| Literature DB >> 33853989 |
Jie Jin1, Qingjun Jiang1, Jun Bai1, Lefeng Qu1.
Abstract
PURPOSE: To evaluate the safety and efficacy of a novel vascular-friendly thoracic stent graft for patients with acute complicated type B aortic dissection (ac-TBAD).Entities:
Keywords: TEVAR; aortic dissection; complication; true lumen rate; vascular-friendly stent graft
Mesh:
Year: 2021 PMID: 33853989 PMCID: PMC8343027 DOI: 10.5761/atcs.oa.20-00383
Source DB: PubMed Journal: Ann Thorac Cardiovasc Surg ISSN: 1341-1098 Impact factor: 1.520
Baseline characteristics and operation-related information (N = 63)
| Characteristics | N = 63 (%) |
|---|---|
| Demographic data | |
| Male sex | 52 (82.5) |
| Age (years) | 54.2 ± 12.4 |
| Risk factors | |
| Smoking | 42 (66.7) |
| Hypertension | 51 (81.0) |
| Diabetes | 7 (11.1) |
| Hyperlipidemia | 15 (23.8) |
| Comorbidities | |
| Coronary artery disease | 6 (9.5) |
| Peripheral artery disease | 9 (14.3) |
| Chronic renal failure | 16 (25.3) |
| Pulmonary disease | 9 (14.3) |
| Complications | |
| Impending rupture | 2 (3.2) |
| Acute renal failure | 12 (19.0) |
| Visceral ischemia | 15 (23.8) |
| Refractory hypertension | 7 (11.1) |
| Refractory pain | 27 (42.9) |
| Technical success | 63 (100) |
| Anesthesia (n) | |
| General | 33 (52.7) |
| Local | 30 (47.6) |
| Duration of procedure (min) | 57.3 ± 24.2 |
| Intraoperative blood loss (ml) | 55.4 ± 42.6 |
| Transfusion | 16 (25.4) |
| No. of stent graft (n) | |
| 1 | 53 (84.1) |
| 2 | 10 (15.9) |
| Reconstruction of LSA (n) | 20 (31.7) |
Categorical variables are presented as number (%). Continuous variables are presented as mean ± standard deviation.
RBS: restrictive bare stent
Thirty-day and follow-up outcomes
| Variables | N = 63 (%) |
|---|---|
| Length of stay (days) | 10.1 ± 5.2 |
| Perioperative complications (n) | |
| Access site events | 6 (9.5) |
| Fever | 13 (20.6) |
| Type I/III endoleak | 0 |
| Type II endoleak | 2 (3.1) |
| Cerebral events | 0 |
| Spinal cord drainage | 2 (3.1) |
| Retrograde type A dissection | 0 |
| Cardiac events | 1 (1.6) |
| Respiratory complications | 6 (9.5) |
| Access site complications | 6 (9.5) |
| In-hospital mortality (n) | 0 |
| Mean follow-up time (months) | 30.1 ± 18.9 |
| Median follow-up period (months) | 30.2 (range, 1.3–70.6) |
| All-cause mortality during follow-up | 2 (3.1) |
| Aorta-related mortality | 1 (1.6) |
| Myocardial infarction | 1 (1.6) |
| Complications during follow-up | |
| Stent displacement / fracture | 0 |
| Type I/III endoleak | 1 (1.6) |
| Stent-induced new entry | 0 |
| Retrograde type A dissection | 1 (1.6) |
| Secondary intervention | 1 (1.6) |
True lumen rate of aortic
| Follow-up N = 63 (%) | ||||
|---|---|---|---|---|
| Baseline | 1 month | 6 months | 12 months | |
| LSA Level | 67.9% ± 24.4% | 86.3% ± 16.2% | 95.7% ± 5.0% | 98.0% ± 2.6% |
| Middle-thoracic aorta level | 54.8% ± 22.1% | 84.7% ± 12.4% | 92.8% ± 6.9% | 96.9% ± 3.1% |
| Coeliac trunk level | 67.4% ± 26.7% | 91.7% ± 7.2% | 96.8% ± 3.1% | 98.4% ± 1.6% |
LSA: Left subclavian artery