Pseudohyperchloremia and Negative Anion Gap - Think Salicylate!

Michael R Wiederkehr1, Raul Benevides2, Carol A Santa Ana3, Michael Emmett4.   

Abstract

BACKGROUND: Pseudohyperchloremia results in a very low or negative anion gap. Historically, the most common cause of this artifact was bromide poisoning. Bromide salts have been removed from most medications and bromism has become very uncommon. More recently, the introduction of chloride ion selective sensing electrodes (Cl-ISE) has generated a new cause of pseudohyperchloremia-salicylate poisoning. We describe 5 such patients and quantitate the error generated by this measurement artifact.
METHODS: The magnitude of artifactual hyperchloremia generated by high salicylate levels was quantified in 5 patients by measuring chloride concentration with several Cl-ISEs from different manufacturers and with Cl-ISEs of different "ages," and comparing these results to measurements with a chloridometer (coulometric titration), which is free of the salicylate artifact.
RESULTS: Cl-ISEs from different manufacturers generated a wide range of artifactual chloride concentration elevation. Furthermore, the same Cl-ISE generated increasingly severe pseudohyperchloremia as it was repeatedly reused over time and "aged."
CONCLUSIONS: Salicylate interferes with measurement of the blood chloride concentration when a Cl-ISE is used. The severity of this artifact is related to the salicylate level, the specific Cl-ISE, and the "age" of the electrode. Toxic blood salicylate levels can generate marked pseudohyperchloremia, and consequently, an artifactual very small or negative anion gap. The large anion gap metabolic acidosis typical of salicylate poisoning is masked by this artifact. Salicylate has become the most common cause of pseudohyperchloremia, and physicians should immediately consider salicylate poisoning whenever the combination of hyperchloremia and a very small or negative anion gap is reported by the laboratory.
Copyright © 2021 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Anion gap; Metabolic acidosis; Pseudohyperchloremia; Salicylate poisoning

Year:  2021        PMID: 33864761     DOI: 10.1016/j.amjmed.2021.03.017

Source DB:  PubMed          Journal:  Am J Med        ISSN: 0002-9343            Impact factor:   4.965


  2 in total

1.  Data on the clinical, analytical, and laboratory factors associated with negative anion gaps at an academic medical center.

Authors:  Joseph M Laakman; Zachary J Fleishhacker; Matthew D Krasowski
Journal:  Data Brief       Date:  2022-06-06

2.  Pseudohyperchloremia caused by the long-term use of phenobarbital and sodium bromide compound tablets: a case report.

Authors:  Xi Jiang; Meng-Na Liu; Zhi-Yuan Zhou; Wei-Jia Lin; Zhao-Fan Luo
Journal:  Ann Transl Med       Date:  2022-08
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.