Literature DB >> 33815535

Letter to the Editor on "Percutaneous Dilational Tracheostomy without Bronchoscopy: Placement is Safe, Reintervention is Not."

Micah T Long1, Melissa Murray2.   

Abstract

Entities:  

Year:  2020        PMID: 33815535      PMCID: PMC8008404     

Source DB:  PubMed          Journal:  Tanaffos        ISSN: 1735-0344


× No keyword cloud information.
To the Editors: We read with great interest the recent report entitled “Performing Percutaneous Dilational Tracheostomy Without Using Fiberoptic Bronchoscope”. This descriptive study reviewed 70 cases of the aforementioned procedure and concluded that it was efficient and safe (1). This is supported by prior work, (2,3) noting however that this technique is frequently preempted by any suspicion of difficult airway anatomy, for prior history of airway problems and for those patients who have had a previous tracheostomy. Thus, some authors have argued that bronchoscopic guidance improves safety (4). Nonetheless, due to the expanding performance of percutaneous dilational tracheostomy without bronchoscopy, some physicians may attempt to manipulate or replace newly placed tracheostomy tubes without bronchoscopy for problems such as cuff rupture. This may pose a severe risk to patients.

Summary

Percutaneous dilational tracheostomy is a safe approach to the surgical airway in critically ill adults and may be performed safely without bronchoscopic guidance. Manipulation of an existing, newly placed tracheostomy, however, mandates measures to increase safety due to the potential for airway edema. A postoperative patient suffered from a tracheostomy tube cuff rupture requiring tube exchange two days after percutaneous tracheostomy. The airway was remarkably swollen and rescuing a complication via direct or indirect laryngoscopy would likely have been impossible. To improve safety in case of tracheostomy exchange failure, we loaded an endotracheal tube onto the bronchoscope, entered the subglottic airway and performed the procedure under bronchoscopic visualization. This would enable orotracheal reintubation rescue via the bronchoscope in case of emergency. While initial percutaneous tracheostomy may be performed safely without bronchoscopy, when intervening on a recently placed tracheostomy it is crucial to enhance safety with bronchoscopy.

Case Report

Written consent was obtained from the patient to discuss the following case. We cared for a 69-year-old man with a past medical history of severe chronic obstructive pulmonary disease with chronic exertional dyspnea, coronary artery disease, prior transient ischemic attack, hypertension and active smoking. He presented to the hospital for management of a symptomatic type-IV thoracoabdominal aortic aneurysm. His postoperative course was complicated, including a massive pulmonary embolism and two extubation failures. An 8.0 cuffed Shiley tracheostomy tube was placed on post-procedure day 18 using a Ciaglia Blue Rhino® G2 advanced percutaneous tracheostomy introducer and tray (Cook Medical, Bloomington, IN) (5). Two days later the tracheostomy was slowly but continually leaking, causing derecruitment and hypoxia. Prior intubations were performed via indirect laryngoscopy showing easy views of normal anatomy, and the patient had no external neck swelling, edema or anatomic abnormalities. Nonetheless, bronchoscopic visualization of the airway was exquisitely challenging (Figures 1 & 2) and severe enough to have precluded direct or indirect laryngoscopy guided orotracheal reintubation of the airway in case of emergency. To improve safety in case of tracheostomy exchange failure, we loaded an endotracheal tube onto the bronchoscope, entered the subglottic airway and performed the procedure under bronchoscopic visualization. This would enable orotracheal reintubation rescue via the bronchoscope. Thankfully, the procedure was uncomplicated. It was discovered that the tracheostomy cuff had a small puncture hole.
Figure 1.

Initial bronchoscopic view of the glottis above the level of the vocal cords with significant edema and poor visualization of anatomic landmarks.

Figure 2.

Improved visualization of vocal cords with suctioning and jaw thrust. A) Left arytenoid cartilage, B) inflated tracheostomy balloon.

Initial bronchoscopic view of the glottis above the level of the vocal cords with significant edema and poor visualization of anatomic landmarks. Improved visualization of vocal cords with suctioning and jaw thrust. A) Left arytenoid cartilage, B) inflated tracheostomy balloon.

Discussion

We conclude that added caution is mandated when manipulating a recently placed tracheostomy as replacement of the tracheostomy could fail, even with the use of a bougie catheter (6), which would require orotracheal intervention “from above.” Even with previously unremarkable intubations and airway anatomy, tracheostomy placement can cause or coincide with glottic edema that could cause failure of direct or indirect laryngoscopy. A catastrophic complication can be avoided by using bronchoscopy with a loaded endotracheal tube.
  5 in total

1.  Percutaneous tracheostomy: to bronch or not to bronch--that is the question.

Authors:  La Scienya M Jackson; James W Davis; Krista L Kaups; Lawrence P Sue; Mary M Wolfe; John F Bilello; Deborah Lemaster
Journal:  J Trauma       Date:  2011-12

2.  Should a percutaneous dilational tracheostomy be guided with a bronchoscope?

Authors:  S Abdulla; A Conrad; S Vielhaber; R Eckhardt; W Abdulla
Journal:  B-ENT       Date:  2013       Impact factor: 0.082

3.  Performing Percutaneous Dilational Tracheostomy without using Fiberoptic Bronchoscope.

Authors:  Siamak Yaghoubi; Nilofar Massoudi; Mohammad Fathi; Navid Nooraei; Marzieh Beygom Khezri; Sareh Abdollahi
Journal:  Tanaffos       Date:  2020-01

4.  Safety of bedside percutaneous tracheostomy in the critically ill: evaluation of more than 3,000 procedures.

Authors:  Bradley M Dennis; Matthew J Eckert; Oliver L Gunter; John A Morris; Addison K May
Journal:  J Am Coll Surg       Date:  2013-02-08       Impact factor: 6.113

5.  Assessment of healthcare professionals' knowledge of managing emergency complications in patients with a tracheostomy.

Authors:  P Casserly; E Lang; J E Fenton; M Walsh
Journal:  Br J Anaesth       Date:  2007-07-03       Impact factor: 9.166

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.