Stress analyses of glenoid component designs.

T E Orr1, D R Carter, D J Schurman.   

Abstract

Metal backing of glenoid components for total shoulder replacements and the use of bony ingrowth surfaces on these components have recently been introduced. In this study, finite element analyses were performed to determine the stress fields in the natural glenoid and to calculate the change in bone stresses after implantation of glenoid components of various designs. The effects of metal backing, keel geometry, and superior constraint on bone stresses indicate that stress distributions on the natural glenoid corresponded to bone morphology. Metal-backing the glenoid component may cause slight improvement in stress transfer to cortical bone. Altered fin geometry better stabilized the glenoid component. Superior restraints on the component intending to prevent subluxation increase stresses and may cause earlier loosening than encountered with unconstrained components.

Entities:  

Mesh:

Year:  1988        PMID: 3383490

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  7 in total

1.  Finite element analysis of the strain distribution in the humeral head tubercles during abduction: comparison of young and osteoporotic bone.

Authors:  Ph Clavert; M Zerah; J Krier; P Mille; J F Kempf; J L Kahn
Journal:  Surg Radiol Anat       Date:  2006-08-26       Impact factor: 1.246

Review 2.  [Anatomical total shoulder replacement in glenohumeral osteoarthritis : Indications, current implants, and clinical results].

Authors:  O Lorbach
Journal:  Orthopade       Date:  2018-05       Impact factor: 1.087

Review 3.  Rotator cuff deficient arthritis of the glenohumeral joint.

Authors:  Alec A Macaulay; R Michael Greiwe; Louis U Bigliani
Journal:  Clin Orthop Surg       Date:  2010-11-05

4.  New design of a cementless glenoid component in unconstrained shoulder arthroplasty: a prospective medium-term analysis of 143 cases.

Authors:  D Katz; J Kany; P Valenti; P Sauzières; P Gleyze; K El Kholti
Journal:  Eur J Orthop Surg Traumatol       Date:  2012-10-27

5.  No correlation between radiolucency and biomechanical stability of keeled and pegged glenoid components.

Authors:  Andreas Voss; Knut Beitzel; Elifho Obopilwe; Stefan Buchmann; John Apostolakos; Jessica Di Venere; Michael Nowak; Mark P Cote; Anthony A Romeo; Augustus D Mazzocca
Journal:  BMC Musculoskelet Disord       Date:  2017-05-25       Impact factor: 2.362

6.  Three-Dimensional Morphometric Analysis of Penetrative Depth and Size of Nonarthritic and Degenerative Arthritic Glenoids: Implications for Glenoid Replacement in Shoulder Arthroplasty.

Authors:  Hong Jun Jung; Tae-Seok Nam; Daewon Park; In-Ho Jeon
Journal:  Clin Orthop Surg       Date:  2020-05-14

7.  The study of 2-dimensional computed tomography scans of the glenoid anatomy in relation to reverse shoulder arthroplasty in the Southern Chinese population.

Authors:  Amanda Mun Yee Slocum; Yuk Chuen Siu; Chun Man Ma; Tun Hing Lui
Journal:  JSES Int       Date:  2021-04-03
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.