Literature DB >> 33724262

Changes of intraocular lens position induced by Nd:YAG capsulotomy.

Martin Fus1,2, Sarka Pitrova1,2, Klara Maresova3, Jan Lestak1,2.   

Abstract

AIMS: The aim of this study was to evaluate the changes in IOL position (axial shift, tilt and decentration) in the capsular bag after Nd:YAG laser capsulotomy and to create a user-friendly software that is optimized for the output of available imaging technology, then verify the relationship between the biometric parameters of the eye and the extent of changes in the IOL position. PATIENTS AND METHODS: The study included 35 artephakic eyes that underwent laser capsulotomy. Before and at least one hour after capsulotomy, all patients underwent the following assessments: optical biometry (Lenstar LS900), OCT with anterior segment module (Optovue Avanti) and IOL photography in infrared mode using reference unit (Verion). The original software solution was designed for the graphical evaluation of the differences between centering, axial displacement and tilt of IOL. Changes in IOL position were evaluated as simple differences before and after the laser procedure and as differences in absolute values (abs).
RESULTS: The following results show the average differences in IOL position: ACD=0.02±0.23 mm (abs=0.11±0.20 mm), IOL decentration in x-axis = -0.001±0.091 mm (abs=0.065±0.063 mm), IOL decentration in y-axis =0.012±0.119 mm (abs=0.048±0.53 mm), tilt in horizontal plane TILT-H=0.11±0.83° (abs=0.51±0.66°) and tilt in vertical plane TILT-V = -0.14±0.49° (abs=0.26±0.44°). All changes were insignificant (P>>0.05). In total, 74.29% of IOLs showed a hypermetropic shift. A moderate positive correlation was found between the absolute differences in horizontal tilt and keratometry (r=0.45). Relationship with other changes and parameters were weak.
CONCLUSION: In conclusion, the use of the original software solution developed by the authors showed that changes in IOL centering, axial displacement and tilt occur after Nd:YAG capsulotomy, but average differences were insignificant. In almost 3/4 cases, there was a hypermetropic axial displacement of the IOL.

Entities:  

Keywords:  IOL axial shift; IOL centration; IOL position; IOL tilt; NdYAG laser; capsulotomy; posterior capsule opacification

Mesh:

Year:  2021        PMID: 33724262     DOI: 10.5507/bp.2021.014

Source DB:  PubMed          Journal:  Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub        ISSN: 1213-8118            Impact factor:   1.648


  6 in total

Review 1.  Effect of intraocular lens design on posterior capsule opacification.

Authors:  Wolf Buehl; Oliver Findl
Journal:  J Cataract Refract Surg       Date:  2008-11       Impact factor: 3.351

2.  Intraocular lens tilt and decentration after Nd:YAG laser posterior capsulotomy: Femtosecond laser capsulorhexis versus manual capsulorhexis.

Authors:  Esat Cinar; Berna Yuce; Fatih Aslan; Gökhan Erbakan; Cem Küçükerdönmez
Journal:  J Cataract Refract Surg       Date:  2019-11       Impact factor: 3.351

Review 3.  Use of Nd:YAG laser capsulotomy.

Authors:  Tariq M Aslam; Hilary Devlin; Baljean Dhillon
Journal:  Surv Ophthalmol       Date:  2003 Nov-Dec       Impact factor: 6.048

4.  Decentration and Tilt of Intraocular Lens after Posterior Capsulotomy.

Authors:  Mehmet Murat Uzel; Serdar Ozates; Mustafa Koc; Ayse Guzin Taslipinar Uzel; Pelin Yılmazbaş
Journal:  Semin Ophthalmol       Date:  2018-02-27       Impact factor: 1.975

5.  The effects of Nd:YAG laser capsulotomy on anterior segment parameters in patients with posterior capsular opacification.

Authors:  Zafer Oztas; Melis Palamar; Filiz Afrashi; Ayse Yagci
Journal:  Clin Exp Optom       Date:  2014-09-04       Impact factor: 2.742

6.  Effect of Nd:YAG laser posterior capsulotomy on intraocular pressure, refraction, anterior chamber depth, and macular thickness.

Authors:  Anil Parajuli; Purushottam Joshi; Prabha Subedi; Chandni Pradhan
Journal:  Clin Ophthalmol       Date:  2019-06-06
  6 in total