| Literature DB >> 33779041 |
William S Ferris1, Wesley S Culberson1, Jennifer B Smilowitz1,2, John E Bayouth2.
Abstract
PURPOSE: Radixact Synchrony corrects for target motion during treatment by adjusting the jaw and MLC positions in real time. As the jaws move off axis, Synchrony attempts to adjust for a loss in output due to the un-flattened 6 MV beam by increasing the jaw aperture width. The purpose of this work was to assess the impact of the variable-width aperture on delivered dose using measurements and simulations.Entities:
Keywords: radixact; synchrony; tomotherapy
Year: 2021 PMID: 33779041 PMCID: PMC8130229 DOI: 10.1002/acm2.13234
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Fig. 1Schematic of collimation and jaw compensation on Radixact. Beam’s eye view (a) and side view (b) with centered jaws and central three leaves open. Beam’s eye view (c) and side view (d) with jaws compensating in the positive IEC‐Y direction. Note that there are 64 total MLC leaves on Radixact and drawings are not to scale.
Fig. 2Photograph of the measurement setup with Edge diode aligned perpendicular to the direction of couch travel. A cutout was made in the 5 mm thick bolus material to reduce the air gap around the detector. The 10 cm of solid water on top of the gel and detector was removed for the photograph. The detector volume was centered in IEC‐X.
Jaw aperture widths used in this work. Aperture settings refer to the physical aperture at 23 cm from the MV target. The extreme locations correspond to shifts in IEC‐Y at isocenter of ± 2.0 cm and ± 1.25 cm for the 7‐mm and 20‐mm apertures, respectively.
| Jaw aperture width (mm) | |||
|---|---|---|---|
| Compensation mode | Centered | Positive extreme | Negative extreme |
| Fixed‐width | 7.00 | 7.00 | 7.00 |
| Synchrony | 7.00 | 7.75 | 7.75 |
| Fixed‐width | 20.00 | 20.00 | 20.00 |
| Synchrony | 20.00 | 20.15 | 20.15 |
Fig. 3Profiles at 10‐cm depth and 85‐cm source‐to‐surface distance measured with an Edge diode. The y‐axis displays the signal relative to the maximum signal for the profile measured with the centered aperture. The colored horizontal ticks indicate the level of the FWHM for each curve. Displayed at the off‐axis locations are profiles acquired with varying jaw aperture widths (see Table 1).
Beam characteristics (jaw‐offset peak, width, and integral factors) for measured profiles at 10‐cm depth. The JOIFs were calculated with a 10% dose threshold. The pair of values in each entry indicates the value for the negative extreme and positive extreme profile, respectively.
| Jaw | Compensation type | JOPF | JOWF | JOIF |
|---|---|---|---|---|
| 2.5 cm | Synchrony | 0.984/0.983 | 1.007/1.006 | 0.978/0.976 |
| Fixed‐width | 0.984/0.982 | 0.999/0.998 | 0.981/0.978 | |
| 1.0 cm | Synchrony | 0.970/0.964 | 1.019/1.020 | 0.992/0.983 |
| Fixed‐width | 0.934/0.925 | 0.944/0.945 | 0.894/0.883 |
Fig. 4Simulated dose profiles in the IEC‐Y direction in the frame of reference of a cylindrical target from a helical IMRT delivery. Profiles are centered in X and Z. The vertical lines indicate the limits of the target in IEC‐Y (5‐cm diameter, 5‐cm length cylinder). The prescription dose was 50 Gy. Negative and positive refer to the target shifted to the positive or negative extreme location throughout treatment and compensated with jaw motion. The resolution is 0.5 mm in the IEC‐Y direction. Gamma profiles (1%, 1 mm) are shown for the Synchrony plans compared to planned.
Gamma pass rates for simulated treatment plans comparing the motion compensated dose distributions to planned. All dose differences are relative to the global maximum dose. A dose threshold of 10% of the prescription dose (50 Gy) is applied to all data. Gamma analyses are 3D. The extreme locations correspond to shifts in IEC‐Y at isocenter of ± 2.0 cm and ± 1.25 cm for the 1.0‐cm and 2.5‐cm jaws, respectively.
| Jaw | Position | Fixed‐Width | Synchrony | ||
|---|---|---|---|---|---|
| 3%, 2 mm | 1%, 1 mm | 3%, 2mm | 1%, 1 mm | ||
| 2.5 cm | Negative extreme | 100 | 94.67 | 100 | 99.20 |
| Positive extreme | 100 | 94.44 | 100 | 98.70 | |
| 1.0 cm | Negative extreme | 62.16 | 48.11 | 100 | 98.89 |
| Positive extreme | 62.01 | 47.02 | 100 | 99.05 | |
Universal tolerance limit recommended by TG‐218 for IMRT QA is > 95% of points passing 3%, 2 mm, with a 10% dose threshold.
Fig. 5Jaw‐offset peak factors (JOPFs) as a function of target shift in IEC‐Y for the variable‐aperture jaw behavior during Synchrony treatments. Profiles were calculated with the TPS at a depth of 1.5‐cm and 85‐cm SSD. The values are an average of positive and negative JOPFs. Values at the extreme positions agree with the measured values in Table 2.