PURPOSE: Temporary balloon occlusion of the inferior vena cava to lower cardiac output is a relatively infrequently used technique to induce controlled systemic hypotension. In this technical note, we describe the feasibility, reliability, and safety of partial occlusion of right atrial inflow and the effect on systemic blood pressure during the deployment of a thoracic stentgraft.Entities:
Keywords: Occlusion balloon; Right atrial inflow occlusion; Thoracic endovascular aorta repair
Mesh:
Year: 2021 PMID: 33709270 PMCID: PMC8189963 DOI: 10.1007/s00270-020-02758-1
Source DB: PubMed Journal: Cardiovasc Intervent Radiol ISSN: 0174-1551 Impact factor: 2.740
Fig. 1Angiogram showing an inflated occlusion balloon (asterisk) for partial occlusion of right atrial inflow in a patient with a false lumen aneurysm (arrowhead). The intended proximal landing is zone 2 covering the left subclavian artery
Fig. 2Fluoroscopic image (A) demonstrates an inflated compliant occlusion balloon in the right atrium with a mixture of iodine contrast and saline. Fluoroscopic image (B) demonstrates the retracted and deformed balloon at the atriocaval junction
Patient characteristics
| Gender | Male |
| Hypertension | |
| Heart disease | Non ischemic cardiomyopathy ( |
| Proximal sealing zone | 0 ( 1 ( 2 ( 3 ( |
| Indication | Chronic dissection with false lumen aneurysm ( Acute dissection with visceral ischemia ( Peri-prosthetic bleeding after surgical arch repair ( Acute complicated aorta dissection with rupture ( Atherosclerotic aneurysm ( |
| Thoracic stentgraft | Valiant Captivia ( Custom-made fenestrated arch graft ( |
| Length of proximal stentgraft | Median 202 mm (range 150–217 mm) |
| Occlusion balloon | Reliant balloon catheter, Medtronic* compatible with 12 French introducer sheath ( CODA balloon catheter, COOK**compatible with 14 French introducer sheath ( |
| Venous access | Open ( Percutaneous ( |