Taiji Nakano, Akira Nakajima, Hiroto Watanabe, Ayaka Osada, Yasuhiro Namura, Takayuki Yoneyama, Eiji Tanaka, Mitsuru Motoyoshi.
Abstract
OBJECTIVES: To examine the torque moment that occurs between esthetic brackets and bendable alloy (stainless steel [SS], titanium-molybdenum [Ti-Mo], and titanium-niobium [Ti-Nb]) wires.
MATERIALS AND METHODS: This study examined ceramic (CR), zirconium oxide (ZC), polycarbonate (PC), and conventional metallic brackets (MT) (upper, 0.018-inch and 0.022-inch slots) combined with SS, Ti-Mo, and Ti-Nb wires using elastic module ligation. The torque moments delivered by various wire and bracket combinations were measured using a torque gauge apparatus. The wire torque angles at 5-40° were examined.
RESULTS: The torque value increased in the order of CR, ZC, MT, and PC brackets for both 0.018-inch and 0.022-inch slots. The fracture points of the CR and ZC brackets combined with SS and Ti-Mo wires were approximately more than 30° and 35°, respectively. No fracture points were detected in the combination of ZC brackets and Ti-Nb wires.
CONCLUSIONS: The current study identified the material characteristics of CR, ZR, and PC brackets during torque tooth movements. The present results demonstrate a characteristic combined effect between different esthetic brackets and bendable alloy wires. © 0000 by The EH Angle Education and Research Foundation, Inc.
OBJECTIVES: To examine the torque moment that occurs between esthetic brackets and bendable alloy (stainless steel [SS], titanium-molybdenum [Ti-Mo], and titanium-niobium [Ti-Nb]) wires.
MATERIALS AND METHODS: This study examined ceramic (CR), zirconium oxide (ZC), polycarbonate (PC), and conventional metallic brackets (MT) (upper, 0.018-inch and 0.022-inch slots) combined with SS, Ti-Mo, and Ti-Nb wires using elastic module ligation. The torque moments delivered by various wire and bracket combinations were measured using a torque gauge apparatus. The wire torque angles at 5-40° were examined.
RESULTS: The torque value increased in the order of CR, ZC, MT, and PC brackets for both 0.018-inch and 0.022-inch slots. The fracture points of the CR and ZC brackets combined with SS and Ti-Mo wires were approximately more than 30° and 35°, respectively. No fracture points were detected in the combination of ZC brackets and Ti-Nb wires.
CONCLUSIONS: The current study identified the material characteristics of CR, ZR, and PC brackets during torque tooth movements. The present results demonstrate a characteristic combined effect between different esthetic brackets and bendable alloy wires. © 0000 by The EH Angle Education and Research Foundation, Inc.
Entities:
Keywords:
Bendable wire; Dental materials; Esthetic bracket; Fracture point; Third-order bend; Torque moment
Year: 2021
PMID: 33822872 DOI: 10.2319/081820-729.1
Source DB: PubMed Journal: Angle Orthod ISSN: 0003-3219 Impact factor: 2.079