Literature DB >> 33800790

A Cost-Effective Inertial Measurement System for Tracking Movement and Triggering Kinesthetic Feedback in Lower-Limb Prosthesis Users.

McNiel-Inyani Keri1, Ahmed W Shehata2, Paul D Marasco3,4, Jacqueline S Hebert2,5,6, Albert H Vette1,6.   

Abstract

Advances in lower-limb prosthetic technologies have facilitated the restoration of ambulation; however, users of such technologies still experience reduced balance control, also due to the absence of proprioceptive feedback. Recent efforts have demonstrated the ability to restore kinesthetic feedback in upper-limb prosthesis applications; however, technical solutions to trigger the required muscle vibration and provide automated feedback have not been explored for lower-limb prostheses. The study's first objective was therefore to develop a feedback system capable of tracking lower-limb movement and automatically triggering a muscle vibrator to induce the kinesthetic illusion. The second objective was to investigate the developed system's ability to provide kinesthetic feedback in a case participant. A low-cost, wireless feedback system, incorporating two inertial measurement units to trigger a muscle vibrator, was developed and tested in an individual with limb loss above the knee. Our system had a maximum communication delay of 50 ms and showed good tracking of Gaussian and sinusoidal movement profiles for velocities below 180 degrees per second (error < 8 degrees), mimicking stepping and walking, respectively. We demonstrated in the case participant that the developed feedback system can successfully elicit the kinesthetic illusion. Our work contributes to the integration of sensory feedback in lower-limb prostheses, to increase their use and functionality.

Entities:  

Keywords:  device development; kinematic feedback; kinesthetic feedback; kinesthetic illusion; lower-limb prostheses; sensory feedback; wearable sensor

Mesh:

Year:  2021        PMID: 33800790      PMCID: PMC7961441          DOI: 10.3390/s21051844

Source DB:  PubMed          Journal:  Sensors (Basel)        ISSN: 1424-8220            Impact factor:   3.576


  29 in total

1.  ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion--part I: ankle, hip, and spine. International Society of Biomechanics.

Authors:  Ge Wu; Sorin Siegler; Paul Allard; Chris Kirtley; Alberto Leardini; Dieter Rosenbaum; Mike Whittle; Darryl D D'Lima; Luca Cristofolini; Hartmut Witte; Oskar Schmid; Ian Stokes
Journal:  J Biomech       Date:  2002-04       Impact factor: 2.712

2.  Dynamic analysis of above-knee amputee gait.

Authors:  Tae Soo Bae; Kuiwon Choi; Daehie Hong; Museong Mun
Journal:  Clin Biomech (Bristol, Avon)       Date:  2007-02-23       Impact factor: 2.063

3.  Proprioception from a neurally controlled lower-extremity prosthesis.

Authors:  Tyler R Clites; Matthew J Carty; Jessica B Ullauri; Matthew E Carney; Luke M Mooney; Jean-François Duval; Shriya S Srinivasan; Hugh M Herr
Journal:  Sci Transl Med       Date:  2018-05-30       Impact factor: 17.956

4.  The contribution of muscle afferents to kinaesthesia shown by vibration induced illusions of movement and by the effects of paralysing joint afferents.

Authors:  G M Goodwin; D I McCloskey; P B Matthews
Journal:  Brain       Date:  1972       Impact factor: 13.501

5.  Prolonged muscle vibration reducing motor output in maximal voluntary contractions in man.

Authors:  L G Bongiovanni; K E Hagbarth; L Stjernberg
Journal:  J Physiol       Date:  1990-04       Impact factor: 5.182

6.  Characterizing the effects of amplitude, frequency and limb position on vibration induced movement illusions: Implications in sensory-motor rehabilitation.

Authors:  Jonathon S Schofield; Michael R Dawson; Jason P Carey; Jacqueline S Hebert
Journal:  Technol Health Care       Date:  2015       Impact factor: 1.285

7.  Human Joint Angle Estimation with Inertial Sensors and Validation with A Robot Arm.

Authors:  Mahmoud El-Gohary; James McNames
Journal:  IEEE Trans Biomed Eng       Date:  2015-02-12       Impact factor: 4.538

8.  Differences in gait characteristics between persons with bilateral transtibial amputations, due to peripheral vascular disease and trauma, and able-bodied ambulators.

Authors:  Po-Fu Su; Steven A Gard; Robert D Lipschutz; Todd A Kuiken
Journal:  Arch Phys Med Rehabil       Date:  2008-07       Impact factor: 3.966

9.  An Ankle-Foot Prosthesis for Rock Climbing Augmentation.

Authors:  Emily A Rogers; Matthew E Carney; Seong Ho Yeon; Tyler R Clites; Dana Solav; Hugh M Herr
Journal:  IEEE Trans Neural Syst Rehabil Eng       Date:  2021-02-25       Impact factor: 3.802

10.  IMU-based joint angle measurement for gait analysis.

Authors:  Thomas Seel; Jörg Raisch; Thomas Schauer
Journal:  Sensors (Basel)       Date:  2014-04-16       Impact factor: 3.576

View more