Literature DB >> 33714899

A study for lens capsule tearing during capsulotomy by finite element simulation.

Shaofeng Han1, Changyan He1, Ke Ma2, Yang Yang3.   

Abstract

BACKGROUND AND
OBJECTIVE: During capsulotomy, the force applied to the anterior capsule is a crucial parameter controlling capsule tears, that affects the clinical performance. This study aims to investigate the tear force in capsulotomy and analyze the effects of different tearing conditions on the tear force.
METHODS: A three-dimensional model of the human lens was constructed based on published clinical data using the finite element (FE) method. The lens model consisted of four layers: the anterior and posterior lens capsule, the cortex, and the nucleus. Distortion energy failure criterion combined with the bilinear interface law was used to express the crack propagation process at the edge of the anterior lens capsule. At the clamping position, a local coordinate system was established to parameterize the capsule tearing. The simulation results were then validated by conducting a capsulorhexis experiment using isolated porcine eyes with force-sensing forceps.
RESULTS: The simulation results showed a good agreement with the experimental data of two porcine specimens (No. 6 and 9) during a stable tearing process (p-values = 0.76 and 0.10). The mean force differences between the experimental data and the simulation were 3.10 ± 2.24 mN and 2.14 ± 1.73 mN, respectively. The tear direction with a minimum mean tear force was at θ1 = 0° and θ2 = 30°. The tear velocity was not significantly different to the variation in the tear force. However, an appropriate capsulorhexis diameter was found to contribute to the reduction of tear force.
CONCLUSIONS: The outcome of this paper demonstrates that our FE model could be used in modeling lens capsule tearing and the theoretical study of tear mechanism.
Copyright © 2021 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Anterior capsule; Capsulotomy; Crack propagation; Lens model; Tear direction; Tear force

Year:  2021        PMID: 33714899     DOI: 10.1016/j.cmpb.2021.106025

Source DB:  PubMed          Journal:  Comput Methods Programs Biomed        ISSN: 0169-2607            Impact factor:   5.428


  2 in total

1.  Computational Modeling of Ophthalmic Procedures: Computational Modeling of Ophthalmic Procedures.

Authors:  William J Foster; Brian W Berg; Steven N Luminais; Amir Hadayer; Shlomit Schaal
Journal:  Am J Ophthalmol       Date:  2022-03-28       Impact factor: 5.488

2.  An automatic drug injection device with spatial micro-force perception guided by an microscopic image for robot-assisted ophthalmic surgery.

Authors:  Zhen Li; Pan Fu; Bing-Ting Wei; Jie Wang; An-Long Li; Ming-Jun Li; Gui-Bin Bian
Journal:  Front Robot AI       Date:  2022-08-03
  2 in total