Johan Torle1, Janni Kjærgaard Thillemann2, Emil Toft Petersen2, Frank Madsen3, Kjeld Søballe4, Maiken Stilling5.
Abstract
BACKGROUND: A modular polyethylene (PE) inlay in total knee arthroplasty (TKA) may wear on both sides. PE particles may induce osteolysis, which can lead to implant loosening. The aim of this study was to determine if PE wear in monobloc TKA differs from that ofmodular TKA at 60-month follow-up. PATIENTS AND METHODS: In a prospective, patient-blinded trial, 50 patients were randomized to hybrid TKA surgery with either acementless high-porosity trabecular-metal tibial component with a monobloc UHMWPE inlay (MONO-TM) or a cementless low-porosity screw-augmented titanium fiber-mesh tibial component with a modular UHMWPE inlay (MODULAR-FM). Radiostereometry was used to measure PE wear and tibial component migration.Entities:
Keywords: Cementless; Knee arthroplasty; Modular; Monobloc; Polyethylene wear; Radiostereometry
Mesh:
Substances:
Year: 2021 PMID: 33743263 DOI: 10.1016/j.knee.2021.02.033