Audiological benefits and performance improvements of Baha® attract implantation in patients with unilateral hearing loss.

Tae Hoon Kong1,2, Jihyeon Lee1,2, Chanbeom Kwak3,4, Woojae Han3,4, O-Hyeon Gwon2,5, Young Joon Seo1,2.   

Abstract

OBJECTIVES: To evaluate the overall functional and audiological benefits in unilateral hearing loss, as well as performance improvements with Cochlear™ Baha® Attract System with Baha® 5 sound processor.
METHODS: We performed a prospective study. Fourteen consecutive adult patients with unilateral hearing loss who were audiologically eligible for Baha® Attract were included in this study. All patients were followed-up, and a series of tests were performed in the sound field before surgery, one, two, three, and 6 months after surgery.
RESULTS: Patients were categorized based on their hearing impairment type and severity. Ten patients had conductive hearing loss (CHL), or mixed hearing loss (MHL), and four patients had sensorineural hearing loss (SNHL). The hearing threshold in the sound field aided with Baha® Attract was similar to the preoperative bone conduction hearing threshold of the better ear. The results of the speech audiometry improved post-operatively, but they did not match that of the better ear before surgery. The overall score of the hearing-in-noise test was gradually improved from one month to 6 months after surgery. The results of the sound localization test for SNHL patients were not as good as those for CHL or MHL. All the questionnaires demonstrated additional benefits that implied performance improvements.
CONCLUSIONS: The Baha® Attract provides enough auditory benefits and performance improvements for hearing impaired patients, though there are different benefits according to the type of hearing loss. Clinicians need to counsel the unilateral hearing loss patients for Baha® Attract System carefully, according to their type of hearing loss.

Entities:  

Keywords:  Audiological benefit; Baha attract; Bone conduction; Correction of hearing impairment; Unilateral hearing loss

Year:  2021        PMID: 33752582     DOI: 10.1080/14670100.2021.1903713

Source DB:  PubMed          Journal:  Cochlear Implants Int        ISSN: 1467-0100


  1 in total

1.  Percutaneous Bone-Anchored Hearing Implant: Is It Clinically Useful in Korean?

Authors:  Sung Min Koh; Young Sang Cho; Ga-Young Kim; Mini Jo; Hye Yoon Seol; Il Joon Moon
Journal:  J Korean Med Sci       Date:  2022-06-13       Impact factor: 5.354

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.