| Literature DB >> 33850948 |
Ali Dinari1, Seyedeh Sahar Mortazavi Farsani2, Soheila Mohammadi3, Farhood Najafi4, Mahdi Abdollahi1.
Abstract
BACKGROUND: Dynamic light scattering (DLS) and electron microscopy (EM) are the most practical techniques for nanoparticles (NPs) characterization. However, the impediments which involved the sample preparation method lead to failure in provided results of mentioned device analysis. These problems will be intensifying, if the examined samples are the soft nanocarriers such as organic ones or biological samples.Entities:
Keywords: Dynamic light scattering (DLS), EM; Soft nanocarriers; Triple steps
Year: 2020 PMID: 33850948 PMCID: PMC8035413 DOI: 10.30498/IJB.2020.2645
Source DB: PubMed Journal: Iran J Biotechnol ISSN: 1728-3043 Impact factor: 1.671
DLS results of filtered and sonicated various samples.
| Product name | Dilution | Beforre filtration and sonication | After filtration and sonication |
|---|---|---|---|
| Size; nm (PDI) | Size; nm (PDI) | ||
| 1/10 | 579.7 (0.527) | 68.18 (0.572) | |
| 1/100 | 397.7 (0.420) | 20.17 (0.429) | |
| 1/10 | 525.1 (0.594) | 98.38 (1.00) | |
| 1/100 | 374.1 (0.538) | 39.71 (0.418) | |
| 1/10 | 971.1 (0. 629) | 129.4 (0.516) | |
| 1/100 | 651.9 (0.673) | 203.9 (0.373) |
Figure 1The effects of sonication and filtration on TEM results A) Particles agglomeration in nonsonicated sample. B) Particles with proper dispersion. C) Smeared particles by dust in non-filtered sample. D) Filtered once.
Figure 2EM images of polymeric nanocarriers. A) SEM image of aggregated nanogels in high concentration (excess of 100 µg/ml-1) samples. B) The same sample prepared in optimized method (filtration, dilution (50 to 100 µg.ml-1) and sonication (15 min)). C) TEM image of agglomerated particles (micelles) in high concentration (excess of 100 µg.ml-1) samples. D) The same sample prepared in optimized method.