Alba Santos1, Nicolás Macías2, Almudena Vega2, Soraya Abad2, Tania Linares2, Inés Aragoncillo2, Leonidas Cruzado3, Cristina Pascual4, Marian Goicoechea2, Juan Manuel López-Gómez2.
Abstract
BACKGROUND: Low-molecular-weight heparins (LMWHs) are easily dialysable with high-flow membranes; however, it is not clear whether the LMWH dose should be adjusted according to the membrane type and dialysis technique. This study aimed to evaluate the influence of the dialyser on anticoagulation of the extracorporeal dialysis circuit.Entities:
Keywords: anticoagulation; antifactor Xa activity; expanded haemodialysis; high-flux haemodialysis; low-molecular-weight heparin; online haemodiafiltration
Year: 2020 PMID: 33841857 PMCID: PMC8023216 DOI: 10.1093/ckj/sfaa057
Source DB: PubMed Journal: Clin Kidney J ISSN: 2048-8505
Main characteristics of dialysers used in this study
| Dialyzer | FxCorDiax80® | Theranova 500® | FxCorDiax 1000® |
|---|---|---|---|
Membrane | Helixone | Polyarylethersulfone and polyvinyl pyrrolidone blend | Helixone |
| Membrane surface area, m2 | 1.8 | 2 | 2.3 |
| KUF, mL/h/mmHg | 64 | 59 | 75 |
| Inner diameter of hollow fibre, µm | 185 | 180 | 210 |
| Membrane thickness, µm | 35 | 35 | 35 |
| Sterilization | Steam | Steam | Steam |
| β2-microglobulin Sc | 0.9 | 1 | 0.9 |
KUF, ultrafiltration coefficient; Sc, sieving coefficient.
Baseline characteristics of the study population
| General characteristics | |
|---|---|
| Sex, male, % | 92 |
| Age, years | 60.1 ± 4.6 |
| CKD aetiology, % | |
| Diabetes | 15.4 |
| Vascular | 7.7 |
| Glomerular | 30.8 |
| Loss of renal mass | 15.4 |
| Others | 30.8 |
| Dialysis vintage, months | 54.7 ± 40.1 |
| Dry weight, kg | 71.7 ± 5.2 |
| Laboratory parameters | |
| Haemoglobin, g/dL | 11.1 ± 1.13 |
| Platelets, 103/mcl | 174.4 ± 58.1 |
| Serum proteins levels, g/dL | 6.5 ± 0.4 |
Data are presented as mean ± SD, or %. CKD, chronic kidney disease.
Coagulation tests and clotting parameters
| HF-HD | HDx | HDF | P-value | |
|---|---|---|---|---|
| Pre-dialysis APTT, s | 27.7 ± 1.95 | 27.5 ± 0.6 | 27.7 ± 1.6 | NS |
| Pre-dialysis aXa, U/mL | 0.026± 0.016 | 0.038 ± 0.034 | 0.039 ± 0.008 | NS |
| Post-dialysis APTT, s | 30.5 ± 0.7 | 28.2 ± 0.64 | 28.8 ± 0.73 | 0.009 |
| 0.009 | ||||
| Post-dialysis aXa, U/mL | 0.26 ± 0.02 | 0.22 ± 0.01 | 0.21 ± 0.02 | 0.024 |
| 0.05 | ||||
| Time to clotting, min | 14.2 ± 1.2 | 17.1 ± 1.6 | 13.9 ± 1.6 | NS |
| Dialyser clotting score: ‘Clean’, | 11 (84.6) | 8 (72.7) | 5 (45.5) | NS |
| ‘Medium’, | 2 (15.4) | 3 (27.3) | 6 (54.5) | NS |
Data are presented as mean ± SD or n (%). NS, not significant; T-Wilcoxon for paired groups with Holm correction for multiple comparisons.
HF-HD versus HDF.
HF-HD versus HDx.