Literature DB >> 33823619

Semi-automated cerebral aneurysm segmentation and geometric analysis for WEB sizing utilizing a cloud-based computational platform.

Ansaar T Rai1, Ryan G Brotman2, Gerald R Hobbs3, SoHyun Boo1.   

Abstract

BACKGROUND: Accurate aneurysm measurements are important for selecting the WEB device. The objective was to validate a cloud-based platform, SurgicalPreview (SP) against manual measurements for aneurysm analysis.
METHODS: Two sets of measurements each for SP and manual methods were obtained for 40 aneurysms. Reliability and agreement were assessed with intra-class correlation coefficient (ICC) and Bland-Altman plots respectively. Kappa coefficient was used to assess agreement for predicting WEB size.
RESULTS: There was good reliability for repeat SP measurements: aneurysm diameter (ICC-1, 95%CI 0.98-1), height (ICC-1, 95%CI 0.99-1) and neck diameter (ICC-0.96, 95%CI 0.93-0.98). There was good reliability for the two manual diameter (ICC-0.97, 95%CI 0.9-0.97) and height (ICC-0.93, 95%CI 0.87-0.96) measurements and moderate for neck diameter (ICC-0.76, 95%CI 0.54-0.87). There was greater agreement for SP versus manual repeat measurements on Bland-Altman plots. Reliability between the SP and manual methods was good for aneurysm diameter (ICC-0.98, 95%CI 0.95-1) and height (ICC-0.96, 95%CI-0.93-0.98) and moderate for neck. (ICC-0.6, 95%CI -0.22-0.87). The Bland-Altman plots confirmed better agreement between the two methods for the aneurysm diameter and height than the neck. There was strong agreement between the methods for predicting the WEB diameter (Kappa-0.84, 95%CI 0.71-0.97) and moderate for predicting WEB height (Kappa-0.66, 95%CI 0.43-0.89). There was moderate agreement for predicted versus deployed WEB diameter: SP (Kappa-0.56, 95%CI 0.38-0.74), Manual (Kappa-0.53, 95%CI 0.34-0.71).
CONCLUSION: The SurgicalPreview® had greater agreement for repeat measurements. There was good reliability between the two methods for predicting WEB diameter and height and moderate agreement between predicted versus deployed WEB diameter.

Entities:  

Keywords:  WEB; computation; sizing

Mesh:

Year:  2021        PMID: 33823619      PMCID: PMC8673900          DOI: 10.1177/15910199211009111

Source DB:  PubMed          Journal:  Interv Neuroradiol        ISSN: 1591-0199            Impact factor:   1.610


  18 in total

1.  Cerebrovascular segmentation from TOF using stochastic models.

Authors:  M Sabry Hassouna; A A Farag; Stephen Hushek; Thomas Moriarty
Journal:  Med Image Anal       Date:  2006-02       Impact factor: 8.545

2.  Comparison of Pipeline Embolization Device Sizing Based on Conventional 2D Measurements and Virtual Simulation Using the Sim&Size Software: An Agreement Study.

Authors:  J M Ospel; G Gascou; V Costalat; L Piergallini; K A Blackham; D W Zumofen
Journal:  AJNR Am J Neuroradiol       Date:  2019-02-07       Impact factor: 3.825

3.  Predicting Factors of Angiographic Aneurysm Occlusion after Treatment with the Woven EndoBridge Device: A Single-Center Experience with Midterm Follow-Up.

Authors:  F Cagnazzo; R Ahmed; R Zannoni; C Dargazanli; P-H Lefevre; G Gascou; I Derraz; C Riquelme; A Bonafe; V Costalat
Journal:  AJNR Am J Neuroradiol       Date:  2019-09-19       Impact factor: 3.825

4.  Physician training protocol within the WEB Intrasaccular Therapy (WEB-IT) study.

Authors:  Adam Arthur; Daniel Hoit; Alexander Coon; Josser E Delgado Almandoz; Lucas Elijovich; Saruhan Cekirge; David Fiorella
Journal:  J Neurointerv Surg       Date:  2017-08-16       Impact factor: 5.836

5.  The Contour-Early Human Experience of a Novel Aneurysm Occlusion Device.

Authors:  P Bhogal; I Lylyk; J Chudyk; N Perez; C Bleise; P Lylyk
Journal:  Clin Neuroradiol       Date:  2020-01-28       Impact factor: 3.649

6.  Cerebrovascular geometry in the anterior circulation: an analysis of diameter, length and the vessel taper.

Authors:  Ansaar T Rai; Jeffery P Hogg; Brendan Cline; Gerald Hobbs
Journal:  J Neurointerv Surg       Date:  2012-04-04       Impact factor: 5.836

7.  Dimensions of the posterior cerebral circulation: an analysis based on advanced non-invasive imaging.

Authors:  Ansaar T Rai; Daniel Rodgers; Eric A Williams; Jeffery P Hogg
Journal:  J Neurointerv Surg       Date:  2012-11-23       Impact factor: 5.836

8.  Comparison of operating room variables, radiation exposure and implant costs for WEB versus stent assisted coiling for treatment of wide neck bifurcation aneurysms.

Authors:  Ansaar T Rai; Ryan C Turner; Ryan G Brotman; SoHyun Boo
Journal:  Interv Neuroradiol       Date:  2020-10-20       Impact factor: 1.610

9.  A vessel segmentation method for multi-modality angiographic images based on multi-scale filtering and statistical models.

Authors:  Pei Lu; Jun Xia; Zhicheng Li; Jing Xiong; Jian Yang; Shoujun Zhou; Lei Wang; Mingyang Chen; Cheng Wang
Journal:  Biomed Eng Online       Date:  2016-11-08       Impact factor: 2.819

10.  Clinical and Anatomical Follow-up in Patients With Aneurysms Treated With the WEB Device: 1-Year Follow-up Report in the Cumulated Population of 2 Prospective, Multicenter Series (WEBCAST and French Observatory).

Authors:  Laurent Pierot; Laurent Spelle; Andrew Molyneux; James Byrne
Journal:  Neurosurgery       Date:  2016-01       Impact factor: 4.654

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.