Comparative study of automatic blood-gas analysers and their use in analysing arterial and capillary samples.

P Rubin, S Bradbury, K Prowse.   

Abstract

Three automatic blood-gas analysers were compared for ease of use; calibration; reproducibility and accuracy of results; maintenance; fault-finding; and use of expert technician time. Results obtained from arterial and capillary blood were compared with duplicate values obtained with a semi-automatic analyser controlled and calibrated with tonometered blood. No analyser was fully automatic, and all three needed maintenance by expert technicians. Difficulties were encountered when inexperienced operators used the machines. One automatic blood-gas analyser gave aberrant values for oxygen pressure (PO2) due to electrode dysfunction that was not indicated by the fault-finding system. A second analyser gave significantly lower values for blood pH than the standard machine. A comparison of pH, carbon dioxide pressure (PCO2), and PO2 measured in 40 simultaneous paired samples of arterial and arterialised capillary blood showed no significant difference for pH or PCO2, but the PO2 values were significantly lower in the capillary samples over the range studied. We conclude that all machines perform satisfactorily in terms of blood-gas analysis, but none may be regarded as fully automatic. Some degree of technical supervision is essential, as is proper training for all potential users.

Entities:  

Mesh:

Substances:

Year:  1979        PMID: 33741      PMCID: PMC1597570          DOI: 10.1136/bmj.1.6157.156

Source DB:  PubMed          Journal:  Br Med J        ISSN: 0007-1447


  5 in total

1.  Arterialized ear lobe blood samples for blood gas tensions.

Authors:  S G Spiro; I R Dowdeswell
Journal:  Br J Dis Chest       Date:  1976-10

2.  Regional quality control survey of blood-gas analysis.

Authors:  B D Minty; J F Nunn
Journal:  Ann Clin Biochem       Date:  1977-09       Impact factor: 2.057

3.  Measuring gas in blood.

Authors:  W Else
Journal:  Nature       Date:  1972-10-20       Impact factor: 49.962

4.  Evaluation of a new thin film tonometer.

Authors:  C Chalmers; B D Bird; J G Whitwam
Journal:  Br J Anaesth       Date:  1974-04       Impact factor: 9.166

5.  Use of capillary blood in measurement of arterial PO2.

Authors:  J MacIntyre; J N Norman; G Smith
Journal:  Br Med J       Date:  1968-09-14
  5 in total
  1 in total

1.  Oxygen tension measurement using an automatic blood gas analyser.

Authors:  J Becket; C Orchard; M K Chakrabarti; G M Hall; I D Gillies; P J Bourdillon
Journal:  J R Soc Med       Date:  1981-08       Impact factor: 18.000

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.