Literature DB >> 33766043

Clinical evaluation the success rate and complications of fluoroscopically guided removal of tracheal tube metallic stents.

Zong-Ming Li1, De-Chao Jiao1, Xin-Wei Han2, Hui-Bin Lu1, Ke-Wei Ren1, Hong Yang1.   

Abstract

BACKGROUND: Long-term placement of airway stents has a high probability of restenosis of the airway due to granulation tissue hyperplasia, and it is difficult to remove the stent. Our aim is to evaluate the success rate and complications of removal of tracheal tube metallic stents under fluoroscopic guidance, and to compare the difference between uncovered stent and covered stent.
METHODS: We retrospectively reviewed 45 cases (31 males and 14 females; age, 12-71 years) of tracheal metallic stent removal performed at our center between January 2014 and December 2019. Covered stents were applied in 36 cases, and uncovered stents were applied in 9 cases. In the covered stent group, 15 patients presented with granulation tissue at both ends; 3 cases, with stent fracture; and 2, with stent intolerance due to severe airway foreign body sensation. In the uncovered stents group, all patients presented with granulation tissue formation; 2 patients, with stent fracture; and 1 patient, with stent intolerance.
RESULTS: A total of 41 (91.1%) stents were successfully removed (34 [94.4%] in the covered stent group and 7 [77.8%] in the uncovered stent group). The average duration of stent placement was 3.2 ± 0.7 and 2.5 ± 1.2 months in the covered stent group and uncovered stent group, respectively. With regard to the complications, hemoptysis occurred in 4 cases (average blood volume lost, 100 ml), tracheal mucosa tear occurred in 5 cases, tracheal collapse requiring emergency airway stent placement occurred in 1 case, and tracheal rupture requiring emergency surgical suture occurred in 1 case. No procedure-related deaths occurred in either group.
CONCLUSIONS: It is safe to remove the metal stent of the tracheal tube under the guidance of fluoroscopy, with low complications, and can avoid the long-term placement of the airway stent.

Entities:  

Keywords:  Removal; Stent; Tracheal; Tubular

Mesh:

Year:  2021        PMID: 33766043      PMCID: PMC7993507          DOI: 10.1186/s13019-021-01444-8

Source DB:  PubMed          Journal:  J Cardiothorac Surg        ISSN: 1749-8090            Impact factor:   1.637


  8 in total

Review 1.  Airway stenting: Technological advancements and its role in interventional pulmonology.

Authors:  Nicolas Guibert; Hideo Saka; Hervé Dutau
Journal:  Respirology       Date:  2020-03-11       Impact factor: 6.424

2.  Perspective on airway stenting in inoperable patients with tracheoesophageal fistula after curative-intent treatment for esophageal cancer.

Authors:  Udit Chaddha; D Kyle Hogarth; Septimiu Murgu
Journal:  J Thorac Dis       Date:  2019-05       Impact factor: 2.895

3.  Successful removal of self-expanding metallic stent after deployment for tubercular bronchostenosis.

Authors:  Barney Thomas Jesudason Isaac; Ashima Datey; Devasahayam Jesudas Christopher
Journal:  Indian J Tuberc       Date:  2016-05-04

4.  Removal of a self-expanding metallic Y stent in the management of reversible central airway obstruction due to mediastinal lymphoma.

Authors:  Gopal Chawla; Karan Madan; Hariharan Iyer; Pawan Tiwari; Vijay Hadda; Anant Mohan; Saurabh Mittal
Journal:  Adv Respir Med       Date:  2020

5.  Removal of metallic tracheobronchial stents in lung transplantation with flexible bronchoscopy.

Authors:  Oren Fruchter; Yael Raviv; Benjamin D Fox; Mordechai R Kramer
Journal:  J Cardiothorac Surg       Date:  2010-09-12       Impact factor: 1.637

6.  Efficacy and safety of airway stenting to treat anastomotic complications after lung transplant: a cohort study.

Authors:  Kevin C Ma; Mingyang Li; Andrew R Haas; Anthony R Lanfranco; Edmund K Moon; David M DiBardino
Journal:  J Thorac Dis       Date:  2020-07       Impact factor: 2.895

7.  Metallic stent insertion and removal for post-tracheotomy and post-intubation tracheal stenosis.

Authors:  Yonghua Bi; Zepeng Yu; Jianzhuang Ren; Xinwei Han; Gang Wu
Journal:  Radiol Med       Date:  2018-10-24       Impact factor: 3.469

8.  An outcome analysis of self-expandable metallic stents in central airway obstruction: a cohort study.

Authors:  Fu-Tsai Chung; Hao-Cheng Chen; Chun-Liang Chou; Chih-Teng Yu; Chih-Hsi Kuo; Han-Pin Kuo; Shu-Min Lin
Journal:  J Cardiothorac Surg       Date:  2011-04-08       Impact factor: 1.637

  8 in total