Noninvasive sampling of the distal airspace via HME-filter fluid is not useful to detect SARS-CoV-2 in intubated patients.

Joerg Reifart1,2, Christoph Liebetrau3,4, Christian Troidl3,4, Katharina Madlener5, Andreas Rolf3,4.   

Abstract

Entities:  

Keywords:  BAL; HME; Infection; PCR; SARS-CoV-2

Mesh:

Year:  2021        PMID: 33785049      PMCID: PMC8008771          DOI: 10.1186/s13054-021-03549-x

Source DB:  PubMed          Journal:  Crit Care        ISSN: 1364-8535            Impact factor:   9.097


× No keyword cloud information.
Sampling for SARS-CoV-2 is often carried out via bronchoscopy in intubated patients if upper respiratory samples at nasopharyngeal and oropharyngeal sites are negative or cannot be readily obtained [1]. PCR for detection of SARS-CoV-2 has been reported to have the highest sensitivity in lower respiratory tract samples [2]. Because bronchoscopy is an aerosol-generating procedure, alternative sampling methods with a similar sensitivity are of value. Previously, it has been demonstrated that heat moisture exchanger (HME) filter fluid had a composition approximating that of the fluid in the distal airspace (which is obtained during bronchoalveolar lavage) and can be used to test for pathogens [3-5]. The present investigation examined whether performing PCR for SARS-CoV-2 on unprocessed HME filter fluid is a viable method to test patients for infection instead of performing bronchoalveolar lavage via bronchoscopy. Patients who had tested positive for SARS-CoV-2 and were treated with invasive ventilation for at least 12 h were enrolled prospectively (Fig. 1). The enrollment period was from May 2020 to December 2020. The initial test yielding the positive result was conducted using a nasopharyngeal swab.
Fig. 1

Flowchart of study design

Flowchart of study design HME filters were swabbed with a regular swab on the septic side of the filter. The swabs were then placed in virus medium. SARS-CoV-2 PCR test systems used were either the BD MAX™ System with BioGX SARS-CoV-2 reagents (BD Life Sciences, Sparks, Maryland, USA) or the Hain Lifescience FluoroType® SARS-CoV-2 plus (Hain Lifescience GmbH, Nehren, Germany). Results of the tests of samples from the HME filters were compared with the results from the nasopharyngeal swab. A sample size of 60 patients appeared to be adequate to compare the two different test methods. This report provides data for the initial 4 patients included in this prospective population. Informed consent was acquired. If the patient was not able to provide informed consent, inclusion occured via an investigator consilium ("Giessener Lösung"). The ethics board of the state of Hessen, Germany, approved the study (AZ 79/20). The data collected from all patients were pseudonymized and entered into a database. The trial was retrospectively registered on November 1st 2020 (DRKS registry; registration: DRKS00023494). Four patients with positive nasopharyngeal swabs were enrolled in the study before an interim analysis of the results showed that it was unlikely that a sensitivity of greater than 90% would be reached. The enrollment was stopped prematurely. Of the 4 HME filter samples assayed, 3 did not test positive for SARS-CoV-2. All patients still tested positive in a bronchoalveolar lavage sample at a later point in time. Further results are provided in Table 1.
Table 1

Patient and testing data

VariablePatient 1Patient 2Patient 3Patient 4
Age, y72857966
Female sexYesNoNoYes
Body-mass index, kg/m [2]33272433
CT/chest X-ray morphology for COVID-19YesYesYesYes
Time from first positive PCR test until HME swab, h7419222235
Time on ventilator until HME swab, h6819212159
Time on ventilator (total), h494243369159
SARS-CoV-2-positive HME PCR samplesYesNoNoNo
SARS-CoV-2-positive PCR samples after HME PCR testing (BAL)YesYesYesYes
Therapy with corticosteroidYesYesYesYes
Therapy with azithromycinYesYesYesYes
Antiviral medicationNoNoNoNo
Death within 30 daysNoYesYesNo
Patient and testing data Bronchoscopy with bronchoalveolar lavage is commonly used to test for SARS-CoV-2 in intubated patients. This study was designed to determine whether sampling HME filter fluid may present a feasible and safer alternative. Our preliminary data showed that HME filter fluid alone, without any additional processing, is not suitable for the detection of SARS-CoV-2 via PCR. Given the manifest disease in the lung as observed in CT scans, a positive SARS-CoV-2 swab in the upper respiratory tract without the lungs being affected is unlikely in patients needing invasive ventilation. Hence, the reason for the negative results remain unclear. Unfortunately, the study did not control for prior inhalation therapy. Some of the fluid could have been nebulized saline instead of precipitated moisture from the distal airspace carrying viral RNA. Even so, viral RNA should still be detectable under these conditions. Further processing of the samples, including the extraction of more liquid via centrifugation of the HME filters, might yield better results, though processing steps would render the technique less applicable and appealing for widespread use [5].
  5 in total

1.  Non-invasive detection of pulmonary pathogens in ventilator-circuit filters by PCR.

Authors:  Richard J Isaacs; Ken Debelak; Patrick R Norris; Judith M Jenkins; Jeffrey C Rooks; Todd R Young; Addison K May; Erik M Boczko
Journal:  Am J Transl Res       Date:  2012-01-05       Impact factor: 4.060

2.  A pilot study of the noninvasive assessment of the lung microbiota as a potential tool for the early diagnosis of ventilator-associated pneumonia.

Authors:  Addison K May; Jacob S Brady; Joann Romano-Keeler; Wonder P Drake; Patrick R Norris; Judith M Jenkins; Richard J Isaacs; Erik M Boczko
Journal:  Chest       Date:  2015-06       Impact factor: 9.410

3.  Detection of SARS-CoV-2 in Different Types of Clinical Specimens.

Authors:  Wenling Wang; Yanli Xu; Ruqin Gao; Roujian Lu; Kai Han; Guizhen Wu; Wenjie Tan
Journal:  JAMA       Date:  2020-05-12       Impact factor: 56.272

4.  Novel Method for Noninvasive Sampling of the Distal Airspace in Acute Respiratory Distress Syndrome.

Authors:  J Brennan McNeil; Ciara M Shaver; V Eric Kerchberger; Derek W Russell; Brandon S Grove; Melissa A Warren; Nancy E Wickersham; Lorraine B Ware; W Hayes McDonald; Julie A Bastarache
Journal:  Am J Respir Crit Care Med       Date:  2018-04-15       Impact factor: 21.405

5.  American Association for Bronchology and Interventional Pulmonology (AABIP) Statement on the Use of Bronchoscopy and Respiratory Specimen Collection in Patients With Suspected or Confirmed COVID-19 Infection.

Authors:  Momen M Wahidi; Carla Lamb; Septimiu Murgu; Ali Musani; Samira Shojaee; Ashutosh Sachdeva; Fabien Maldonado; Kamran Mahmood; Matthew Kinsey; Sonali Sethi; Amit Mahajan; Adnan Majid; Colleen Keyes; Abdul H Alraiyes; Arthur Sung; David Hsia; George Eapen
Journal:  J Bronchology Interv Pulmonol       Date:  2020-10
  5 in total
  2 in total

1.  Sensitivity and specificity of heat and moisture exchange filters sampling for SARS-CoV-2 detection in mechanically ventilated COVID-19 patients.

Authors:  Toufik Kamel; Clémence Guillaume; Grégoire Muller; Lekbir Baala; Thierry Boulain
Journal:  J Microbiol Immunol Infect       Date:  2022-04-14       Impact factor: 10.273

2.  SARS-CoV-2 viral load in heat and humidity exchange filters during invasive mechanical ventilation of patients with COVID-19.

Authors:  Giorgio Enzo Fulceri; Fabio Morecchiato; Alberto Antonelli; Manuela Bonizzoli; Michele Cecchi; Gian Maria Rossolini; Adriano Peris
Journal:  Br J Anaesth       Date:  2022-09-12       Impact factor: 11.719

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.