Investigations on the near surface dose for three 10-MV x-ray beam accelerators with emphasis on the reduction of electron contamination.

B M Rao1, S G Prasad, K Parthasaradhi, Y Lee, R Ruparel, R Garces.   

Abstract

A study of three 10-MV x-ray clinical accelerators with emphasis on the reduction of electron contamination was conducted. This study, which was performed with different types of trays and filters, suggests that, at 100-cm source-surface distance (SSD), Pb can be used as an effective filter material up to 30 X 30 cm2; however, due to its transparency, a Clear-Pb tray is useful for field sizes up to a 20 X 20 cm2. Percent depth doses for a few selected depths and field sizes at this nominal SSD were examined. No significant differences, with the exception of the location of Dmax, amongst the three accelerators were noticed.

Entities:  

Mesh:

Year:  1988        PMID: 3386598     DOI: 10.1118/1.596256

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071