Specific absorption rate and temperature in neonate models resulting from exposure to a 7T head coil.

Shaihan J Malik1,2, Jeffrey W Hand1, Ryan Satnarine1, Anthony N Price1,2, Joseph V Hajnal1,2.   

Abstract

PURPOSE: To investigate safe limits for neonatal imaging using a 7T head coil, including both specific absorption rate (SAR) and temperature predictions.
METHODS: Head-centered neonate models were simulated using finite-difference time domain-based electromagnetic and thermal solvers. The effects of higher water content of neonatal tissues compared with adults, position shifts, and thermal insulation were also considered. An adult model was simulated for comparison.
RESULTS: Maximum and average SAR are both elevated in the neonate when compared with an adult model. When normalized to B 1 + , the SAR experienced by a neonate is greater than an adult by approximately a factor of 2; when normalized to net forward power (forward-reflected), this increases to a factor of 2.5-3.0; and when normalized to absorbed power, approximately a factor of 4. Use of age-adjusted dielectric properties significantly increases the predicted SAR, compared with using adult tissue properties for the neonates. Thermal simulations predict that change in core temperature/maximum temperature remain compliant with International Electrotechnical Commission limits when a thermally insulated neonate is exposed at the SAR limit for up to an hour.
CONCLUSION: This study of two neonate models cannot quantify the variability expected within a larger population. Likewise, the use of age-adjusted dielectric properties have a significant effect, but while their use is well motivated by literature, there is uncertainty in the true dielectric properties of neonatal tissue. Nevertheless, the main finding is that unlike at lower field strengths, operational limits for 7T neonatal MRI using an adult head coil should be more conservative than limits for use on adults.
© 2021 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  7T MRI; RF safety; neonatal imaging

Year:  2021        PMID: 33811667     DOI: 10.1002/mrm.28784

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  3 in total

1.  A Preliminary Study for Reference RF Coil at 11.7 T MRI: Based on Electromagnetic Field Simulation of Hybrid-BC RF Coil According to Diameter and Length at 3.0, 7.0 and 11.7 T.

Authors:  Jeung-Hoon Seo; Jun-Young Chung
Journal:  Sensors (Basel)       Date:  2022-02-15       Impact factor: 3.576

2.  Evaluation of specific absorption rate and heating in children exposed to a 7T MRI head coil.

Authors:  Shaihan J Malik; Jeffrey W Hand; David W Carmichael; Joseph V Hajnal
Journal:  Magn Reson Med       Date:  2022-06-06       Impact factor: 3.737

3.  Towards an integrated neonatal brain and cardiac examination capability at 7 T: electromagnetic field simulations and early phantom experiments using an 8-channel dipole array.

Authors:  Jérémie Clément; Raphaël Tomi-Tricot; Shaihan J Malik; Andrew Webb; Joseph V Hajnal; Özlem Ipek
Journal:  MAGMA       Date:  2022-01-08       Impact factor: 2.533

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.