The concurrent validity and intrarater reliability of a hand-held dynamometer for the assessment of neck strength in semi-professional rugby union players.

Adam Ashall1, Nick Dobbin2, Cari Thorpe3.   

Abstract

OBJECTIVES: The main objective of this study was to determine the concurrent validity between a hand-held (HHD) and mounted hand-held dynamometer (MHHD) for assessing isometric neck strength.
DESIGN: Observational design.
SETTING: Semiprofessional rugby club. PARTICIPANTS: Nineteen semi-professional rugby players (age = 26 ± 5 years, stature = 186.5 ± 6.5 cm, body mass = 98.7 ± 12.8 kg). MAIN OUTCOME MEASURES: Concurrent validity (limits of agreement, correlation) between HHD and MHHD, the intrarater reliability (intra-class correlation, ICC) and comparison between playing positions.
RESULTS: Absolute peak and mean peak force were systematically lower when using the HHD compared to MHHD, with the mean bias ranging from -1.8 to -3.8 kgf (P < 0.05). Differences were not evident for flexion when applying the correction equations (-0.5 to 2.1 kgf, P > 0.05) but remained for extension. Correlations between methods were large-to-very large; the ICCs for both methods were good (ICC = 0.72-0.89), with no difference between positions (P < 0.05).
CONCLUSION: The concurrent validity of HHD was considered acceptable when compared to the MHHD and the correction equation applied. Both methods are reliable and useful for assessing neck strength in rugby players, though, caution is needed when determining strength during neck extension.
Copyright © 2021 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Concussion playing position; Injury; Physiotherapy; Rehabilitation; Rugby

Year:  2021        PMID: 33794446     DOI: 10.1016/j.ptsp.2021.03.007

Source DB:  PubMed          Journal:  Phys Ther Sport        ISSN: 1466-853X            Impact factor:   2.365


  2 in total

1.  Reliability of Repeated Isometric Neck Strength in Rugby Union Players Using a Load Cell Device.

Authors:  Christian Chavarro-Nieto; Martyn Beaven; Nicholas Gill; Kim Hébert-Losier
Journal:  Sensors (Basel)       Date:  2022-04-08       Impact factor: 3.847

2.  Effects of Knee Joint Angle and Contraction Intensity on the Triceps Surae Stiffness.

Authors:  Ming Lin; Weixin Deng; Hongying Liang; Suiqing Yu; Qin Xu; Chunlong Liu
Journal:  Front Bioeng Biotechnol       Date:  2022-06-22
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.