Literature DB >> 33876847

Effective design of barrier enclosure to contain aerosol emissions from COVID-19 patients.

Dan Daniel1, Marcus Lin2, Irvan Luhung3, Tony Lui4, Anton Sadovoy1, Xueqi Koh1, Anqi Sng1, Tuan Tran2, Stephan C Schuster3, Xian Jun Loh1, Oo Schwe Thet5, Chee Keat Tan4.   

Abstract

Facing shortages of personal protective equipment, some clinicians have advocated the use of barrier enclosures (typically mounted over the head, with and without suction) to contain aerosol emissions from coronavirus disease 2019 (COVID-19) patients. There is, however, little evidence for its usefulness. To test the effectiveness of such a device, we built a manikin that can expire micron-sized aerosols at flow rates close to physiological conditions. We then placed the manikin inside the enclosure and used a laser sheet to visualize the aerosol leaking out. We show that with sufficient suction, it is possible to effectively contain aerosol from the manikin, reducing aerosol exposure outside the enclosure by 99%. In contrast, a passive barrier without suction only reduces aerosol exposure by 60%.
© 2021 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  COVID-19; aerosol; airborne transmissions; barrier enclosure; infection control

Mesh:

Year:  2021        PMID: 33876847      PMCID: PMC8250690          DOI: 10.1111/ina.12828

Source DB:  PubMed          Journal:  Indoor Air        ISSN: 0905-6947            Impact factor:   6.554


Barrier enclosure with active suction can be used to minimize aerosol spread from COVID‐19 patients. Barrier can be made from cheap material such as acrylic and suction can be provided by wall units commonly found in hospitals, that is, this approach can potentially be scaled up. Given the growing evidence that COVID‐19 is airborne, a well‐designed barrier enclosure can provide additional protection for healthcare workers.

INTRODUCTION

The prolonged nature of the coronavirus disease 2019 (COVID‐19) pandemic has resulted in global shortages of personal protective equipment (PPE), especially N95 respirators. As a result, some clinicians have resorted to using barrier enclosures typically mounted on the hospital bed over the patient's head to contain any aerosol emissions, especially during aerosol generating procedures such as intubation. , , Different barrier designs with varying levels of complexities have been proposed, ranging from a simple carton box, plastic drapes and boxes , , to a custom‐built acrylic box with active suction and filtration system. Despite all these innovations, there is little validation on the effectiveness of the various designs. To assess the performance of barrier enclosures, some groups have resorted to simulating a cough/sneeze (for example by using an exploding balloon or a water spray ) and comparing the splatter pattern of droplets with and without the enclosure. Such an approach can confirm the effectiveness of the enclosures at blocking larger respiratory droplets (tens of microns to millimeters in diameters), but not for small micron‐sized aerosol droplets which are airborne and can potentially travel over much longer distances. To detect aerosol spread, some groups have successfully used particle counters to measure the concentration of micron and sub‐micron droplets leaking out of the barrier enclosure , and concluded that active suction is critical to effectively contain aerosol droplets. However, the amount of suction required to achieve containment remains unexplored. This is crucial because COVID‐19 patients are often subjected to treatment modalities involving high gas flow rates, for example, high flow nasal cannula (HFNC) therapy to provide supplemental oxygen, , , which can quickly disperse bioaerosol over large distances, increasing the infection risk to healthcare workers. In this paper, we would like to establish the design criteria—in particular, the minimum suction required—for a barrier enclosure to effectively contain aerosol emissions especially with high (and medically relevant) gas flow rates. We used two complementary techniques (one qualitative and the other quantitative) to assess the performance of the barrier enclosure. The first is to use laser sheets to visualize the aerosol flow from a custom‐built manikin expiring micron‐sized water‐glycerin droplets at flow rates close to physiological conditions. The second is to collect the aerosol leaking out using air samplers and subsequently to quantify the collected amount using spectrofluorometry. Given the growing evidence that COVID‐19 is airborne and can spread through aerosol, , , a well‐designed barrier enclosure can be useful as an additional layer of protection for healthcare workers.

MATERIALS AND METHODS

Barrier enclosure design

The barrier enclosure was designed by local institutions of higher learning and used by a hospital in Singapore. The physical enclosure (width, breadth, and height of 92 cm × 60 cm × 70 cm) is made of acrylic (Figure 1A), with a plastic drape at the front, four openings (two at the headend and one on either side) for access to patients, and two suction ports (one on either side). Suction is provided by two wall units (BeaconMedaes Gem 10) typically found in a hospital, each with a maximum suction rate of Q suction = 60 L min−1.
FIGURE 1

A, Barrier enclosure design used to contain aerosols from (B) patients who typically require supplementary oxygen treatment (C) delivered through a nasal cannula

A, Barrier enclosure design used to contain aerosols from (B) patients who typically require supplementary oxygen treatment (C) delivered through a nasal cannula COVID‐19 patients typically require supplemental oxygen therapy, with high flow nasal cannula (HFNC) oxygen therapy (Figure 1b,c) showing positive medical outcomes. , , HFNC is a heated and humidified system (typicaly 100% humidity at 34–37°C) that delivers oxygen passively at a prescribed concentration (between 21% and 100%) at high flow rates up to Q O2 = 60 L min−1. In this study, we looked at the effectiveness of aerosol containment for diferent Q O2 = 0–60 L min−1 (with oxygen concentration set at 75%) and Q suction = 0–120 L min−1. The HFNC was placed over the nose of a custom‐built manikin which can emits aerosol at flow rates close to physiological conditions as described in the next section.

Manikin design and aerosol generation

Figure 2A shows a schematic of the custom‐built manikin. A 400 W fog machine, typically used in entertainment venues, was used to heat up and generate micron‐sized aerosol from a glycerin‐water mixture (50 v%) inside a 50 cm × 50 cm × 60 cm box within 3 seconds. Food‐grade glycerin was purchased from a baking shop. The addition of glycerin which has a boiling point of 290°C minimizes the evaporation of the aerosol droplets.
FIGURE 2

A, Schematic of the custom‐built manikin placed inside the barrier enclosure. B, Micron‐sized glycerin‐water aerosol droplets expelled from the manikin and visualized by shining a blue laser sheet at the sagittal plane (See also Video S1). C, Size distribution of aerosol droplets

A, Schematic of the custom‐built manikin placed inside the barrier enclosure. B, Micron‐sized glycerin‐water aerosol droplets expelled from the manikin and visualized by shining a blue laser sheet at the sagittal plane (See also Video S1). C, Size distribution of aerosol droplets A manual resuscitator or an Ambu bag was then used to transfer the aerosol from the box through the mouth of the manikin into the barrier. The Ambu bag was compressed by hand and the volume pushed out during each compression cycle (0.63 ± 0.03 L) was determined by releasing the expelled air into a jar filled with water and measuring the volume of water displaced. The length of the compression cycle (3 s) was chosen to closely mimic the physiological parameters of the human breath, that is, Q air = 13 L min−1. We did not find significant difference in the expelled volumes between different human operators. In a typical experiment, the Ambu bag was compressed for 3 seconds and re‐inflated for another 3 seconds to mimic continuous exhalation of aerosol for 20 minutes (Figure 2B and Video S1). A metronome was used to help the operator keep in time. The manikin does not simulate the inhalation cycle, but this should not affect the results greatly, since aerosol is generated mostly during the exhalation cycle. Aerosol typically refers to droplets <5 μm in diameter and thought to be responsible for airborne transmission of diseases, since they can stay suspended in air for a long period of time. , Aerosol generated during respiratory activities (such as talking and coughing) and in healthcare settings (such as intubation) is thought to vary between 0.1 and 10 μm. , , The droplets generated by the fog machine span this range with a mean diameter of about 1 μm, as measured using a particulate matter sensor Sensirion SPS30 (Figure 2C).

Visualizing aerosol flow using laser sheets

The aerosol flow can be visualized by shining 2D laser sheets. The laser sheet is generated by passing a laser light (Class 3B, 100 mW power) through a 4‐mm‐diameter cylindrical rod lens (Edmund Optics). Similar techniques have been used to visualize aerosol leak from masks and faceshields.

Quantifying aerosol leak

We first dissolved a small amount of fluorescein sodium salt (Sigma Aldrich) at a concentration of 0.5 g L−1 in the water‐glycerin solution. The aerosol (glycerol‐water droplets with fluorescein) leaking out of the barrier enclosure was collected using SASS 3100 Air Samplers fitted with standard filter cartridges from Research International running at a flow rate of 60 L min−1 for 20 minutes. The hospital room we were using had a total air change rate per hour of 10, that is, it took about 6 minutes for the air in the room to be well‐mixed, shorter than the experimental time. SASS 3100 air sampler has been used previously to collect bioaerosols from the air. The filter paper was then placed in a tube of 2 mL of deionized (DI) water and shaken with a vortex mixer for 1 minute to dissolve the trapped fluorescein. Fluorescein has an excitation and emission wavelengths of 490 and 514 nm, respectively. The fluorescein concentration (and hence the amount of fluorescein) in the 2 mL solution was then determined using a spectrofluorometer (Duetta, HORIBA scientific) by comparing its fluorescence intensity at 514 nm with those from calibration standards of known concentrations (See Figure 3 for the calibration curve). The minimum concentration that can be measured using spectrofluorometer is 0.02 μg L−1 or 0.04 ng in 2 mL solution.
FIGURE 3

Calibration curve for different concentration standards. Dashed line is the best fit curve with a slope of 1, that is, intensity is linearly proportional to concentration

Calibration curve for different concentration standards. Dashed line is the best fit curve with a slope of 1, that is, intensity is linearly proportional to concentration

RESULTS AND DISCUSSIONS

Minimum suction required for effective containment of aerosol

For aerosol to be effectively contained, we need to generate a negative pressure inside the enclosure. This is achieved when the flow rate of the suction Q suction exceeds the sum of the oxygen flow rate Q O2 and the expiration rate of the patient/manikin Q air, that is, For example, at the maximum oxygen flow rate Q O2 = 60 L min−1, we are able to contain aerosol emissions by turning on the two suction wall units with a combined Q suction = 120 L min−1 (Figure 4a and Video S2). To visualize the aerosol flow, we shone two laser sheets: a blue laser sheet at the sagittal z‐y plane and a green laser sheet at the transverse x‐z plane in front of the barrier. As expected, there is no aerosol leakage and the green laser sheet is not visible since there is minimal aerosol outside the barrier to scatter the light. We also observed fresh air, which was free from aerosol and therefore appeared dark, continually being drawn through gaps at the bottom of the plastic drape.
FIGURE 4

A, Aerosol is effectively contained when Q suction > Q O2 + Q air. B, Otherwise, aerosol leakage from the enclosure can be readily observed. C, Phase diagram for effective (blue filled dots) and ineffective aerosol containment (unfilled red dots)

A, Aerosol is effectively contained when Q suction > Q O2 + Q air. B, Otherwise, aerosol leakage from the enclosure can be readily observed. C, Phase diagram for effective (blue filled dots) and ineffective aerosol containment (unfilled red dots) In contrast, with just one (Q suction = 60 L min−1) or no suction (Q suction = 0 L min−1), the aerosol cannot be contained and spread quickly throughout the entire room (with a floor area of about 30 m2) within minutes. The aerosol leaking out scatters light strongly and renders the green laser sheet visible (Figure 4b and Video S3). The enclosure chamber is also now completely filled with aerosol, and there is no fresh air being drawn in. To test the effectiveness of the aerosol containment under different conditions, we varied the suction rate Q suction from 0 (no suction) to 60 and 120 L min−1 (1 and 2 wall suction units, respectively), while at the same time subjecting the manikin to different oxygen flow rates Q O2 = 0, 30 and 60 L min−1 (Figure 4c). Experimentally, we found that Equation 1 correctly predicts the criterion for effective aerosol confinement (The transition from effective to ineffective containment as predicted by Equation 1 is indicated by the gray line in Figure 4c). In our experiment, the suction ports are located on the two sides of the enclosure. The exact positionings of the suction port, together with the detailed geometry of the enclosure, need to be optimized to avoid deadspots, where there are little circulation and potential accumulation of aerosol. This can be done using detailed computational fluid dynamics study and could be part of a future study.

Reduction in aerosol exposure

To assess the level of protection afforded by the barrier enclosure with and without suction, we added a small amount of fluorescein to the water‐glycerin solution. The level of aerosol exposure can then be quantified by measuring the amount of fluorescein collected by filters of two air samplers placed 50 and 110 cm away from the barrier (samplers 1 and 2 in Figure 5A, respectively). The amount of fluorescein trapped by the filter (Figure 5B) can be deduced by spectrofluorometry. See Section 1D, for experimental details.
FIGURE 5

A, To collect the aerosol leaking out, we placed two air samplers outside the barrier enclosure. B, Aerosol droplets (with added fluorescein) were trapped by the filter on the air sampler, which can then be detected using spectrofluorometer. C, The amount of fluorescein collected by air samplers 1 and 2 for enclosure barrier with and without suction can then be compared to the control, that is, no barrier. Error bars are the standard deviation for triplicates

A, To collect the aerosol leaking out, we placed two air samplers outside the barrier enclosure. B, Aerosol droplets (with added fluorescein) were trapped by the filter on the air sampler, which can then be detected using spectrofluorometer. C, The amount of fluorescein collected by air samplers 1 and 2 for enclosure barrier with and without suction can then be compared to the control, that is, no barrier. Error bars are the standard deviation for triplicates We found that for HFNC oxygen therapy with no barrier at Q O2 = 60 L min−1, the amount of fluorescein collected after 20 minutes by samplers 1 and 2 are 15 ± 5 ng and 11 ± 6 ng, respectively (Figure 5C). With a passive barrier and no suction (corresponding to Figure 4B and case b in Figure 4C), the amount of fluorescein collected by samplers 1 and 2 was reduced by about 60% to 5 ± 2 and 4 ± 1 ng, respectively. At maximum suction of Q suction = 120 L min−1 (corresponding to Figure 4A and case a in Figure 4C), the amount of fluorescein reaching the two samplers was reduced by 99% to 0.15 ± 0.08 and 0.19 ± 0.01 ng, respectively. The amount of fluorescein collected by the two air samplers is similar to each other, because the aerosol droplets are uniformly distributed within the experimental time of 20 minutes. Finally, we would like to point out that although a passive barrier with no suction provides some level of protection, large amount of aerosol can accumulate inside the barrier over time, which will be released into the room if the barrier were to be dismantled, for example during a medical emergency.

CONCLUSIONS

In short, we have established the criterion for effective containment of aerosol for barrier enclosure, namely that the suction rate must exceed the oxygen flow rate and the expiration rate of the human breath. We show explicitly that for high (and medically relevant) oxygen flowrate of 60 L min−1, it is possible to significantly reduce aerosol exposure outside the enclosure by 99% with sufficient suction. Given that the barrier enclosure can be made from readily available material such as acrylic and that suction points are commonly found in hospital rooms, we believe such a device can potentially be scaled up and provide additional protection for healthcare workers.

CONFLICT OF INTERESTS

Ngee Ann Polytechnic and Ng Teng Fong general Hospital have filed a patent submission for the barrier enclosure design.

AUTHORS’ CONTRIBUTIONS

D.D., L.X.J., C.K.T., and O.S.T. conceived the research idea and supervised the research. O.S.T. and C.K.T designed and built the barrier enclosure. D.D., M.L., T.L., A.S., X.K., and T.T. contributed to the laser visualization experiment. I.L., S.S., and A.S. performed the air sampling and subsequent fluorescein quantification. D.D. and C.K.T. wrote up the manuscript. All authors have read and approved the manuscript.

PEER REVIEW

The peer review history for this article is available at https://publons.com/publon/10.1111/ina.12828. Video S1 Click here for additional data file. Video S2 Click here for additional data file. Video S3 Click here for additional data file.
  18 in total

Review 1.  High flow nasal cannula compared with conventional oxygen therapy for acute hypoxemic respiratory failure: a systematic review and meta-analysis.

Authors:  B Rochwerg; D Granton; D X Wang; Y Helviz; S Einav; J P Frat; A Mekontso-Dessap; A Schreiber; E Azoulay; A Mercat; A Demoule; V Lemiale; A Pesenti; E D Riviello; T Mauri; J Mancebo; L Brochard; K Burns
Journal:  Intensive Care Med       Date:  2019-03-19       Impact factor: 17.440

2.  More on Barrier Enclosure during Endotracheal Intubation.

Authors:  Pete G Kovatsis; Clyde T Matava; James M Peyton
Journal:  N Engl J Med       Date:  2020-05-05       Impact factor: 91.245

3.  Modified Wake Forest Type Protective Shield for an Asymptomatic, COVID-19 Nonconfirmed Patient for Intubation Undergoing Urgent Surgery.

Authors:  Sakura Kinjo; Matthew Dudley; Norihiro Sakai
Journal:  Anesth Analg       Date:  2020-08       Impact factor: 5.108

Review 4.  Small droplet aerosols in poorly ventilated spaces and SARS-CoV-2 transmission.

Authors:  G Aernout Somsen; Cees van Rijn; Stefan Kooij; Reinout A Bem; Daniel Bonn
Journal:  Lancet Respir Med       Date:  2020-05-27       Impact factor: 30.700

5.  It Is Time to Address Airborne Transmission of Coronavirus Disease 2019 (COVID-19).

Authors:  Lidia Morawska; Donald K Milton
Journal:  Clin Infect Dis       Date:  2020-12-03       Impact factor: 9.079

6.  A Carton-Made Protective Shield for Suspicious/Confirmed COVID-19 Intubation and Extubation During Surgery.

Authors:  Yu Yung Lai; Chia Ming Chang
Journal:  Anesth Analg       Date:  2020-07       Impact factor: 5.108

7.  Effectiveness of a negative-pressure patient isolation hood shown using particle count.

Authors:  Angela L Lang; Kendrick M Shaw; Rodrigo Lozano; Jingping Wang
Journal:  Br J Anaesth       Date:  2020-05-15       Impact factor: 9.166

8.  Microbial communities in the tropical air ecosystem follow a precise diel cycle.

Authors:  Elena S Gusareva; Enzo Acerbi; Kenny J X Lau; Irvan Luhung; Balakrishnan N V Premkrishnan; Sandra Kolundžija; Rikky W Purbojati; Anthony Wong; James N I Houghton; Dana Miller; Nicolas E Gaultier; Cassie E Heinle; Megan E Clare; Vineeth Kodengil Vettath; Carmon Kee; Serene B Y Lim; Caroline Chénard; Wen Jia Phung; Kavita K Kushwaha; Ang Poh Nee; Alexander Putra; Deepa Panicker; Koh Yanqing; Yap Zhei Hwee; Sachin R Lohar; Mikinori Kuwata; Hie Lim Kim; Liang Yang; Akira Uchida; Daniela I Drautz-Moses; Ana Carolina M Junqueira; Stephan C Schuster
Journal:  Proc Natl Acad Sci U S A       Date:  2019-10-28       Impact factor: 11.205

9.  Visualizing droplet dispersal for face shields and masks with exhalation valves.

Authors:  Siddhartha Verma; Manhar Dhanak; John Frankenfield
Journal:  Phys Fluids (1994)       Date:  2020-09-01       Impact factor: 3.521

View more
  5 in total

1.  Airborne SARS-CoV-2 surveillance in hospital environment using high-flowrate air samplers and its comparison to surface sampling.

Authors:  Alicia Xy Ang; Irvan Luhung; Bintou A Ahidjo; Daniela I Drautz-Moses; Paul A Tambyah; Chee Keng Mok; Kenny Jx Lau; Sai Meng Tham; Justin Jang Hann Chu; David M Allen; Stephan C Schuster
Journal:  Indoor Air       Date:  2021-09-14       Impact factor: 6.554

Review 2.  SARS-CoV-2 in wastewater: From detection to evaluation.

Authors:  Danwei Zhang; Solco S Faye Duran; Wei Yang Samuel Lim; Chee Kiang Ivan Tan; Wun Chet Davy Cheong; Ady Suwardi; Xian Jun Loh
Journal:  Mater Today Adv       Date:  2022-01-25

3.  Polylactic acid face masks: Are these the sustainable solutions in times of COVID-19 pandemic?

Authors:  Xiang Yun Debbie Soo; Suxi Wang; Chee Chuan Jayven Yeo; Jiuwei Li; Xi Ping Ni; Lu Jiang; Kun Xue; Zibiao Li; Xunchang Fei; Qiang Zhu; Xian Jun Loh
Journal:  Sci Total Environ       Date:  2021-10-19       Impact factor: 7.963

4.  Design and development of multilayer cotton masks via machine learning.

Authors:  Y Leow; J K Shi; W Liu; X P Ni; P Y M Yew; S Liu; Z Li; Y Xue; D Kai; X J Loh
Journal:  Mater Today Adv       Date:  2021-11-01

5.  Effectiveness of a novel semi-closed barrier device with a personalized exhaust in cough aerosol simulation according to particle counts and visualization of particles.

Authors:  Ryohei Matsui; Hiroshi Sasano; Takafumi Azami; Hisako Yano; Hiromi Yoshikawa; Yota Yamagishi; Takahiro Goshima; Yuka Miyazaki; Kazunori Imai; Marechika Tsubouchi; Yoichi Matsuo; Shuji Takiguchi; Tomonori Hattori
Journal:  Indoor Air       Date:  2022-02       Impact factor: 6.554

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.