Conventional stylet-driven leads with extendable helix can be implanted successfully for Left Bundle Branch Area Pacing (LBBAP) with a low acute complication rate. We report 2 cases in which lead repositioning after a first unsuccessful attempt to LBBAP was associated with fracture of the helix rotating mechanism and failure to fully extract the pacing lead. This article is protected by copyright. All rights reserved. This article is protected by copyright. All rights reserved.Entities:
Keywords: Complication; His bundle pacing; Left bundle area pacing; Physiologic pacing
Year: 2021 PMID: 33825263 DOI: 10.1111/jce.15030