abc |
| Procare 27G | - | 5.00 (5.00; 6.00) Ade | 4.00 (4.00; 4.00) Be |
| Septoject XL 27G | - | 4.00 (3.00; 5.00) Af | 4.00 (3.00; 5.00) Ae |
| Terumo 27G | - | 6.00 (5.00; 7.00) Abcd | 5.00 (5.00; 5.00) Bde |
At a 30° angle, no needles were broken. Groups with medians followed by different letters (uppercase horizontally and lowercase vertically) differ significantly from each other (P ≤ 0.05). P (brand) < 0.000; P (angle) = 0.000; P (brand × angle) = 0.0079.