This work examines the stable isotope fractionation of carbon and oxygen in gaseous, supercritical, and liquid carbon dioxide systems at temperatures from -27.1 to +43.5 °C. For pressurized single-, supercritical-, and dual-phase carbon dioxide, both carbon and oxygen isotope fractionations can be measured and are significant when subjected to variations within this temperature range. The δ 13C and δ 18O values ranged from -41.55 to -41.38 ‰ (VPDB) and -27.74 to -24.9 ‰ (VPDB), respectively, for gas-phase carbon dioxide from 9.3 to 39 °C. A pressure variation of 27.58 barg to 34.48 barg was measured throughout this temperature range. In order to evaluate the effect of supercritical formation and liquefaction on the stable isotope values, cylinders were filled to varying pressures. When stored at cold temperatures, the δ13C value as measured in the headspace of the liquid phase varied from -41.23 to -41.13 ‰ (VPDB) and -41.50 to -41.44 ‰ (VPDB) in the supercritical phase. The δ18O value was between -25.51 and -25.36 ‰ (VPDB) in the liquid phase and between -24.79 and -24.77 ‰ (VPDB) in the supercritical phase. Temperatures in these experiments were selected to mimic outdoor conditions (winter and summer) that stable isotope laboratory practitioners may encounter when storing compressed carbon dioxide cylinders containing stable isotope working reference gases. The carbon and oxygen isotope composition of carbon dioxide gas within these pressurized cylinders return to their precooled isotope values within ∼24 h when warmed to laboratory temperatures (∼24 °C). A headspace analysis performed immediately after the carbon dioxide cylinder was removed from the cold environment yielded δ13C values that were relatively enriched, while δ18O values were relatively depleted. This is likely an effect of 12C and 18O being preferentially partitioned in the liquid phase within the cylinder. As the cylinder warmed, both liquid and gas equilibrated, and carbon and oxygen homogenized isotopically. As the cylinder was heated into the supercritical phase, a slight opposite isotope effect at higher pressure and temperatures was noted. That is, a slight 13C depletion and 18O enrichment were observed in the gas phase. However, these isotope variations were just slightly outside of the analytical error. Additionally, a separate gas-phase carbon dioxide cylinder was kept at a constant laboratory temperature as a control. This carbon dioxide showed no measurable carbon or oxygen isotope variation throughout the duration of the experimental work. The measured isotope fractionation was significantly higher comparing the phase transition from the gaseous to liquid phase versus the gaseous phase to supercritical phase. The proper handling of pressurized carbon dioxide cylinders used as reference gases for an isotope ratio mass spectrometer includes using carbon dioxide at pressures of less than ∼34.88 barg to ensure that the gas is present as a single phase, storing the gas in a temperature-controlled environment, and allowing the gaseous carbon dioxide to equilibrate to ambient conditions for 24-48 h if storage in a controlled ambient environment is not feasible.
This work examines the stable isotope fractionation of carbon and oxygen in gaseous, supercritical, and liquid carbon dioxide systems at temperatures from -27.1 to +43.5 °C. For pressurized single-, supercritical-, and dual-phase carbon dioxide, both carbon and oxygen isotope fractionations can be measured and are significant when subjected to variations within this temperature range. The δ 13C and δ 18O values ranged from -41.55 to -41.38 ‰ (VPDB) and -27.74 to -24.9 ‰ (VPDB), respectively, for gas-phase carbon dioxide from 9.3 to 39 °C. A pressure variation of 27.58 barg to 34.48 barg was measured throughout this temperature range. In order to evaluate the effect of supercritical formation and liquefaction on the stable isotope values, cylinders were filled to varying pressures. When stored at cold temperatures, the δ13C value as measured in the headspace of the liquid phase varied from -41.23 to -41.13 ‰ (VPDB) and -41.50 to -41.44 ‰ (VPDB) in the supercritical phase. The δ18O value was between -25.51 and -25.36 ‰ (VPDB) in the liquid phase and between -24.79 and -24.77 ‰ (VPDB) in the supercritical phase. Temperatures in these experiments were selected to mimic outdoor conditions (winter and summer) that stable isotope laboratory practitioners may encounter when storing compressed carbon dioxide cylinders containing stable isotope working reference gases. The carbon and oxygen isotope composition of carbon dioxide gas within these pressurized cylinders return to their precooled isotope values within ∼24 h when warmed to laboratory temperatures (∼24 °C). A headspace analysis performed immediately after the carbon dioxide cylinder was removed from the cold environment yielded δ13C values that were relatively enriched, while δ18O values were relatively depleted. This is likely an effect of 12C and 18O being preferentially partitioned in the liquid phase within the cylinder. As the cylinder warmed, both liquid and gas equilibrated, and carbon and oxygen homogenized isotopically. As the cylinder was heated into the supercritical phase, a slight opposite isotope effect at higher pressure and temperatures was noted. That is, a slight 13C depletion and 18O enrichment were observed in the gas phase. However, these isotope variations were just slightly outside of the analytical error. Additionally, a separate gas-phase carbon dioxide cylinder was kept at a constant laboratory temperature as a control. This carbon dioxide showed no measurable carbon or oxygen isotope variation throughout the duration of the experimental work. The measured isotope fractionation was significantly higher comparing the phase transition from the gaseous to liquid phase versus the gaseous phase to supercritical phase. The proper handling of pressurizedcarbon dioxide cylinders used as reference gases for an isotope ratio mass spectrometer includes using carbon dioxide at pressures of less than ∼34.88 barg to ensure that the gas is present as a single phase, storing the gas in a temperature-controlled environment, and allowing the gaseous carbon dioxide to equilibrate to ambient conditions for 24-48 h if storage in a controlled ambient environment is not feasible.
With the advent and growth of continuous
flow-isotope ratio mass
spectrometry (CF-IRMS) techniques over the past few decades,[1] there has been an increase in the use of stable
isotope analysis in areas such as food authentication and environmental
research.[2] At the heart of this technology
is the requirement for reliable and stable working IRMS reference
gas standards. IRMS reference gases are typically contained in high-pressure
compressed gas cylinders and connected directly to the IRMS (or an
offline aliquot is taken) for use as working gases for direct comparison
to unknown samples during stable isotope measurements in either CF-IRMS
or dual inlet-IRMS modes. Furthermore, some of these cylinders can
contain more than one physical phase at common storage temperatures
depending on the pressure and temperature (i.e., gas, supercritical,
and liquid), such as carbon dioxide (CO2) and nitrous oxide
(N2O). However, over the decades, little attention has
been paid to the behavior of these multiphase working gases in the
compressed state, particularly with respect to their potential isotope
fractionation at variable temperatures when they are sampled as a
gas phase for stable isotope analysis. In order to better understand
this behavior, a series of experiments were conducted using single-
and dual-phase compressed gas cylinders (30 L) of CO2.
Carbon dioxide within compressed gas cylinders remains an important
anchor for isotope ratio measurements of carbon- and oxygen-bearing
compounds, and thus, understanding the stable isotope composition
of CO2 reference gas over a range of environmental conditions
is paramount to the operation of a scientifically credible stable
isotope laboratory.[3]The physical
state of any substance is governed by both its pressure
and temperature. An understanding of the relationships thus provides
the pressure and temperature for phase transitions. For carbon dioxide,
the critical point occurs at 30.98 °C and 73.8 barg[4] (barg or bar gauge pressure is the pressure in
bars above ambient pressure). Above this temperature and pressure,
all of the carbon dioxide will be present in the supercritical phase,
that is, an intermediate phase between the gas and liquid states.
Above the critical point, a gas cannot be liquefied, no matter the
pressure applied.[4] Like a gas, a supercritical
fluid will expand to fill the cylinder; however, it is important to
note that the density of the supercritical state is significantly
higher than that of the gas phase.[5]Grootes et al.[6] focused on the equilibrium
fractionation of carbon and oxygen isotopes in the CO2 (l)–CO2(g) system. They showed that at isotopic equilibrium, 18O tends to be preferentially enriched in the liquid phase
(CO2(l)) and depleted in the gas phase (CO2(g)), while 13C tends to be enriched in the gas phase (CO2(g)) and depleted in the liquid phase (CO2(l)).
The objectives for this work were to design and execute experiments
to determine if carbon- or oxygen-isotope fractionation occurs when
sampling CO2 gas from single-, supercritical-, and dual-phase
compressed gas cylinders stored at different temperatures. The range
of temperatures was selected to represent conditions that may be experienced
as stable isotope laboratory practitioners move stored CO2 cylinders earmarked for isotope reference use from an outdoor facility
to their indoor laboratories. These results provide a basis for recommending
proper cylinder handling and storage procedures for CO2 reference gas for stable isotope laboratory use. Results from these
experiments clearly indicate that certain precautions should be taken
when dealing with dual-, supercritical-, and single-phase compressed
CO2 cylinders being used as IRMS working laboratory isotope
reference gases.
Results and Discussion
CO2 Cylinder
Temperature and Pressure Variations
All data for samples
taken from the three cylinders are shown in Tables –3. The CO2 gas showed
systematic and measurable pressure changes when pressurized cylinders
were exposed to temperature fluctuations. The temperature vs time
graph for all three cylinders is illustrated in Figure . Prior to the beginning of the experiment,
all three cylinders were stored in the laboratory for ∼4 weeks
to ensure thermal equilibrium. The temperature of all three cylinders
was 24.3 °C; cylinder A had a pressure of 34.48 barg, and cylinders
B and C had a pressure of 57.22 barg. While in the freezer (−26.3
°C), the pressure of cylinder A dropped to 15.86 barg. Once the
cylinder was removed from the freezer, the pressure gradually rose
to 27.58 barg at 9.3 °C and eventually leveled out to 34.48 barg
when warmed to ambient laboratory temperatures. When placed outdoors,
the temperature reached 39.9 °C and the pressure remained at
∼34.48 barg.
Table 1
Summary of CO2 Cylinder
“a” Experimental Data: BG = Background Temperature,
Pressure, δ13C and δ18O Composition
of Cylinder “A’s” CO2 before the Beginning
of the Experiment
cylinder A
time (h)
T °C
δ13CVPDB ‰
δ18OVPDB ‰
P (barg)
state
BG
24.3
–41.54
–24.77
34.48
gas
0
–26.3
–41.34
–25.07
15.86
gas/liquid
2
9.3
–41.52
–24.79
27.58
gas
24
23.0
–41.54
–24.90
34.48
gas
49
24.4
–41.55
–24.80
34.48
gas
144
23.9
–41.53
–24.78
34.48
gas
480
23.7
–41.51
–24.80
34.48
gas
888
24.1
–41.47
–24.82
34.48
gas
1344
34.5
–41.41
–24.77
34.48
gas
1536
39.4
–41.43
–24.80
34.48
gas
1550
38.7
–41.43
–24.87
34.48
gas
1574
39.9
–41.44
–24.89
34.48
gas
2320
39.0
–41.38
–24.74
34.48
gas
Table 3
Summary of CO2 Cylinder
“C” Experimental Data: BG = Background Temperature,
Pressure, δ13C and δ18O Composition
of Cylinder “C’s” CO2a
cylinder
C
time (h)
T °C
δ13CVPDB ‰
δ18OVPDB ‰
P (barg)
state
BG
24.3
–41.45
–25.92
57.22
gas
0
23.6
–41.41
–24.83
57.22
gas
2
23.8
–41.40
–24.96
57.22
gas
23
24.3
–41.40
–24.90
57.22
gas
44
22.5
–41.41
–24.95
57.22
gas
143
23.4
–41.39
–24.93
57.22
gas
480
23.7
–41.36
–24.95
57.22
gas
888
23.8
–41.30
–24.99
57.22
gas
1344
23.8
–41.30
–25.03
57.22
gas
1538
23.6
–41.34
–25.02
57.22
gas
1550
23.3
–41.30
–25.01
57.22
gas
1574
23.3
–41.32
–24.96
57.22
gas
2324
22.4
–41.31
–24.96
57.22
gas
This Cylinder Was
Subjected to neither
Cold nor Heat and Remained within the Ambient Conditions of the Laboratory
throughout the Duration of the Experiment.
Figure 1
Temperature vs time for cylinders A, B, and C. Background
temperatures
were measured prior to the beginning of the experiments. While in
the freezer, the pressure of the CO2 in cylinders A and
B dropped to a low of 15.86 barg. The pressure in cylinder A gradually
rose to 27.58 barg at 9.3 °C and eventually leveled out to 34.48
barg when warmed to laboratory temperature and when placed outdoors
in the summer heat (up to 39.9 °C). Cylinder B behaved similarly
to A; however, due to the higher starting pressure and slightly higher
outdoor temperature (43.5 °C), for cylinder B, the pressure rose
to 93.08 barg. Cylinder C saw virtually no temperature change from
the background throughout the duration of the experiment.
Temperature vs time for cylinders A, B, and C. Background
temperatures
were measured prior to the beginning of the experiments. While in
the freezer, the pressure of the CO2 in cylinders A and
B dropped to a low of 15.86 barg. The pressure in cylinder A gradually
rose to 27.58 barg at 9.3 °C and eventually leveled out to 34.48
barg when warmed to laboratory temperature and when placed outdoors
in the summer heat (up to 39.9 °C). Cylinder B behaved similarly
to A; however, due to the higher starting pressure and slightly higher
outdoor temperature (43.5 °C), for cylinder B, the pressure rose
to 93.08 barg. Cylinder C saw virtually no temperature change from
the background throughout the duration of the experiment.This Cylinder Was
Subjected to neither
Cold nor Heat and Remained within the Ambient Conditions of the Laboratory
throughout the Duration of the Experiment.While in the freezer (−27.1 °C), the pressure
in cylinder
B dropped to 15.85 barg, like cylinder A and cylinder B following
the same pattern with respect to temperature and pressure when removed
from the freezer (Figure ). After 2 h in the warm laboratory ambient environment, cylinder
B warmed to −0.6 °C and the pressure rose to 34.48 barg.
However, unlike the lower-pressure cylinder A, this higher-pressure
cylinder rose to much higher pressures when placed outdoors in the
heat of the summer (up to ∼93.00 barg at 43.5 °C). It
should be noted that the initial pressure was 57.22 barg for cylinder
B and 34.48 barg for cylinder A. Cylinder B experienced a slightly
higher outdoor temperature (43.5 °C for cylinder B vs 39.9 °C for cylinder A).
CO2 Carbon and
Oxygen Isotope Variations
The CO2 gas showed systematic
and measurable isotope variations
when cylinders A and B were exposed to temperature fluctuations. The
δ13C value of CO2 in cylinder A (gas only)
increased, while the δ18O value decreased when the
cylinder was immediately removed from the freezer (Figures and 3). Once the cylinder at 24.3 °C and 34.98 barg (cylinder A)
was placed in the freezer at −26.3 °C, the pressure of
the cylinder decreased to 15.86 barg and liquefied,[5] as seen in cylinder A. We interpret these trends in carbon
and oxygen isotopes as an effect of 18O and 12C being concentrated within the liquid phase and conversely depleted
in the gas phase (initially). Like cylinder A, CO2 in cylinder
B also showed an increase in the δ13C value and a
decrease in the δ18O value when first sampled and
removed from the freezer. As with cylinder A, initially the 18O and 12C are being concentrated within the liquid phase
and conversely depleted in the gas phase. Therefore, when the headspace
CO2 sample was taken from the cold cylinder, it tends to
be fractionated (13C enriched and 18O depleted).
Essentially the liquid is being sequestered when cold where an isotope
fractionation is observed. During this liquid sequestration, the gas,
which is being sampled as headspace from the cylinder, tends to be
more enriched in 13C and depleted in 18O. These
observations are consistent with those seen by Grootes et al.[6] Unlike cylinder A, however, cylinder B, which
contained higher pressure, shows a greater degree of isotope fractionation
in both 13C and 18O. As cylinders A and B warmed
to room temperature over ∼24 h, both carbon and oxygen homogenized
isotopically and returned to their initial isotopic values. In a compressed
gas cylinder, the molecules of each component within the gas mixture
are constantly in motion, according to the kinetic theory of gases,
which essentially helps to maintain the homogeneity of the mixture.[7]
Figure 2
Change in δ13C vs time relative to VPDB
for CO2 cylinders A, B, and C. The CO2 gas showing
a clear 13C enrichment in both cylinders A and B when first
removed
from the freezer, an indication that CO2 liquid is being
sequestered while cold. Cylinder B, being dual phase, and thus higher
pressure, also shows a 13C enrichment in the gas samples
when initially removed from the freezer, a clear indication of liquid
sequestration of the CO2 while cold. During heating, the
trend seen during cooling may be slightly reversed. Cylinder C was
subjected to neither cold nor heat and shows no 13C variations.
Figure 3
Change in δ18O vs time relative
to
VPDB for CO2 cylinders A, B, and C. The CO2 gas
showing a clear 18O depletion in both cylinders A and B
when first removed from the freezer, an indication that CO2 liquid is being sequestered while cold. Cylinder B, being dual phase,
and thus higher pressure, also shows a 18O depletion in
the gas samples when initially removed from the freezer, a clear indication
of liquid sequestration of the CO2 while cold. During heating,
the trend seen during cooling may be slightly reversed. Cylinder C
was subjected to neither cold nor heat and shows no 18O
variations.
Change in δ13C vs time relative to VPDB
for CO2 cylinders A, B, and C. The CO2 gas showing
a clear 13C enrichment in both cylinders A and B when first
removed
from the freezer, an indication that CO2 liquid is being
sequestered while cold. Cylinder B, being dual phase, and thus higher
pressure, also shows a 13C enrichment in the gas samples
when initially removed from the freezer, a clear indication of liquid
sequestration of the CO2 while cold. During heating, the
trend seen during cooling may be slightly reversed. Cylinder C was
subjected to neither cold nor heat and shows no 13C variations.Change in δ18O vs time relative
to
VPDB for CO2 cylinders A, B, and C. The CO2 gas
showing a clear 18O depletion in both cylinders A and B
when first removed from the freezer, an indication that CO2 liquid is being sequestered while cold. Cylinder B, being dual phase,
and thus higher pressure, also shows a 18O depletion in
the gas samples when initially removed from the freezer, a clear indication
of liquid sequestration of the CO2 while cold. During heating,
the trend seen during cooling may be slightly reversed. Cylinder C
was subjected to neither cold nor heat and shows no 18O
variations.After continued periodic measurement
of the CO2 in both
cylinders A and B for 846 h (36 days), the carbon and oxygen isotope
composition remained constant, within analytical error, that is, ±0.05
‰ for δ 13C and ± 0.07 ‰ for δ 18O (Figures and 3). After this period, both cylinders
A and B were placed outside and heated for ∼984 h (∼41
days). Temperatures rose to a high of 39.9 °C for cylinder A
and 43.5 °C for cylinder B. It should be noted that the difference
in temperature between cylinders A and B while outdoors is likely
due to each cylinder’s exposure to direct sunlight and cloud
cover, while the temperature, pressure, and samples were being taken
from each. Furthermore, the temperature inside of the cylinder might
not have achieved uniform distribution during this period. Analyses
were conducted throughout this period. For heated cylinders (up to
43.5 °C for cylinder B), there may be a slight inverse isotope
effect at higher pressure and temperature, as opposed to that seen
in low temperature CO2. Both carbon and oxygen isotopes
in the CO2 gas in cylinder B tended to show a slight opposite
trend as that seen in the cold samples when they were heated outdoors
(13C depletion and an 18O enrichment); however,
these isotope variations are not significantly outside of the analytical
error of the overall measurements.Cylinder C, which was identical
to cylinder B in terms of filling
pressure, was used as a control by maintaining it in the laboratory
environment through the duration of the experiments. This CO2 cylinder was subjected to neither cold nor heat. The CO2 from cylinder C shows neither 13C enrichment nor 18O depletion (Figures and 3), within analytical error. Isotope
trends for both 13C and 18O are also consistent
with no liquid isotope fractionation (±0.05 ‰ for δ 13C and ± 0.07 ‰ for δ 18O).
Conclusions
The work described herein addresses understanding
the behavior
of the carbon and oxygen isotopes in pressurizedCO2 cylinders
when those cylinders are subjected to variations in temperatures that
may be encountered during normal earth surface conditions. This understanding
is essential since these compressed gases are used as isotope reference
materials for routine daily operation in stable isotope laboratories
worldwide and since isotope variations can occur due to temperature
variations when moving cylinders from, for example, cold outdoor conditions
to a warm indoor laboratory. Results from these experiments clearly
indicate that certain precautions should be taken when dealing with
both dual-phase and single-phase CO2 when used as a laboratory
isotope reference gas.This work shows that for CO2, isotope fractionations
can be significant and measurable. The carbon and oxygen isotope composition
of CO2 gas within these cylinders quickly returned to their
precooled values when warmed to laboratory temperatures. Initial cold
cylinder samples, when first removed from a freezer, tended to be
the most depleted in 18O and enriched in 13C.
We interpret these results as an effect of the 18O and 12C being concentrated within the liquid phase and consequently
depleted in the gas phase that was sampled. As the cylinder warmed
to laboratory temperature, both liquid and vapor equilibrated isotopically.
For cylinders that were heated outdoors up to +43.5 °C, we observed
a slight inverse isotope effect at higher pressure, as opposed to
that seen in low-temperature CO2.To eliminate these
isotope fractionations, we recommend the following
precautions be taken when pressurizedCO2 is used as a
laboratory isotope reference gas:Utilize single-phase CO2 (i.e., less than
∼34.88 barg).Store CO2 indoors, if possible.Always stabilize
gaseous CO2 cylinders to
laboratory temperature for at least 24–48 h if stored in either
cold or hot conditions.
Experimental Section
Three separate compressed gas
cylinders of CO2 were utilized in this experiment, as outlined
in Table . All cylinders
used in this study were 6061 aluminum alloys and had internal volumes
of 30 L (Luxfer, Riverside, CA). Ceodeux (Mount Pleasant, PA) brass
diaphragm valves were used. Research-grade carbon dioxide (99.999%)
was obtained from Airgas (Plumsteadville, PA). Three CO2 cylinders were utilized to determine the variation of δ13C and δ18O values
at different temperatures, pressures, and phases. Initial headspace
samples of CO2 gas were taken from all three cylinders
prior to the beginning of each experiment and analyzed for δ13C and δ18O. Measurements continued while
the cylinders were subjected to variable temperatures.
Table 4
Summary of CO2 Cylinder
Physical Conditions Used for this Experiment
cylinder
phase
room temperature
pressure
temperature
ranges
A
gas
34.48 barg
–27.1 to 39.9 °C
B
gas
57.22 barg
–27.1 to 43.5 °C
C
gas
57.22 barg
22.5
to 24.3 °C
Two single-phase CO2 cylinders with differing pressures
(cylinders A and B) were placed in a top loading freezer and cooled
for ∼192 h (∼8 days) at −27.1 °C, ensuring
that fractionation would occur as the cylinder contents came to thermal
equilibrium with the temperature of the freezer.[7,8] At
this temperature, only cylinder B contained liquefied CO2.[4] The two cylinders were subsequently
removed from the freezer and brought into the laboratory, and the
samples of the gas phase were immediately taken from the headspace
of each cylinder. The cylinders were then permitted to warm to laboratory
temperature (22.5° to 24.3 °C) over an 864 h (36-day) period.
These same two cylinders were then placed outdoors where temperatures
ranged to a maximum of between 39.9 and 43.5 °C, where the headspace
was also periodically sampled for another 984 h (41 days). A third
gas-phase CO2 cylinder (cylinder C) remained indoors as
a control at laboratory temperature throughout the duration of the
experiment.The temperature of the CO2 cylinder was
taken using
a digital infrared thermometer (Extech Instruments, Model 4250A) at
30.5 cm from the cylinder body until the temperature reading stabilized
(∼20 s.). The precision of the measurements was ±0.8 °C.Headspace CO2 samples were taken by flowing gas from
the cylinder at a low delivery pressure (∼0.25–0.34
barg) through a stainless steel gas aliquot assembly (∼5 mL
in sample volume) consisting of two valves that were connected to
the regulator on the cylinder being sampled (Figure ).[9] After flowing
the gas through the aliquot assembly with both valves 1 and 2 open
for ∼30 s, the valve furthest from the regulator was closed
(valve 2). Immediately thereafter, valve 1 was closed and the regulator
pressure was adjusted to a zero-outlet pressure. The cylinder valve
was subsequently closed, and the entire aliquot assembly was removed
from the regulator and connected to the inlet portion of the IRMS
by 1/4” compression connection (Swagelok, Solon, Ohio) for
isotope analysis. The volume between the 1/4” compression connection
and valve 2 on the gas aliquoter was evacuated by the inlet system
within the IRMS and then a few millimoles of CO2 gas was
expanded into the dual inlet bellows of the IRMS as matched by the
working reference gas.
Figure 4
Gas sample aliquot assembly for both capturing gas flow
from the
CO2 cylinders and transferring to the inlet of the dual
inlet IRMS for simultaneous carbon and oxygen stable isotope analyses.
Gas sample aliquot assembly for both capturing gas flow
from the
CO2 cylinders and transferring to the inlet of the dual
inlet IRMS for simultaneous carbon and oxygen stable isotope analyses.All gas samples were analyzed via a dual inlet
on a Thermo Scientific
253 Plus 10 kV IRMS. Carbon and oxygen isotopes are reported relative
to VPDB[10] with both internal and external
standard deviations of measurements typically better than ±0.05
‰ for δ 13C and ± 0.07 ‰ for δ 18O.
Table 2
Summary of CO2 Cylinder
“B” Experimental Data: BG = Background Temperature,
Pressure, δ13C and δ18O Composition
of Cylinder “B’s” CO2 before the Beginning
of the Experiment