Tebogo R Tsekeli1, Tsholofelo I Sebokolodi1, Hassan Karimi-Maleh1,2, Omotayo A Arotiba1,3. 1. Department of Chemical Sciences, University of Johannesburg, Doornfontein, Johannesburg 2028, South Africa. 2. University of Electronic Science and Technology of China, Chengdu 611731, China. 3. Centre for Nanomaterials Science Research, University of Johannesburg, Johannesburg 2028, South Africa.
Abstract
Silver nanoparticles (AgNPs) were synthesized separately and loaded onto the expanded layers of exfoliated graphite (EG) to form a silver nanoparticle-exfoliated graphite nanocomposite (AgNPs-EG). The AgNPs-EG was compressed into a pellet (0.6 cm in diameter) and used to prepare an electrochemical sensor for bisphenol A (BPA) in standard samples and in thermal paper. The synthesized materials were characterized by ultraviolet-visible spectrophotometry, X-ray diffraction spectroscopy, scanning electron microscopy, and energy-dispersive X-ray. The electrochemical behavior of BPA on the AgNPs-EG sensor was investigated by cyclic voltammetry and square wave voltammetry. Under optimized experimental conditions, the oxidation peak current was linearly proportional to bisphenol A concentrations in the range from 5.0 to100 μM, with a coefficient of determination (R2 ) of 0.9981. The obtained limit of detection of the method was 0.23 μM. The fabricated sensor was able to overcome electrode fouling with good reproducibility (RSD = 2.62%, n = 5) by mechanical polishing of the electrode on emery paper. The proposed method was successfully applied to determine bisphenol A in thermal paper samples and demonstrated good accuracy of 93.1 to 113% recovery.
Silver nanoparticles (AgNPs) were synthesized separately and loaded onto the expanded layers of exfoliated graphite (EG) to form a silver nanoparticle-exfoliated graphite nanocomposite (AgNPs-EG). The AgNPs-EG was compressed into a pellet (0.6 cm in diameter) and used to prepare an electrochemical sensor for bisphenol A (BPA) in standard samples and in thermal paper. The synthesized materials were characterized by ultraviolet-visible spectrophotometry, X-ray diffraction spectroscopy, scanning electron microscopy, and energy-dispersive X-ray. The electrochemical behavior of BPA on the AgNPs-EG sensor was investigated by cyclic voltammetry and square wave voltammetry. Under optimized experimental conditions, the oxidation peak current was linearly proportional to bisphenol A concentrations in the range from 5.0 to100 μM, with a coefficient of determination (R2 ) of 0.9981. The obtained limit of detection of the method was 0.23 μM. The fabricated sensor was able to overcome electrode fouling with good reproducibility (RSD = 2.62%, n = 5) by mechanical polishing of the electrode on emery paper. The proposed method was successfully applied to determine bisphenol A in thermal paper samples and demonstrated good accuracy of 93.1 to 113% recovery.
Bisphenol A (BPA) is
one of the highest volume chemicals produced
globally due to its extensive industrial use, such as in the manufacturing
of polycarbonates, epoxy resins, and thermal papers. In the latter
applications, the compound is used as a color developer, together
with a thermochromic dye and an organic solvent, to create a print
on faxes, ATM receipts, bus tickets, receipts, and labels stuck on
retail products.[1] Scientific evidence shows
that BPA is an estrogenic hormone and an endocrine disruptor, which
can easily transfer from the thermal paper (either occupationally
or via everyday handling) to fingers and penetrates to a point where
it reaches the bloodstream, thereby inducing endocrine disruption.[2] Even at trace levels, human exposure to BPA leads
to adverse health issues such asinfertility, diabetes, brain damage,
and various kinds of cancer.[3−5] Despite these facts, recent studies
have shown that BPA is still used in thermal paper printing all over
the world. This risky exposure and the toxicity of BPA in general,
necessitate BPA detection and monitoring.Until recently, BPA
detection was achieved through chromatographic
methods such as high-performance liquid chromatography (HPLC),[6] liquid chromatography-mass spectrometry (LC–MS),[7] and gas chromatography coupled with mass spectrometry
(GC–MS).[8] These methods are associated
with a number of drawbacks that limit their applications. For example,
they require skilled operators, large sample volumes, and are time
consuming and expensive to maintain. On the other hand, electrochemical
methods have extensively been reported in the literature and have
proven to offer various advantages toward BPA detection due to their
high sensitivity, selectivity, low cost, and rapid response. This
is because BPA presents two hydroxyls moieties in its molecular structure,
which can be oxidized at the electrode’s surface, rendering
the possibility for its electrochemical detection.[9]The oxidation of BPA, however, is an irreversible
process that
produces polymeric films, which foul electrode surfaces, and thus,
resulting in diminished oxidative peak currents. It has been proposed
that the polymeric films are due to electrochemical polymerization,
and this phenomenon is considered as the main problem in the electrochemical
detection of all phenolic compounds in general.[7,10,11] Hence, the development of electrochemical
sensors with better properties such as low costs, minimal, or no fouling
effects is still a challenge. To avoid or minimize this issue, several
approaches that include the use of catalytic, anti-fouling nanomaterials
have been reported. For instance, Wannapob et al. proposed a porous gold-based electrochemical sensor for BPA detection
and their electrode provided enhanced surface area and exhibited good
electrolytic activity toward the oxidation of BPA. This work exhibited
a detection limit as low as 2.0 nM.[12] Furthermore,
a more comprehensive study by Ragavan et al., reported
a vast range of nanomaterials including quantum dots, graphene, carbon
nanotubes, and molecularly imprinted polymer-based electrochemical
sensors for BPA.[13] Unfortunately, the issue
of electrode fouling still remains a concern that needs to be mitigated.When compared with other noble metals, silver has received special
attention due to its high electrical conductivity, good catalytic
activity, low toxicity, simple synthesis, and low cost.[14−16] These properties have led to a vast number of studies reporting
on the electrodetection of BPA using silver nanoparticle (AgNPs) composites.
For example, Campos et al. modified a glassy carbon
electrode with a nanocomposite of SiO2/GO/AgNPs to develop
an electrochemical sensor for BPA. The presence of AgNPs led to enhanced
BPA oxidation currents and significantly lowered the detection limit
of the sensor.[17] Baccarin et al. decorated a graphite-polyurethane composite electrode with silver
nanoparticles to fabricate a pen-based electrochemical sensor for
BPA detection. Improved electrokinetics of the sensor was obtained
under optimized AgNPs contents, and the sensor exhibited a BPA detection
limit of 0.24 μM.[18] Other materials
composited with silver nanoparticles for enhanced electrochemical
detection such ascarbon black-silver nanoparticles[19] and Printex nanocarbon-silver nanoparticles[20] have also been reported. The above works motivates
this paper, where silver nanoparticles were composited with exfoliated
graphite (EG) for the fabrication of an electrochemical sensor for
BPA detection. EG, possessing properties such as high electrical conductivity
and electron transport, mechanical strength, and high surface area,
has made a niche for itself as a suitable candidate for various applications,
particularly as a substrate in electrochemistry.[21−28] With compressibility and mechanical strength, EG can be easily compacted
into a working electrode that allows surface regeneration by polishing
on emery paper, as a way of tackling fouling. Based on these inherent
properties of both EG and AgNPs, as well as the motivation from the
above works, we report the use of an AgNPs-EG nanocomposite-based
electrode for the development of an electrochemical sensor for BPA
detection in thermal paper samples.
Results and Discussion
Characterization
of the AgNPs-EG Nanocomposite
The
successful synthesis of AgNPs was confirmed by a characteristic strong
absorbance peak at around 393 nm (Figure S1b),[29,30] and an average particle size (prior to intercalation
into the EG) of 20 nm was estimated from zetasizer. The field emission
scanning electron images of EG and AgNPs-EG nanocomposite are shown
in Figure a,b, respectively.
From Figure a, an
accordion-like structure and the expanded layers of graphite with
open cavities suitable for the entrapment of particles, can be seen.
The presence of AgNPs on EG after modification was evidenced as evenly
distributed high-contrast nanoparticles on the surface of the layers
of graphite on EG (Figure b). However, some agglomerations were observed on the AgNPs-EG
nanocomposite at a higher SEM magnification (Figure S1a). To further confirm the loading of AgNPs and formation
of the nanocomposite, EDX analysis was conducted (Figure d). The presence of the Ag
peak confirmed the presence of AgNPs on the surface of EG.
Figure 1
SEM images
of (a) EG and (b) AgNPs-EG, (c) EDX of AgNPs-EG, and
XRD patterns of (d) EG and (e) AgNPs-EG (Inset shows amplified Ag
peaks suppressed by high intensity EG peaks).
SEM images
of (a) EG and (b) AgNPs-EG, (c) EDX of AgNPs-EG, and
XRD patterns of (d) EG and (e) AgNPs-EG (Inset shows amplified Ag
peaks suppressed by high intensity EG peaks).The XRD pattern of the pristine exfoliated graphite (Figure d) shows two prominent diffraction
peaks at 2θ = 26.33 and 54.52°, corresponding to the (002)
and (004) planes, respectively. These XRD peaks are attributed to
the characteristic diffraction peaks of graphite and crystal phase,
which are in agreement with previous reports.[24,31] For the AgNPs-EG nanocomposite (Figure e and inset), well-defined peaks were observed
at 2θ = 37.74, 44.27, 64.10, 77.25, and 86.83°, corresponding
to (111), (200), (220), (311), and (222) planes of the face centered
cubic (fcc) structure of metallic silver, respectively (JCPDS 04–0783)[15,32−34] while still retaining the characteristic peaks of
EG.
Electrochemical Characterization the Prepared Electrodes
Cyclic voltammetry (CV) and square wave voltammetry (SWV) were used
to compare and investigate the electrochemical behaviors of bare EG
and AgNPs-EG electrodes in [Fe(CN)6]3–/4– as a redox probe and in phosphate buffer solution (PBS). SWV was
chosen because it is a more sensitive and faster technique than differential
pulse voltammetry. It is especially sensitive to surface-confined
reactions. Thus, it is often preferred for organic detection owing
to its ability to limit/omit adsorption effect (such asBPA electro-oxidation).[35] The pristine EG electrode (Figure a) displayed a peak potential
separation (ΔEp) of about 290 mV,
which reduced to about 180 mV after loading with AgNPs. This indicates
that the kinetics of the electron transfer of the redox probe is faster
at the modified EG electrode as a result of the AgNPs. The SWV responses
of the two electrodes in BPA solution are presented in Figure b. A 44.54% current enhancement
and shift toward lower peak potential on the AgNPs-EG electrode compared
to that of the EG electrode can be observed. The marked increase in
the electro-oxidation peak of BPA shows the electrocatalytic nature,
large surface area, and excellent conductivity conferred on EG by
the AgNPs. The reduction of anodic potential denotes an increase in
the electron transfer rate constant from BPA.[36] Furthermore, the reduction in the peak potential of BPA at the AgNPs-EG
electrode suggests that a lower energy is needed to oxidize BPA; this
is a further evidence of the electrocatalytic nature of AgNPs. A reduced
peak potential can also help in eliminating some interference from
other electroactive species. The presence of AgNPs in the AgNPs-EG
electrode was further proven electrochemically by CV using PBSas
the electrolyte (Figure c). The Ag+/Ag redox behavior of the AgNPs-EG electrode
was observed at potentials 380 and 620 mV for the oxidation of AgNPs
and at −80 mV for the reduction of Ag cations. These redox
peaks were not observed on the pristine EG electrode. These results
corroborated those in Figure , in the sense that AgNPs were successfully grafted on the
EG substrate. Figure d shows successive scans at the AgNPs-EG electrode surface to study
electrode surface fouling. The obtained results showed that the intensity
of oxidation peaks of BPA declines cycle after cycle. These results
clearly show that even though the modified electrode responded better
than pristine EG toward BPA (as shown in Figure b), the electrode is still susceptible to
fouling. This observation can be explained as a result of an electrode
fouling phenomenon, in the sense that oxidative and polymeric products
of BPA-oxidation are deposited on the electrode surface, which blocks
further access of BPA to the electrode.[37] The oxidation current signal declined by 74% on the second scan
and further declined by 81% on the third scan. Moreover, an 83% current
decline was recorded on the seventh scan (relative to the first scan),
thus indicating a continuous and progressive deactivation of the electrode.
As a way of tackling this problem, we propose mechanical polishing
of the electrode surface between scans to counteract fouling. The
fouled electrode was polished on P1200-grit emery paper, and the results
showed successful regeneration of the electrode since similar BPA
oxidation peak currents were obtained after every polishing, as shown
in Figure e. This
electrode regeneration method was reproducible, with electrochemical
responses having a relative standard deviation (RSD) of 2.62%. The
minor peaks observed (in Figure. b,d,e) at potentials between 0.20 and 0.25 V are possibly
attributed to byproducts of the electro-oxidation of BPA.[38]
Figure 2
(a) CV curves of the bare EG and AgNPs-EG electrodes in
5 mM [Fe(CN)6]3–/4– containing
0.1 M KCl; (b)
SWV curves of bare EG and AgNPs-EG electrodes; (c) CV characterization
of EG and AgNPs-EG in 10 mM PBS, pH 7.2. (d) Successive SWVs at the
AgNPs-EG electrode; (e) SWVs of the AgNPs-EG electrode at 5 runs with
polishing. SWV experiments were conducted in 5 mM BPA in 10 mM PBS,
pH 10 at 0.05 V amplitude, 25 Hz frequency, and a step potential of
0.01 V.
(a) CV curves of the bare EG and AgNPs-EG electrodes in
5 mM [Fe(CN)6]3–/4– containing
0.1 M KCl; (b)
SWV curves of bare EG and AgNPs-EG electrodes; (c) CV characterization
of EG and AgNPs-EG in 10 mM PBS, pH 7.2. (d) Successive SWVs at the
AgNPs-EG electrode; (e) SWVs of the AgNPs-EG electrode at 5 runs with
polishing. SWV experiments were conducted in 5 mM BPA in 10 mM PBS,
pH 10 at 0.05 V amplitude, 25 Hz frequency, and a step potential of
0.01 V.
Effects of Electrolyte
pH on BPA Electro-Oxidation
SWV was used to study the effects
of electrolyte pH on the peak current
and peak potential in 10 mM PBS, pH 7.0 to 12, containing 5 mM BPA.
As shown in Figure , an increase in the anodic peak current with increasing pH was observed
until pH 9.0 after which a sharp decrease was observed when the solution
pH exceeded pH 9.0. The maximum response to pH was lower than the
pKa of BPA (pKa = 9.73), indicating that the non-dissociated BPA interacted or adsorbed
better than the dissociated BPA on the AgNPs-EG surface.[39] Therefore, pH 9.0 was selected as the optimum
pH value for all subsequent electrochemical experiments. Furthermore,
under this study, the anodic peak potential (Epa) shifted negatively with an increase in pH, conforming to
the equation: Epa (V) = −0.05569
pH + 1.10112 (R = 0.9925). The obtained
slope of 0.05569 V/pH is approximately close the theoretical value
of 0.0576 V/pH and indicates that the electron transfer was accompanied
by an equal number of protons on the AgNPs-EG electrode.[40] Therefore, it can be inferred that the electrochemical
oxidation of BPA at the AgNPs-EG electrode is a two-electron and two-proton
process and this is illustrated in Scheme .
Figure 3
Effects of solution pH on BPA oxidation of 5
mM BPA in 10 mM PBS.
SWV experiments were conducted at 0.05 V amplitude and 25 Hz frequency,
a step potential of 0.01 V, within 0.0 to 1.0 V potential window.
Scheme 1
Possible Reaction Mechanism for BPA Oxidation on the
AgNPs-EG Electrode
Effects of solution pH on BPA oxidation of 5
mM BPA in 10 mM PBS.
SWV experiments were conducted at 0.05 V amplitude and 25 Hz frequency,
a step potential of 0.01 V, within 0.0 to 1.0 V potential window.
Analytical Performance
of the AgNPs-EG Sensor toward BPA Detection
Under the optimal
experimental conditions, the prepared AgNPs-EG
sensor was employed to detect a series of BPA solutions with different
concentrations by SWV and responses are presented in Figure a. The peak currents increased
linearly with an increase in BPA concentrations in the range from
5.0 to 100 μM with a linear regression equation as I (μA) = 0.95988 [BPA] (μM) + 11.8201 (R 0.9981) (Figure b). The detection limit of the method was calculated according
to the equation, LOD = 3 × SD/b, where SD is
the standard deviation of three blank measurements and b is the slope of the calibration plot. From Figure b, a detection limit of 0.23 μM was
calculated for this work. For the control study, SWV measurements
of the EG electrode were also conducted within the same linear concentration
range, under the same optimized electrolyte pH. The calibration plot
of the EG electrode shows a lower sensitivity (as interpreted by a
lower slope, and was not as linear in the concentration range analyzed
(as interpreted from a lower correlation coefficient value of 0.8272))
due to its ineffectiveness in the electro-oxidation of BPA. Table compares the performance
of the AgNPs-EG sensor with those reported previously. It can thus
be inferred that the AgNPs-EG-based sensor exhibits either comparable
or even better response toward BPA detection.
Figure 4
(a) SWVs of the AgNPs-EG
sensor in 10 mM PBS, pH 9.0 containing
different concentrations of BPA (5.0, 20, 30, 50, 80, and 100) and
(b) calibration plot of response currents of AgNPs-EG (solid line)
and EG (broken line) electrodes as a function of BPA concentrations.
Table 1
Comparison of Different Electrochemical
Sensors for BPA Detection
electrode
configuration
linear range
(μM)
detection
limit (μM)
references
(3D-GN)/Cu/Fe3O4
7.2–18
1.70
(41)
PEDOT-GCE
40–410
22.0
(42)
AuNp@MOF
200–1000
37.80
(43)
CNHs-Nafion/GCE
2.0–1000
1.80
(44)
SWCNT/GCE
10–100
7.30
(45)
Gr-AuCuNP/AuE
0.1–30
1.31
(46)
Gr-AuCuNP/AuE
0.1–100
1.91
(46)
MWCNTs/GCE
2.0–30
0.5
(47)
AgNPs-EG
5.0–100
0.23
this work
(a) SWVs of the AgNPs-EG
sensor in 10 mM PBS, pH 9.0 containing
different concentrations of BPA (5.0, 20, 30, 50, 80, and 100) and
(b) calibration plot of response currents of AgNPs-EG (solid line)
and EG (broken line) electrodes as a function of BPA concentrations.
Interference and Stability Studies of the AgNPs Electrode
The influence of interferences on the detection of BPA was studied
in a 10 mM PBS (pH 9.0) electrolyte solution containing BPA in the
presence of possible interfering species like 4-nitrophenol, 3-chlorophenol,
phenol, and 4-phenylphenol. Figure a shows that no marked interference was caused by the
presence of these interfering molecules. Intensity current changes
of only 9.89, 16, 0, and 0.55% were observed when 3-chlorophenol,
4-nitrophenol, 4-phenylphenol, and phenol were present, respectively.
This is an indication that the AgNPs-EG sensor exhibits relatively
reasonable ability to detect BPA without the interference of the chosen
species. Furthermore, the stability of the electrode was conducted
for a period of 2 weeks. After detection of 30 μM BPA in 10
mM PBS (pH 9.0), the electrode was polished, rinsed with deionized
water, and stored for 14 days. The oxidation of the same BPA concentration
after this storage period retained more than 85% of the initial current
response, as shown in Figure b. Hence, the AgNPs-EG sensor exhibits good stability for
BPA detection.
Figure 5
(a) Effect of the presence of 50 μM 3-chlorophenol,
4-nitrophenol,
phenol, and 4-phenylphenol on the signal responses obtained for 50
μM BPA in 10 mM PBS, pH 9.0, (b) stability study of AgNPs-EG
electrode for a period of 2 weeks.
(a) Effect of the presence of 50 μM 3-chlorophenol,
4-nitrophenol,
phenol, and 4-phenylphenol on the signal responses obtained for 50
μM BPA in 10 mM PBS, pH 9.0, (b) stability study of AgNPs-EG
electrode for a period of 2 weeks.
Analysis of Real Thermal Paper Samples
To evaluate
the performance of the AgNPs-EG electrode in practical analytical
applications, the determination of BPA in thermal paper samples was
carried out through a recovery study and the results are shown in Table . The recovery of
BPA was from 93.1 to 113%, indicating that the proposed electrochemical
sensor is reliable, effective, and accurate for practical applications.
Table 2
Detection of BPA in Thermal Papers
using the AgNPs-EG Electrochemical Sensor
sample
added (μM)
(%)
found (μM)
recovery
sample 1
10
11.35 ± 0.45
113
sample
2
10
17.66 ± 1.17
93.1
Conclusions
We have developed a
simple anti-fouling electrochemical sensor
for the quantitative detection of BPA using a silver nanoparticle-modified
exfoliated graphite electrode. The anti-fouling property of the sensor
is by a simple step of mechanical polishing of the electrode for surface
renewal. Relative to the pristine EG electrode, the silver nanoparticles
incorporated the electrode allowed for the oxidation of BPA in less
basic conditions and further improved/lowered the detection limit
of the sensor. The proposed sensor was further applied in the detection
of BPA in thermal paper real samples and exhibited good recoveries.
Furthermore, the sensor exhibited insignificant interferences from
potential BPA co-existing molecules and also showed good stability.
Therefore, this easy to fabricate electrochemical sensor offers a
potential application for BPA environmental monitoring while mitigating
the problem of electrode fouling by BPA.
Experimental Section
Chemicals
and Reagents
Bisphenol A (≥99%), natural
graphite flakes, sulfuric acid (99.98%), trisodium citrate (AR), 4-nitrophenol
(98%), phenol (99%), 4-phenylphenol (97%), 3-chlorophenol (98%), and
potassium chloride (≥99%) were all obtained from Sigma-Aldrich
(South Africa) and were used as received. Nitric acid (≥55%)
was obtained from Honeywell Fluka (South Africa). Silver nitrate (99%)
was obtained from ACE Chemicals (South Africa). Sodium borohydride
(AR) was purchased from Sisco Research Laboratories Pvt. Ltd. (India).
Phosphate buffer solutions (PBS, 10 mM) of different pH values were
prepared with 10 mM sodium dihydrogen phosphate (AR, Sigma-Aldrich)
and 10 mM disodium hydrogen phosphate (AR, Sigma-Aldrich), and their
pH values were adjusted using sodium hydroxide (AR, Sigma-Aldrich)
and hydrochloric acid (32%, Sigma-Aldrich). All the solutions were
prepared using ultrapure water (18.2 MΩ.cm at 22 °C).
Instrumentation and Apparatus
Scanning electron microscopy
(SEM) analysis and elemental composition analysis were conducted by
a Zeiss Gemini 2 Crossbeam 540 FEG SEM with an EDX detector (United
Kingdom). A Malvern Zetasizer (South Africa) was used for the particle
size distribution of AgNPs. The ultraviolet–visible (UV–Vis)
spectra were recorded with an Agilent UV–Vis spectrophotometer
(Malaysia). X-ray diffraction patterns were examined under an X-ray
diffractometer (XRD, Rigaku Ultima IV, Japan). Electrochemical measurements
were conducted using an Ivium compactstat potentiostat (Netherlands)
system connected to a three-electrode system. The electrode system
comprised of an AgNPs-EG (and EG) compressed pellet on copper wire
as a working electrode, a platinum wire as a counter electrode, and
Ag/AgCl (in 3 KCl) as a reference electrode. All solutions were sparged
with argon gas before electrochemical measurements.
Synthesis of
Silver Nanoparticles
AgNPs were synthesized
through the reduction of Ag ions from silver salts using sodium borohydride
and trisodium citrate. Sodium borohydride and sodium citrate acted
as primary and secondary reducing agents, respectively, and also as
stabilizing agents. This was conducted at two different temperatures,
60 and 90 °C. The required amounts of freshly prepared aqueous
solutions of sodium borohydride and trisodium citrate were mixed and
heated to 60 °C for 30 min under magnetic stirring to produce
a homogeneous solution. After this, the required amount of AgNO3 was added to the reduction solution and the mixture immediately
turned light yellow, confirming the formation of AgNPs. Thereafter,
the temperature was increased to 90 °C and kept for 30 min under
continuous stirring. The typical input amounts of each precursors
were 5 mM sodium borohydride, 2 mM trisodium citrate, and 1.22 mM
silver nitrate on a base of 500 mL of deionized water.[48]
Preparation of EG and AgNPS-EG Electrodes
The preparation
of EG followed a two-step procedure: intercalation and exfoliation.
The intercalation was achieved by soaking natural graphite flakes
in a mixture of concentrated HNO3 and H2SO4 (in a volume ratio of 1:3) for 72 h at room temperature.
The intercalated material was washed with deionized water until a
pH of 7 was attained and air dried. Exfoliation was then carried out
by subjecting the dried material to a preheated furnace at of 800
°C for 1 min.[25]The preparation
of a composite of EG and silver nanoparticles (AgNPs-EG) was based
on the following procedure: the prepared exfoliated graphite was poured
in the as-prepared silver colloidal solution and the mixture was subsequently
transferred into a 100 mL Teflon-lined autoclave. The autoclave was
sealed and heated in an oven at 110 °C overnight. After the reaction
duration, the autoclave was allowed to cool to room temperature and
the resulting product was collected by filtration and air dried.About 500 mg of the AgNPs-EG nanocomposite was weighed and compacted
without any binder into a pellet using a hydraulic press at a high
pressure for 12 h. The pellet was employed in the construction of
an electrode by using a copper wire, the conductive silver paint,
and a glass tube. The copper wire was coiled at one end to form a
flat surface, which was scrubbed on a P120-grit sandpaper to remove
any oxides on the surface. The AgNPs-EG pellet was then placed on
the coiled flat surfaced copper wire with the assistance of silver
paint and allowed to dry overnight. The electrode was then inserted
into a glass tube and the edges of the pellet were subsequently covered
with Teflon tape so that the current is contributed by only the basal
plane of the electrode. The same procedure was followed for the fabrication
of the EG electrodes. The diameter of the EG electrode was 0.6 cm.
Fouled electrodes were cleaned by polishing on P1200-grit emery paper
after every scan, to have renewed and activated surfaces, as a way
of tackling electrode fouling.
BPA Extraction and Sample
Preparation
The determination
of BPA in thermal printing paper is a two-step procedure. First, the
compound is extracted from paper samples with polar organic solvents
like methanol or acetonitrile, etc., and second, detection is employed.[49] For this work, two types of thermal papers (travel
tickets and retail store receipts) were collected. The samples were
referred to as sample 1 and 2. BPA was extracted according to a previously
published method with slight modifications.[2] Briefly, thermal papers were cut into 4 mm diameter circular pieces
using a puncher. Then, 1.0 g from each sample was immersed into 10
mL of ethanol overnight at 60 °C for the extraction process.
After centrifugation, supernatants were filtered with a 0.45 μm
filter. Sample solutions were then diluted with PBS and detected by
SWV.
Authors: Paulo A Raymundo-Pereira; Anderson M Campos; Fernando C Vicentini; Bruno C Janegitz; Camila D Mendonça; Leonardo N Furini; Naiza V Boas; Marcelo L Calegaro; Carlos J L Constantino; Sergio A S Machado; Osvaldo N Oliveira Journal: Talanta Date: 2017-06-27 Impact factor: 6.057
Authors: C Y Flores; C Diaz; A Rubert; G A Benítez; M S Moreno; M A Fernández Lorenzo de Mele; R C Salvarezza; P L Schilardi; C Vericat Journal: J Colloid Interface Sci Date: 2010-06-25 Impact factor: 8.128
Authors: Thais S Alves; Jaqueline S Santos; Antonio R Fiorucci; Gilberto J Arruda Journal: Mater Sci Eng C Mater Biol Appl Date: 2019-08-02 Impact factor: 7.328
Authors: D G Mita; A Attanasio; F Arduini; N Diano; V Grano; U Bencivenga; S Rossi; A Amine; D Moscone Journal: Biosens Bioelectron Date: 2007-03-25 Impact factor: 10.618
Authors: Marina Baccarin; Mariani A Ciciliati; Osvaldo N Oliveira; Eder T G Cavalheiro; Paulo A Raymundo-Pereira Journal: Mater Sci Eng C Mater Biol Appl Date: 2020-04-28 Impact factor: 7.328
Authors: Paulo A Raymundo-Pereira; Anderson M Campos; Thiago M Prado; Leonardo N Furini; Naiza V Boas; Marcelo L Calegaro; Sergio A S Machado Journal: Anal Chim Acta Date: 2016-04-27 Impact factor: 6.558