Abstract
We report the development of a highly sensitive ratiometric fluorescent lateral flow immunoassay (RFLFIA) strip for rapid and accurate detection of acute myocardial infarction biomarker, namely heart-type fatty acid binding protein (H-FABP). The RFLFIA strip works in terms of ratiometric change of fluorescence signal, arising from blending of fluorescence emitted by two composite nanostructures conjugated to capture and probe antibodies and inner filter effect of gold nanoparticles. In conjunction with using custom smartphone-based analytical device and tonality analysis, quantitative detection of H-FABP was achieved with a low limit of detection at 0.21 ng mL-1 . The RFLFIA strip can generate a visually distinguishable green-to-red color change around the threshold concentration of H-FABP (6.2 ng mL-1 ), thus allowing the semi-quantitative diagnosis by the naked eye.Entities:
Keywords: H-FABP; immunoassays; point-of-care; ratiometric fluorescence; visual detection
Year: 2021 PMID: 33793060 DOI: 10.1002/anie.202103458
Source DB: PubMed Journal: Angew Chem Int Ed Engl ISSN: 1433-7851 Impact factor: 15.336