| Literature DB >> 33878981 |
Benjamin Mine1, Thomas Bonnet1, Adrien Guenego1, Stéphanie Elens1, Juan V Suarez1, Boris Lubicz1.
Abstract
We report a case of cerebral aneurysm rerupture 24 months after initial bleeding and complete occlusion using a WEB device. A middle-aged patient was transferred at our institution because of a recurrent aneurysmal bleeding. The patient was graded WFNS2 and unenhanced CT showed a modified Fisher grade 3 SAH. DSA showed an increased compaction and an evolutive "proximal recess" of the WEB device associated with a small unprotected part of the aneurysm wall and a bleb pointing towards the haemorrhage. Balloon- and stent-assisted coiling was performed, allowing to completely occlude the recurrence. The patient suffered no neurological deficit neither delayed complication and was discharged at day 16.Entities:
Keywords: Intracranial aneurysm; subarachnoid hemorrhage; woven EndoBridge
Mesh:
Year: 2021 PMID: 33878981 PMCID: PMC8673907 DOI: 10.1177/15910199211011879
Source DB: PubMed Journal: Interv Neuroradiol ISSN: 1591-0199 Impact factor: 1.610