Literature DB >> 33722304

Effect of defibrillation on the performance of an implantable vagus nerve stimulation system.

Imad Libbus1, Scott R Stubbs2, Scott T Mazar2, Scott Mindrebo2, Bruce H KenKnight2, Lorenzo A DiCarlo2.   

Abstract

BACKGROUND: Vagus Nerve Stimulation (VNS) delivers Autonomic Regulation Therapy (ART) for heart failure (HF), and has been associated with improvement in cardiac function and heart failure symptoms. VNS is delivered using an implantable pulse generator (IPG) and lead with electrodes placed around the cervical vagus nerve. Because HF patients may receive concomitant cardiac defibrillation therapy, testing was conducted to determine the effect of defibrillation (DF) on the VNS system.
METHODS: DF testing was conducted on three ART IPGs (LivaNova USA, Inc.) according to international standard ISO14708-1, which evaluated whether DF had any permanent effects on the system. Each IPG was connected to a defibrillation pulse generator and subjected to a series of high-energy pulses.
RESULTS: The specified series of pulses were successfully delivered to each of the three devices. All three IPGs passed factory electrical tests, and interrogation confirmed that software and data were unchanged from the pre-programmed values. No shifts in parameters or failures were observed.
CONCLUSIONS: Implantable VNS systems were tested for immunity to defibrillation, and were found to be unaffected by a series of high-energy defibrillation pulses. These results suggest that this VNS system can be used safely and continue to function after patients have been defibrillated.

Entities:  

Keywords:  Autonomic regulation therapy; Defibrillation; Heart failure; Implantable cardioverter‐defibrillator; Vagus nerve stimulation

Year:  2021        PMID: 33722304      PMCID: PMC7962382          DOI: 10.1186/s42234-021-00064-w

Source DB:  PubMed          Journal:  Bioelectron Med        ISSN: 2332-8886


  8 in total

1.  2013 ACCF/AHA guideline for the management of heart failure: a report of the American College of Cardiology Foundation/American Heart Association Task Force on practice guidelines.

Authors:  Clyde W Yancy; Mariell Jessup; Biykem Bozkurt; Javed Butler; Donald E Casey; Mark H Drazner; Gregg C Fonarow; Stephen A Geraci; Tamara Horwich; James L Januzzi; Maryl R Johnson; Edward K Kasper; Wayne C Levy; Frederick A Masoudi; Patrick E McBride; John J V McMurray; Judith E Mitchell; Pamela N Peterson; Barbara Riegel; Flora Sam; Lynne W Stevenson; W H Wilson Tang; Emily J Tsai; Bruce L Wilkoff
Journal:  Circulation       Date:  2013-06-05       Impact factor: 29.690

Review 2.  2017 ACC/AHA/HFSA Focused Update of the 2013 ACCF/AHA Guideline for the Management of Heart Failure: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Failure Society of America.

Authors:  Clyde W Yancy; Mariell Jessup; Biykem Bozkurt; Javed Butler; Donald E Casey; Monica M Colvin; Mark H Drazner; Gerasimos S Filippatos; Gregg C Fonarow; Michael M Givertz; Steven M Hollenberg; JoAnn Lindenfeld; Frederick A Masoudi; Patrick E McBride; Pamela N Peterson; Lynne Warner Stevenson; Cheryl Westlake
Journal:  Circulation       Date:  2017-04-28       Impact factor: 29.690

3.  2016 ACC/AHA/HFSA Focused Update on New Pharmacological Therapy for Heart Failure: An Update of the 2013 ACCF/AHA Guideline for the Management of Heart Failure: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Failure Society of America.

Authors: 
Journal:  J Card Fail       Date:  2016-07-06       Impact factor: 5.712

4.  Autonomic regulation therapy via left or right cervical vagus nerve stimulation in patients with chronic heart failure: results of the ANTHEM-HF trial.

Authors:  Rajendra K Premchand; Kamal Sharma; Sanjay Mittal; Rufino Monteiro; Satyajit Dixit; Imad Libbus; Lorenzo A DiCarlo; Jeffrey L Ardell; Thomas S Rector; Badri Amurthur; Bruce H KenKnight; Inder S Anand
Journal:  J Card Fail       Date:  2014-09-01       Impact factor: 5.712

5.  Impact of Autonomic Regulation Therapy in Patients with Heart Failure: ANTHEM-HFrEF Pivotal Study Design.

Authors:  Marvin A Konstam; James E Udelson; Javed Butler; Helmut U Klein; John D Parker; John R Teerlink; Patricia M Wedge; Benjamin R Saville; Jeffrey L Ardell; Imad Libbus; Lorenzo A DiCarlo
Journal:  Circ Heart Fail       Date:  2019-11-14       Impact factor: 8.790

6.  ACC/AHA/ESC 2006 guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death--executive summary: A report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death) Developed in collaboration with the European Heart Rhythm Association and the Heart Rhythm Society.

Authors:  Douglas P Zipes; A John Camm; Martin Borggrefe; Alfred E Buxton; Bernard Chaitman; Martin Fromer; Gabriel Gregoratos; George Klein; Arthur J Moss; Robert J Myerburg; Silvia G Priori; Miguel A Quinones; Dan M Roden; Michael J Silka; Cynthia Tracy; Jean-Jacques Blanc; Andrzej Budaj; Veronica Dean; Jaap W Deckers; Catherine Despres; Kenneth Dickstein; John Lekakis; Keith McGregor; Marco Metra; Joao Morais; Ady Osterspey; Juan Luis Tamargo; José Luis Zamorano; Sidney C Smith; Alice K Jacobs; Cynthia D Adams; Elliott M Antman; Jeffrey L Anderson; Sharon A Hunt; Jonathan L Halperin; Rick Nishimura; Joseph P Ornato; Richard L Page; Barbara Riegel
Journal:  Eur Heart J       Date:  2006-09       Impact factor: 29.983

7.  Extended Follow-Up of Patients With Heart Failure Receiving Autonomic Regulation Therapy in the ANTHEM-HF Study.

Authors:  Rajendra K Premchand; Kamal Sharma; Sanjay Mittal; Rufino Monteiro; Satyajit Dixit; Imad Libbus; Lorenzo A DiCarlo; Jeffrey L Ardell; Thomas S Rector; Badri Amurthur; Bruce H KenKnight; Inder S Anand
Journal:  J Card Fail       Date:  2015-11-11       Impact factor: 5.712

  8 in total
  1 in total

1.  Implantable vagus nerve stimulation system performance is not affected by internal or external defibrillation shocks.

Authors:  Imad Libbus; Scott R Stubbs; Scott T Mazar; Scott Mindrebo; Bruce H KenKnight; Lorenzo A DiCarlo
Journal:  J Interv Card Electrophysiol       Date:  2021-08-31       Impact factor: 1.759

  1 in total