| Literature DB >> 33817994 |
Daniel M Gleghorn1, Orisadare Olayiwola1, Anisa Omer1, Roderick Lynn1, David Norcliffe2, Mark Robinson1.
Abstract
Entities:
Keywords: COVID-19; hyperinflammatory; plasma viscosity
Mesh:
Year: 2021 PMID: 33817994 PMCID: PMC8251075 DOI: 10.1111/ijlh.13532
Source DB: PubMed Journal: Int J Lab Hematol ISSN: 1751-5521 Impact factor: 3.450
FIGURE 1Receiver operating characteristic (ROC) curve analysis provided an area under curve (AUC) of 0.914 (95% CI: 0.89‐0.94)
Plasma viscosity results between the greatest sensitivity and specificity for COVID‐19. This includes confidence intervals (CI), likelihood ratios and supplementary Youden's index used as the discriminator
| Plasma viscosity (mPas) | Sensitivity (95% CI) | Specificity (95% CI) | Likelihood ratio (+) | Likelihood ratio (−) | Youden's index |
|---|---|---|---|---|---|
| 1.66 | 100% (97.9‐100) | 66.9% (60.4‐73.1) | 171 | 150 | 0.670 |
| 1.67 | 99.4% (96.8‐99.9) | 66.9% (60.4‐73.1) | 170 | 150 | 0.664 |
| 1.68 | 94.7% (90.2‐97.6) | 66.9% (60.4‐73.1) | 162 | 150 | 0.617 |
| 1.69 | 91.2% (85.9‐95.0) | 66.9% (60.4‐73.1) | 156 | 150 | 0.582 |
| 1.70 | 88.3% (82.5‐92.7) | 66.9% (60.4‐73.1) | 151 | 150 | 0.553 |
| 1.71 | 84.2% (77.9‐89.3) | 66.9% (60.4‐73.1) | 144 | 150 | 0.512 |
| 1.72 | 82.5% (75.9‐87.8) | 66.9% (60.4‐73.1) | 141 | 150 | 0.494 |
| 1.73 | 80.7% (73.9‐86.3) | 69.2% (62.7‐75.2) | 138 | 155 | 0.499 |
| 1.74 | 76.0% (68.9‐82.2 | 75.5% (69.3‐80.9) | 130 | 169 | 0.515 |
| 1.75 | 74.3% (67.0‐80.6) | 79.0% (73.1‐84.2) | 127 | 177 | 0.533 |
| 1.76 | 73.1% (65.8‐79.6) | 83.5% (77.9‐88.1) | 125 | 187 | 0.566 |
| 1.77 | 72.5% (65.2‐79.1) | 87.1% (81.9‐91.2) | 124 | 195 | 0.596 |
| 1.78 | 71.9% (64.6‐78.5) | 90.2% (85.5‐93.7) | 123 | 202 | 0.621 |
| 1.79 | 71.4% (63.9‐77.9) | 91.1% (86.6‐94.5) | 122 | 204 | 0.624 |
| 1.80 | 70.8% (63.3‐77.5) | 91.1% (86.6‐94.5) | 121 | 204 | 0.618 |
| 1.81 | 70.2% (62.7‐76.9) | 93.3% (89.2‐96.2) | 120 | 209 | 0.635 |
| 1.82 | 69.6% (62.1‐76.4) | 95.1% (91.4‐97.5) | 119 | 213 | 0.647 |
|
|
|
|
|
|
|
| 1.84 | 67.6% (59.7‐74.2) | 100% (98.4‐100) | 115 | 224 | 0.673 |
The optimum cut‐off PV of 1.83mPas is highlighted in bold.