A case report of inappropriate inhibition of ventricular pacing due to a unique pacemaker electrogram storage feature.

Habib R Khan1, William K Chan1, Juliana Kanawati1, Raymond Yee1.   

Abstract

BACKGROUND: Modern permanent pacemakers (PPMs) have individual features designed to identify cardiac rhythm abnormalities and improve their performance. Inappropriate pacing inhibition may be an undesired outcome from these features and cause symptoms in patients who require frequent pacing, leading to dizziness, and syncope. Inappropriate inhibition can be difficult to identify in circumstances that are intermittent and difficult to reproduce. CASE
SUMMARY: A 57-year-old female underwent a mitral valve replacement (MVR) for severe mitral stenosis. One month following MVR, she presented with symptomatic third-degree atrioventricular block, and a dual-chamber PPM (Advisa™, Medtronic, Minneapolis, USA) was implanted and programmed DDD 50-130 b.p.m. At the 3-month follow-up, she reported frequent episodes of lightheadedness. She was found to have intermittent ventricular pacing inhibition on a 48-h Holter monitor due to an internal function of the Advisa™ series of PPMs that attempts to store an electrogram (EGM) every 1 h and 30 s. During the EGM storage, an amplified signal from the storage capacitor can result in oversensing by the ventricular channel and inappropriate pacing inhibition. DISCUSSION: To rectify the issue, the ventricular lead sensitivity value was increased from 0.9 mV to 1.2 mV. No instances of inappropriate ventricular pacing inhibition were noted on follow-up. To our knowledge, this is a rare case of inappropriate ventricular pacing inhibition caused by a combination of PPM self-adjusting sensitivity algorithm and oversensing every 1 h and 30 s from an amplified storage capacitor. Physicians should be aware of this possible complication and differentiate it from device or lead malfunction.
© The Author(s) 2021. Published by Oxford University Press on behalf of the European Society of Cardiology.

Entities:  

Keywords:  Advisa; Case report; Dizziness; Oversensing; Pacemaker; Pacing inhibition

Year:  2021        PMID: 33768197      PMCID: PMC7980079          DOI: 10.1093/ehjcr/ytab096

Source DB:  PubMed          Journal:  Eur Heart J Case Rep        ISSN: 2514-2119


This case highlights an important cause of inappropriate inhibition of ventricular pacing due to a non-programmable feature of a pacemaker and can be translated to other manufacturers that might have other proprietary algorithms with similar effect. If a regular inhibition of ventricular pacing is seen in a Medtronic Advisa™ pacemaker, the electrogram storage feature resulting in inhibition should be considered as a cause and sensitivity reduced as a treatment option.

Introduction

All modern permanent pacemakers (PPMs) can function in a demand mode, meaning that pacing stimuli are inhibited when the patient’s intrinsic heart rate exceeds the programmed lower rate limit or when activity is sensed in the implanted cardiac chamber. There are numerous PPM manufacturers with various individual features designed to identify cardiac rhythm abnormalities and improve the devices’ performance., Inappropriate pacing inhibition may be an undesired outcome from these features and cause symptoms in patients who require frequent pacing. Inappropriate pacing inhibition can be challenging to identify in circumstances that are intermittent and difficult to reproduce. We identified a rare case of inappropriate pacing inhibition caused by a non-programmable internal PPM feature resulting in the patient experiencing dizziness.

Case presentation

A 57-year-old Caucasian woman underwent a mitral valve replacement (MVR) for severe mitral stenosis. Her past medical history included obstructive sleep apnoea, Hashimoto’s thyroiditis, and gastro-oesophageal reflux disease. The regular medications included thyroxine 50 mcg daily, warfarin, and Aspirin 81 mg daily. One month following MVR, she presented with symptomatic third-degree atrioventricular (AV) block, and a dual-chamber PPM (Advisa™, Medtronic, Minneapolis, USA) was implanted as per guidelines and programmed DDD 50–130 b.p.m., However, several months after her device implantation, she reported frequent episodes of abrupt lightheadedness that were not orthostatic. Device interrogation parameters were unchanged from the time of implant. P waves were 2.8 mV and R waves were 18 mV. Right atrial and right ventricular lead impedances were 531 and 862 Ohms, respectively. Capture thresholds on both leads were <1 V/0.5 ms. Provocation testing in the clinic with deep respiration and active movement of the arms and the generator pocket did not evoke pacing inhibition or show artefact on either atrial or ventricular leads. Twelve-lead electrocardiogram (ECG) showed atrial tracking with ventricular pacing and no loss of capture. Chest X-ray showed no obvious lead dislodgement or fracture. Since she was stable and had no syncope, she was ultimately sent home with an ambulatory Holter monitor. She had intermittent ventricular pacing inhibition, as shown in documented only on the Holter monitor. Ventricular pacing inhibition was seen with sinus rate tracking of ≤75 b.p.m., was not preceded by any visualized artefact, and did not correspond to standard device features such as Managed Ventricular Pacing or Capture Management. AV delay remained the same during ventricular capture, excluding AV hysteresis. Holter monitor showing two time points of pacing inhibition related to intermittent oversensing in the ventricular lead. This was due to a repetitive feature of electrogram collection every 1 h and 30 s. Stepwise approach to pre-syncope in pacemaker-dependent patients. The pacing inhibition’s recurrent and predictable nature every 1 h and 30 s suggested a device feature as opposed to a hardware malfunction. This was reported to the device manufacturer and discussed with a technical expert from the company. In this case, it was caused by an internal function of the Advisa™ series of PPMs that attempts to store an electrogram (EGM) every 1 h and 30 s. During EGM storage, an amplified signal from the storage capacitor can result in oversensing by the ventricular channel and inappropriate ventricular pacing inhibition. To rectify the issue, the ventricular lead sensitivity value was increased from 0.9 mV to 1.2 mV. No further instances of inappropriate ventricular pacing inhibition were noted on the follow-up ambulatory monitor at 3 months, and the patient’s symptoms resolved.

Discussion

Dizziness or syncope is a concerning symptom in patients with PPMs, requiring urgent attention and thorough investigation. Although there are many potential causes uncovered through careful history and review of medications, additional tests such as a 12-lead ECG, chest X-ray, and device interrogation with provocative manoeuvres may help rule out PPM hardware malfunction (., Ambulatory monitoring is useful in identifying causes of ‘pseudo-malfunction’ as a result of increasingly complex, proprietary device features. These features can cause confusion and sometimes lead to inappropriate pacing inhibition as seen in our patient who was pacing-dependent.,
Figure 2

Stepwise approach to pre-syncope in pacemaker-dependent patients.

Pacemaker oversensing occurs due to multiple factors that can be categorized as extrinsic or intrinsic. Extrinsic factors include lead noise or artefact caused by electromagnetic interference from machinery and sources of high voltage. Intrinsic factors can be from physiologic or non-physiologic sources. Physiologic sources include cross-talk due to oversensing of signal in the opposite chamber.,, Non-physiologic sources include lead fracture or a misaligned connection of the lead connector pin into the device connector block. If all of these causes are ruled out, then ‘pseudo malfunction’ due to device-specific features should be considered., This case report highlights a rare cause of intrinsic non-physiologic noise wherein a signal is generated by a circuit within the PPM that interferes with normal device function. In the Advisa™ series of PPMs, ventricular lead sensing is managed by a self-adjusting function that was originally developed for implantable cardioverter-defibrillators to identify tachyarrhythmias. Sensing filters are usually present in defibrillators to allow for an adjustment of ventricular sensitivity with decay for a programmed length of time (nominally 450 ms). The self-adjusting feature is based on a time-dependent sensitivity decay after a ventricular sensed event. After each ventricular sensed event, the sensitivity decay starts at either 75% of the R wave (18 mV for our case) or a value calculated as 9 times the programmed sensitivity value (0.9 mV in our case = 8.1 mV), whichever is greater during diastole. The Advisa™ model of PPMs also has a feature that collects EGMs every 1 h and 30 s, resulting in intermittent ventricular pacing inhibition as recorded on the Holter monitor in our patient (). As the EGM storage feature turned ON and OFF every 1 h and 30 s, there were instances of an amplified signal generated from the storage capacitor resulting in ventricular oversensing when the self-adjusting sensitivity feature of the ventricular channel approached 0.9 mV in late diastole. The EGM storage feature cannot be programmed OFF in the Advisa™ model and therefore the device manufacturer needs to be made aware of this to fix the underlying problem. The only immediate solution to ventricular oversensing was increasing the ventricular sensitivity value from 0.9 mV to 1.2 mV. This change in sensitivity was possible without any adverse impact on device function as the patient was pacing-dependent, and the ventricular-sensed amplitude was 18 mV. A repeat Holter monitor was performed after 3 months, and there were no further episodes of pacing inhibition. This is a rare case of inappropriate ventricular pacing inhibition caused by a combination of PPM self-adjusting sensitivity feature and oversensing of a recurrent signal produced by the EGM storage amplifier every 1 h and 30 s. Despite thousands of implanted Advisa™ PPMs, this is the second case report of this subtle anomaly that can cause significant symptoms in pacing-dependent patients. Although several proprietary device features can cause unexpected pacing inhibition, the one highlighted in this article is non-programmable and easily missed without further ambulatory monitoring. Technical assistance from the device manufacturer should be sought after when these issues are identified. Cardiologists should be aware of this possible complication and differentiate it from device or lead malfunction.

Lead author biography

Dr Habib R. Khan graduated from University of Peshawar, Pakistan in 2002. After completing resident training, he trained in medicine in Ireland and obtained MRCP Ireland and UK. He is certified by European Examination in General Cardiology, and Cardiac Rhythm Management certification in Devices by British Heart Rhythm Society. He is currently appointed as an Assistant Professor at University of Western, London-Ontario, Canada. Dr Khan regularly publishes in field of cardiovascular medicine.

Supplementary material

Supplementary material is available at European Heart Journal - Case Reports online. Slide sets: A fully edited slide set detailing this case and suitable for local presentation is available online as Supplementary data. Consent: The authors confirm that written consent for submission and publication of this case report including images and associated text has been obtained from the patient in line with COPE guidance. Conflict of interest: None declared. Funding: None declared. Click here for additional data file.
Day 0A 57-year-old woman with severe mitral stenosis underwent mechanical mitral valve replacement.
Day 30She presented with symptomatic intermittent complete heart block requiring implantation of a dual-chamber pacemaker (Advisa™).
Day 122She complained of intermittent unprovoked lightheadedness and was identified to have inappropriate pacing inhibition on a 48-h Holter monitor.
Day 162Readjustment of ventricular chamber sensitivity value and repeat ambulatory electrocardiogram monitor showed resolution of the problem.
Day 256No further pacing inhibition on monitor and symptoms resolved.
  10 in total

Review 1.  Pacing features that mimic malfunction: a review of current programmable and automated device functions that cause confusion in the clinical setting.

Authors:  Michael S Lloyd; Mikhael F El Chami; Jonathan J Langberg
Journal:  J Cardiovasc Electrophysiol       Date:  2009-04

2.  Regular ventricular pacing inhibition due to electronic artifact emitted by pulse generator equipped with a self-adjusting sensitivity function.

Authors:  Takashi Iizuka; Yoshiaki Kaneko; Tadashi Nakajima; Masahiko Kurabayashi
Journal:  Heart Rhythm       Date:  2013-10-15       Impact factor: 6.343

3.  2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA).

Authors:  Michele Brignole; Angelo Auricchio; Gonzalo Baron-Esquivias; Pierre Bordachar; Giuseppe Boriani; Ole-A Breithardt; John Cleland; Jean-Claude Deharo; Victoria Delgado; Perry M Elliott; Bulent Gorenek; Carsten W Israel; Christophe Leclercq; Cecilia Linde; Lluís Mont; Luigi Padeletti; Richard Sutton; Panos E Vardas; Jose Luis Zamorano; Stephan Achenbach; Helmut Baumgartner; Jeroen J Bax; Héctor Bueno; Veronica Dean; Christi Deaton; Cetin Erol; Robert Fagard; Roberto Ferrari; David Hasdai; Arno W Hoes; Paulus Kirchhof; Juhani Knuuti; Philippe Kolh; Patrizio Lancellotti; Ales Linhart; Petros Nihoyannopoulos; Massimo F Piepoli; Piotr Ponikowski; Per Anton Sirnes; Juan Luis Tamargo; Michal Tendera; Adam Torbicki; William Wijns; Stephan Windecker; Paulus Kirchhof; Carina Blomstrom-Lundqvist; Luigi P Badano; Farid Aliyev; Dietmar Bänsch; Helmut Baumgartner; Walid Bsata; Peter Buser; Philippe Charron; Jean-Claude Daubert; Dan Dobreanu; Svein Faerestrand; David Hasdai; Arno W Hoes; Jean-Yves Le Heuzey; Hercules Mavrakis; Theresa McDonagh; Jose Luis Merino; Mostapha M Nawar; Jens Cosedis Nielsen; Burkert Pieske; Lidija Poposka; Frank Ruschitzka; Michal Tendera; Isabelle C Van Gelder; Carol M Wilson
Journal:  Eur Heart J       Date:  2013-06-24       Impact factor: 29.983

Review 4.  Troubleshooting implanted cardioverter defibrillator sensing problems I.

Authors:  Charles D Swerdlow; Samuel J Asirvatham; Kenneth A Ellenbogen; Paul A Friedman
Journal:  Circ Arrhythm Electrophysiol       Date:  2014-12

Review 5.  Troubleshooting implantable cardioverter-defibrillator sensing problems II.

Authors:  Charles D Swerdlow; Samuel J Asirvatham; Kenneth A Ellenbogen; Paul A Friedman
Journal:  Circ Arrhythm Electrophysiol       Date:  2015-02

6.  Double chamber cross-talk, the ultimate double-cross.

Authors:  Fahad Almehmadi; Stephen Duffett; Habib R Khan; George Klein; Raymond Yee
Journal:  Pacing Clin Electrophysiol       Date:  2020-06-29       Impact factor: 1.976

Review 7.  Cardiac Pacemakers: Function, Troubleshooting, and Management: Part 1 of a 2-Part Series.

Authors:  Siva K Mulpuru; Malini Madhavan; Christopher J McLeod; Yong-Mei Cha; Paul A Friedman
Journal:  J Am Coll Cardiol       Date:  2017-01-17       Impact factor: 24.094

8.  2018 ACC/AHA/HRS Guideline on the Evaluation and Management of Patients With Bradycardia and Cardiac Conduction Delay: Executive Summary: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines, and the Heart Rhythm Society.

Authors:  Fred M Kusumoto; Mark H Schoenfeld; Coletta Barrett; James R Edgerton; Kenneth A Ellenbogen; Michael R Gold; Nora F Goldschlager; Robert M Hamilton; José A Joglar; Robert J Kim; Richard Lee; Joseph E Marine; Christopher J McLeod; Keith R Oken; Kristen K Patton; Cara N Pellegrini; Kimberly A Selzman; Annemarie Thompson; Paul D Varosy
Journal:  Circulation       Date:  2018-11-06       Impact factor: 29.690

Review 9.  A review of specialized and automated features in implantable cardiac devices.

Authors:  Hakeem Ayinde; Michael Lloyd
Journal:  J Cardiovasc Electrophysiol       Date:  2019-11-21

10.  Syncope in Patients with Pacemakers.

Authors:  Richard Sutton
Journal:  Arrhythm Electrophysiol Rev       Date:  2015-12-01
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.