Intra- and inter-rater reliability of rectus femoris muscle thickness measured using ultrasonography in healthy individuals.

Yosuke Takahashi1, Yuji Fujino2, Kohei Miura1, Ayumi Toida3, Tadamitsu Matsuda4, Shigeru Makita5.   

Abstract

BACKGROUND: Ultrasonography (US) is a feasible and accessible method for the measurement of skeletal muscle mass. This technique presents acceptable intra-rater reliability; however, there are a few reports on its inter-rater reliability. Additionally, relative reliability should equally be inspected to determine the presence of systematic errors. Therefore, this study aimed to investigate the intra- and inter-rater reliabilities and absolute reliability of rectus femoris muscle thickness as measured using US.
METHODS: The participants included in our study comprised 12 healthy young men (26.5 ± 3.9 years. Rectus femoris muscle thickness was measured from the right side of the thigh using US by two trained physical therapists. Inter- and intra-rater reliabilities were determined using the intraclass correlation coefficient (ICC) (1, 1) and ICC (2, 1) methods, respectively. Absolute reliability was evaluated using Bland - Altman analysis. Additionally, we calculated the minimal detectable change at the 95% level of confidence (MDC95). RESULT: According to the results of the Bland - Altman analysis, no fixed or proportional errors were present. The ICC (1, 1) was 0.95, and the ICC (2, 1) was 0.70. The MDC95 values of rectus femoris thickness for the intra- and inter-rater reliabilities were 2.0 mm and 4.3 mm, respectively.
CONCLUSIONS: In our study, intra- and inter-rater reliabilities were measured at "excellent" and "moderate" levels in the healthy individuals based on a previously defined scale. Moreover, we determined the measurement error for quantifying rectus femoris thickness. Therefore, the measurement of rectus femoris thickness using US could be considered applicable in clinical research.

Entities:  

Keywords:  Rectus femoris muscle; Reliability; Skeletal muscle mass; Ultrasonography

Year:  2021        PMID: 33856566     DOI: 10.1186/s13089-021-00224-8

Source DB:  PubMed          Journal:  Ultrasound J        ISSN: 2524-8987


  1 in total

1.  Occurrence and distribution of fat in human muscles at various age levels; a morphologic and roentgenologic investigation.

Authors:  A FRANTZELL; B E INGELMARK
Journal:  Acta Soc Med Ups       Date:  1951-04
  1 in total
  2 in total

1.  Reliability and Validity of an Ultrasound-Based Protocol for Measurement of Quadriceps Muscle Thickness in Children.

Authors:  Emil Rydell Högelin; Kajsa Thulin; Ferdinand von Walden; Lotta Fornander; Piotr Michno; Björn Alkner
Journal:  Front Physiol       Date:  2022-06-27       Impact factor: 4.755

2.  Skeletal muscle analysis of panoramic ultrasound is reliable across multiple raters.

Authors:  Christopher J Cleary; Omid Nabavizadeh; Kaycie L Young; Ashley A Herda
Journal:  PLoS One       Date:  2022-05-02       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.