Andrea Zaffora1, Francesco Di Franco1, Danilo Virtù1, Francesco Carfì Pavia1, Giulio Ghersi2, Sannakaisa Virtanen3, Monica Santamaria1. 1. Dipartimento di Ingegneria, Università degli Studi di Palermo, Viale delle Scienze, Palermo 90128, Italy. 2. Dipartimento di Scienze e Tecnologie Biologiche, Università degli Studi di Palermo, Chimiche e Farmaceutiche (STEBICEF), Viale delle Scienze, Palermo 90128, Italy. 3. Chair for Surface Science and Corrosion, Department of Materials Science and Engineering, University of Erlangen-Nürnberg, Erlangen 91058, Germany.
Abstract
Coatings were grown on the AZ31 Mg alloy by a hard anodizing process in the hot glycerol phosphate-containing electrolyte. Anodizing conditions were optimized, maximizing corrosion resistance estimated by impedance measurements carried out in Hank's solution at 37 °C. A post anodizing annealing treatment (350 °C for 24 h) allowed us to further enhance the corrosion resistance of the coatings mainly containing magnesium phosphate according to energy-dispersive X-ray spectroscopy and Raman analyses. Gravimetric measurements revealed a hydrogen evolution rate within the limits acceptable for application of AZ31 in biomedical devices. In vitro tests demonstrated that the coatings are biocompatible with a preosteoblast cell line.
Coatings were grown on theAZ31Mg alloy by a hard anodizing process in the hot glycerol phosphate-containing electrolyte. Anodizing conditions were optimized, maximizing corrosion resistance estimated by impedance measurements carried out in Hank's solution at 37 °C. A post anodizing annealing treatment (350 °C for 24 h) allowed us to further enhance the corrosion resistance of the coatings mainly containing magnesium phosphate according to energy-dispersive X-ray spectroscopy and Raman analyses. Gravimetric measurements revealed a hydrogen evolution rate within the limits acceptable for application of AZ31 in biomedical devices. In vitro tests demonstrated that the coatings are biocompatible with a preosteoblast cell line.
Metallic
biomaterials are nowadays widely used in human body as
implants (e.g., artificial joints, stents, and bone plates). However,
serious concerns could arise using permanent metallic implants since
they could cause harmful body reactions (e.g., thrombosis, physical
irritation, restenosis, etc.) and also an inability to adapt to the
growth and changes in thehuman body.[1,2] In these cases,
additional surgery is needed to remove metallic implants but only
after the tissue healing function has been completed. Biodegradable
implants can represent a solution because they can be implanted for
an appropriate period to fix and then disappear, avoiding undesirable
body reactions.[3,4] Magnesium alloys have attracted
increasing attention over the last two decades, of both industrial
and academic world, as a suitable candidate for biodegradable biomedical
devices (e.g., vascular stents and bone plates)[5,6] due
to their peculiar properties, such as high strength/weight ratio,
good thermal and electrical conductivities, excellent vibration, and
shock absorption.[7] Indeed, with respect
to materials such asstainless steel, titanium alloys, and Cr–Co
based alloys, which are typically employed, Mg and its alloys have
a low density (1.74–2.0 g cm–3) and an elastic
modulus between 41 and 45 GPa, and once implanted in thehuman body
(in vivo), the ions and/or particles released as
a consequence of corrosion phenomena are not detrimental to the body.[8]However, the potential clinical applications
of Mg alloys have
been hindered by their poor corrosion resistance. Due to the very
negative reduction potential of Mg and its poor corrosion resistance
in chloride-containing environments (human body fluids or blood plasma),
the degradation rate of magnesium and its alloys is so high that the
mechanical integrity before the diseased or damaged bone tissue healed
is not always maintained. Moreover, since water reduction is the common
cathodic process during corrosion of Mg and its alloys, the high corrosion
rate implies a high H2 evolution rate, with consequent
detrimental gas pocket formation around the implant and alkalization
occurring in the vicinity of the corroding surface as possibly being
deleterious for the surrounding biological environment.[9]The simplest way to slow down corrosion
of Mg and Mg alloys is
to form a coating on themagnesium substrate to provide a barrier
toward the contact between the substrate and the environments. There
are many coating technologies which can be used to coat themagnesium
substrate,[10] such as chemical conversion
coating,[11−13] physical vapor deposition,[14,15] laser surface treatment,[15,16] and anodic oxidation.[17−20] Among these technologies, the latter is one of the most effective
and popular methods, even if the growth of protective anodic layers
on Mg and Mg alloys is difficult due to the unfavorable Pilling–Bedworth
ratio (PBR) for MgO. The latter is defined as the ratio of the volume
of the elementary cell of a metaloxide to the volume of the elementary
cell of the corresponding metal oxidized to produce theoxide, that
is:[21]where V is the molar volume, M is the molecular weight,
ρ is the density, and n is the number of metal
moles for 1 mole of formed oxide.
Theoxide layer would be unprotective if thePBR is less than unity
because the film that forms on themetal surface is porous and/or
cracked, as in the case of magnesium. Thus, in order to enhance the
corrosion resistance of Mg and Mg alloys by anodic oxide layers, incorporation
from the electrolytic bath of species other than O2– during the anodizing process must be promoted. The anodizing occurs
according to a high-field mechanism, thanks to the superimposition
of very high electric field (in the order of a few MV cm–1), which allows a solid-state migration process of the oxidized metallic
cations from the substrates as well as of the anion from the electrolyte,
namely, O2– in aqueous solution derived by water
deprotonation. Other anions added to the electrolytic solution can
also be incorporated even if in a smaller extent than oxygen ions
highly available in an aqueous solution. Replacing wateras a solvent
as well as adding anions that can favor the biocompatibility of the
coating is a promising strategy.In the present work, anodizing
of theAZ31 alloy was carried out
in the hot glycerol (HG) electrolyte containing K2HPO4 and K3PO4, in an attempt to induce
the growth of a magnesium phosphate protective layer on the surface
of AZ31 with a more favourable PBR than that of MgO. The design strategy
for the preparation of these coatings is schematically described in Figure .
Figure 1
Sketch of the anodizing
process setup. Inset: Movement of metallic
cations and oxygen anions during anodic coating growth.
Sketch of the anodizing
process setup. Inset: Movement of metallic
cations and oxygen anions during anodic coating growth.In the attempt to seal the pores of the anodic layers, to
improve
their biocompatibility, and to tune the biodegradation rate, the growth
of hydroxyapatite (HAP) on thesample surface was induced by a dip
coating procedure. The structure and composition of the resultant
films were studied by scanning electron microscopy (SEM), energy-dispersive
X-ray spectroscopy (EDS), and Raman spectroscopy before and after
the immersion. The corrosion resistance of the resulting composite
coatings was characterized in Hank’s solution (HS) at 37 °C
by open circuit potential (OCP) measurements, electrochemical impedance
spectroscopy (EIS), and by recording polarization curves. Accurate
measurement of thehydrogen evolution (HE) was obtained by a method
based on the measurement of the hydrostatic force resulting from the
accumulation of hydrogen in a submerged container.[22]In vitro studies were carried out to evaluate
the cytocompatibility of the anodized AZ31 samples.
Materials and Methods
Materials
Composition of themagnesiumAZ31 alloy is reported in Table .
Table 1
Composition (wt %) of the As-Received
Mg AZ31 Alloy
Al
Zn
Mn
Si
Fe
Cu
Ni
Mg
2.89
0.92
0.05
0.01
0.004
0.002
0.001
balance
Preparation
of AZ31 Coatings and the Anodizing
Process
Before any measurements, AZ31 samples were mechanically
polished with SiC abrasive papers with grit sizes between P1000 and
P2400. Afterward, all samples were cleaned with acetone in an ultrasonic
cleaner for 5 min and then rinsed with distilled water.Anodizing
of AZ31 electrodes was carried out in a stirred HG solution containing
0.6 M K2HPO4 and 0.2 M K3PO4, with controlled temperature conditions in thegalvanostatic unipolar
mode (i.e., at a constant current density) using a two-electrode configuration,
where the working electrode was theAZ31 sample and the counter electrode
was a Pt net with a high surface area. The effect of different operating
parameters on the final properties of the coatings was studied: (i)
anodizing time (60 s, 30, and 60 min), (ii) bath temperature (100,
160, and 200 °C), and (iii) current density (0.5, 2, and 8 mA
cm–2). Post anodizing thermal treatment was carried
out at 350 °C for 24 h. Each anodic growth was repeated at least
three times.
Electrochemical Characterization
To investigate the influence of operating parameters on the properties
of the coating on AZ31 samples, OCP, EIS, and polarization curve measurements
were carried out using a three-electrode configuration, with an Ag/AgCl/sat.
KCl (0.197 V vs SHE) reference electrode, while the
working and counter electrodes were the same as that for the anodizing
process. The characterization electrolyte was theHS, whose composition
is reported in Table .
Table 2
Composition of Hank’s Solution
(pH = 6.67)[23]
component
concentration [g L–1]
NaCl
8
KCl
0.4
NaHCO3
0.35
NaH2PO4·H2O
0.25
Na2HPO4·H2O
0.06
CaCl2·2H2O
0.19
MgCl2
0.19
MgSO4·7H2O
0.06
glucose
1
The bath
temperature was kept constant at 37 ± 1 °C.
Impedance spectra were recorded by superimposing to the constant electrode
potential (OCP) an a.c. signal of 10 mV in a frequency range between
100 kHz and 100 mHz. Polarization curves were recorded in the electrode
potential range between −1.7 V Ag/AgCl and −1.1 V Ag/AgCl
with a scan rate of 1 mV s–1. All electrochemical
measurements were carried out using a PARSTAT 2273 instrument connected
to a PC for data acquisition. EIS spectra were then fitted with ZSimpWin
software. Each experiment was repeated three times.
SEM, EDX, XRD, and Raman Spectroscopy Analyses
The
morphology of thesamples was observed using SEM using a Versa
3D microscope (FEI, Thermo Fischer). The cross sections were prepared
by mounting thesamples vertically in epoxy resin and mechanically
grinding near parallel to the coating/substrate interface direction
with emery paper down to 2400 grit. Images were taken at a voltage
of 30 kV. EDX microanalysis was carried out with a XFlash 6l10 EDX
probe (Bruker) integrated in the microscope. The spectra, obtained
at the voltage of 30 kV, were processed with Esprit software for qualitative
and quantitative analyses. The same software was used for EDX mapping.
The maps were obtained at the voltage of 30 kV with an acquisition
time of 300 s.X-ray diffraction (XRD) measurements were performed
using a PanAnalytical Empyrean diffractometer with a Cu anode (Cu
Kα radiation, λ = 0.15405 nm) equipped with a PIXCel1D
detector (voltage: 40 kV, current: 40 mA). The XRD patterns were collected
over the 2θ angle range of 10–90°.Micro-Raman
analysis was performed through a Renishaw inVia Raman
microscope spectrometer equipped with a microprobe (50×) and
a CCD detector with a Nd:YAG laser with a wavelength of 532 nm.
Measurement of H2 Evolution during
Corrosion in HS
Thesample was mounted below a bottle completely
immersed in HS. The bottle was closed at the bottom through a deformable
diaphragm to allow collection of the evolved hydrogen bubbles. The
bottle and thesample were mounted on a support that was hooked into
a precision laboratory balance (Mettler Toledo XS1003S). An inert
weight was added to the bottle ensuring stable submersion. With this
arrangement, when hydrogen evolves from the electrode surface, it
is collected by the bottle. Due to the hydrostatic force, the weights
of the bottle and sample decrease proportionally to the volume of
evolved hydrogen. Each measurement of H2 evolution was
repeated three times.
Cytotoxicity Tests
For indirect cytotoxicity
tests, thesamples (coated and not-coated) were rinsed with Milli-Q
water and then sterilized by immersion in ethanol for 24 h followed
by UV ray treatment for 2 h (1 h per side). Samples were then incubated
with Dulbecco’s modified Eagle medium (DMEM, Sigma-Aldrich)
at 37 °C for 24 h with a volume/surface ratio of 5 ml/cm2.[24] Afterward, theas-treated media
were collected in 50 mL of falcon and utilized for thecytotoxicity
test. MC3T3-E1 preosteoblastic cells purchased from Sigma-Aldrich
(ECACC) were cultured in DMEM added to 10% fetal bovine serum, 1%
glutamine, and 1% antibiotic at 37 °C and in a 5% CO2 atmosphere. 104 osteoblastic cells were seeded into the
wells of a 24-well culture plate (10 mm diameter) and incubated with
normal DMEM at 37 °C and 5% CO2. After 24 h, the medium
was replaced with treated media. Normal DMEM was used as the positive
control to evaluate the effect of coatings.The proliferation
rate was evaluated using an AlamarBlue cell viability reagent (invitrogen).
Each well was incubated at 37 °C and 5% CO2 for 3
h with 500 μL of AlamarBlue reagent (10×) diluted (1:10)
in DMEM. The resulting fluorescence was read on a plate reader at
an excitation wavelength of 530/25 (peak excitation is 570 nm) and
an emission wavelength of 590/35 (peak of emission is 585 nm). Cytotoxicityassays were carried out after 0, 1, 4, and 7 days of culture. Each
cytotoxicity test was repeated three times.
Results and Discussion
Growth Kinetics and Anodizing
Process Optimization
Figure shows the
dependence of cell voltage on time recorded during the anodizing of
AZ31 carried out galvanostatically at 2 mA cm–2 at
160 °C for 60 min in HG solution.
Figure 2
(a) Dependence of cell
voltage on time during AZ31 anodizing at
2 mA cm–2 for 60 min at 160 °C. Inset: SEM
micrographs related to samples anodized for 100 s and 60 min. (b)
XRD patterns related to samples anodized at 2 mA cm–2 for 100 s, 30 min, and 60 min.
(a) Dependence of cell
voltage on time during AZ31 anodizing at
2 mA cm–2 for 60 min at 160 °C. Inset: SEM
micrographs related to samples anodized for 100 s and 60 min. (b)
XRD patterns related to samples anodized at 2 mA cm–2 for 100 s, 30 min, and 60 min.The effect of anodizing time on the properties of the coating was
studied by interrupting the anodizing process at 100 s, 30 min, and
60 min. The cell voltage rises linearly during the early stage of
anodic polarization with a slope of ∼1.2 V s–1 up to ∼120 V, followed by a much slower increase accompanied
by oscillations due to the onset of dielectric breakdown phenomena.
The slope, dV/dt, of the linear
part of the curve can be linked to the film growth parameters according
to the following relationship[18]where i is the anodic current
density, Ed is the electric field strength, M is the molecular weight of the film constituent, z is the number of electrons, F is the
Faraday constant, and ρ is the film density. As described in
detail in Appendix A, eq allows us to estimate the barrier layer thickness
corresponding to two limiting cases, that is, 110 nm assuming the
formation of MgO (no foreign anions are incorporated from the electrolyte
apart O2– anions) and 335 nm assuming the formation
of Mg3(PO4)2 [only (PO4)3– anions are incorporated from the electrolyte].
The second part of the anodizing process allowed the formation of
a thicker coating, thanks to the formation in a hard anodizing regime
of an outer porous layer.[25]Figure shows the
EIS spectra recorded for anodized AZ31 samples in HS at 37 °C
at theOCP after 1 h of immersion. For comparison, the EIS spectrum
recorded for as-polished AZ31 is also reported. The spectra in the
Nyquist representation for the coated alloy look like depressed semicircles.
Figure 3
EIS spectra
in the Nyquist representation recorded at OCP in HS
for the as-polished and anodized AZ31 samples at different times.
Inset: Zoom of EIS spectra at low real and imaginary parts. Continuous
lines: fitting lines.
EIS spectra
in the Nyquist representation recorded at OCP in HS
for theas-polished and anodized AZ31 samples at different times.
Inset: Zoom of EIS spectra at low real and imaginary parts. Continuous
lines: fitting lines.It is interesting to
note that two different electrical equivalent
circuits (EECs) were used to fit EIS spectra, and they are shown in Figure .
Figure 4
EECs used to model the
electrochemical behavior of the systems
comprising (a) only the barrier layer and (b) barrier + porous layers.
EECs used to model the
electrochemical behavior of the systems
comprising (a) only the barrier layer and (b) barrier + porous layers.In fact, in the case of theAZ31 sample anodized
for 100 s, best
fitting was reached using an EEC comprising just one time constant,
that is, a parallel (RbQb) in series with Rel (electrolyte
resistance), where Q is a constant phase element
modelling the nonideal capacitance of the single-layer coating. This
is due to the barrier nature of the anodic film grown for 100 s, and
the use of the EEC, as shown in Figure a, is in agreement with that reported in the literature,
for which the electrochemical behavior of barrier-like anodic films
can be successfully modelled by a simple parallel RQ.[26−28] Conversely, in the case of AZ31 samples anodized for longer times
(e.g., 30 and 60 min), a more complex EEC is needed to model the electrochemical
behavior of anodic films immersed in HS. In particular, the EEC used
is shown in Figure b, where the parallel (RbQb) is in series with Rp representing
the outer layer resistance, that is, the electrolyte resistance inside
the pores of the anodic films. This part of the electrical circuit
is in parallel with a Qp element representing
the outer porous layer (nonideal) capacitance. It was then added in
series as a final element with the electrolyte resistance Rel. Best fitting parameters, estimated by using
EECs, as shown in Figure a,b, are reported in Table .
Table 3
Fitting Parameters Related to EIS
Spectra Shown in Figure
anodizing
time
Rel [Ω cm2]
Rp [kΩ cm2]
Qp [S sn cm–2]
n
Rb [kΩ cm2]
Qb [S sn cm–2]
n
χ2
100 s
15
7
2.0 × 10–5
0.76
2.5 × 10–3
30 min
15
0.6
4.9 × 10–7
0.76
64
2.5 × 10–6
0.71
7.4 × 10–4
60 min
15
10
8.2 × 10–6
0.70
3
1.5 × 10–4
1
1.1 × 10–2
The highest overall resistance corresponds to an anodizing time
of 30 min, suggesting that the corrosion rate is slowed down synergistically
by both the barrier inner layer and the porous outer layer. The formation
of anodic films during spark anodizing is more complex than the growth
of the anodic film under high field mechanism. As reported above (see eq ), in the latter case,
the thickness of the barrier film is directly proportional to the
applied voltage. This statement is no longer true for anodizing occurring
under sparking regime since thickening of the inner (so called dense
layer) and outer porous layers is determined by several types of discharge
and by plasma reactions occurring as a consequence of the discharges
themselves.[29] The fitting of EIS spectra
suggests that a longer duration of the sparking anodizing may induce
damage in the inner layer and induce the formation of a thicker porous
layer with a detrimental effect on the corrosion resistance.It is interesting to note, by looking at the inset of Figure , that an inductive
loop is present in the low-frequency range of the EIS spectrum of
as-polished AZ31. The presence of an inductive loop in the EIS spectrum
is typically related to the electrochemical processes that involve
more than one reaction step, where one step comprises the formation
of adsorbed species on the surface leading to a dependence of reaction
kinetics on the electrode surface coverage.[26,30,31] This behavior is usually observed in the
case of pure Mg and Mg-alloy electrodes, for which theoxide/hydroxide
film that spontaneously grows on the surface is quite incoherent and
does not cover all of the substrate surface, not offering good corrosion
protection to themetal, mostly in acidic or neutral environment,
asHS. Moreover, the cathodic process usually coupled to theMg oxidation
is thewater reduction that leads to gaseous HE that typically occurs
following a multistep kinetic path, involving the presence of adsorbed
species.[32,33]Anodizing process was also carried
out at different temperatures
(i.e., 100 and 200 °C), as shown in Figure a, where the comparison of the growth curves
is reported for a galvanostatic polarization at 2 mA cm–2 for 30 min.
Figure 5
(a) Dependence of cell potential on time during AZ31 anodizing
at 2 mA cm–2 for 30 min at different temperatures.
(b) EIS spectra in the Nyquist representation recorded at OCP in HS
for anodized AZ31 samples at different temperatures. Continuous lines:
fitting lines.
(a) Dependence of cell potential on time during AZ31 anodizing
at 2 mA cm–2 for 30 min at different temperatures.
(b) EIS spectra in the Nyquist representation recorded at OCP in HS
for anodized AZ31 samples at different temperatures. Continuous lines:
fitting lines.Regardless of the anodizing bath
temperature, the cell voltage
followed the same dependence on time shown in Figure , that is, an initial linear dependence related
to the growth of a barrier layer followed by voltage oscillations
due to sparking phenomena. It is noteworthy to mention that the slope
of the linear part of the growth curve strongly depends on the bath
temperature, suggesting a different growth efficiency and leading
to a different structure and/or a different barrier layer thickness
(see eq ). Moreover,
at the highest temperature (namely 200 °C), the dissolution rate
became too high to keep the final formation voltage (as in the case
of growth at 100 and 160 °C), with a decrease of cell voltage
value for t > 12 min. In spite of the corrosion
potential
values being the same regardless of the anodizing bath temperature
(−1.48 V vs Ag/AgCl for the anodic film grown
at 160 °C and −1.50 V vs Ag/AgCl for
the coatings grown at 100 and 200 °C), a clear difference in
the electrochemical behavior could be assessed by looking at the EIS
spectra, recorded at OCP, as shown in Figure b. For the fitting procedure, EEC was used,
as shown in Figure b, since the structure of the coating obtained for 30 min was porous,
regardless of the bath temperature during the anodic film growth.
EIS fitting parameters are reported in Table .
Table 4
Fitting Parameters
Related to EIS
Spectra Shown in Figure b
anodizing temperature
Rel [Ω cm2]
Rp [kΩ cm2]
Qp [S sn cm–2]
n
Rb [kΩ cm2]
Qb [S sn cm–2]
n
χ2
100 °C
17
0.20
5.3 × 10–6
0.69
14
1.6 × 10–6
0.89
4.8 × 10–3
160 °C
15
0.6
4.9 × 10–7
0.76
64
2.5 × 10–6
0.71
7.4 × 10–4
200 °C
13
0.06
2.2 × 10–5
0.63
5.7
3.3 × 10–6
0.93
2.4 × 10–3
The best performance, in terms of overall system resistance, was
shown by the coating grown at 160 °C, while the worst one wasassociated with the coating grown at 200 °C (Rb = 5.7 kΩ cm2), as the temperature of
the anodizing bath plays a major role in the quality of the coating
grown on AZ31 samples.AZ31 coupons were also anodized for 30
min at 0.5, 2, and 8 mA
cm–2 at 160 °C in order to assess the effect
of current density on the quality of the prepared coatings (see Figure a). As expected,
the higher the slope of the voltage–time curve, the higher
the applied current density, and at 8 mA cm–2, the
voltage increases significantly for a long anodizing time probably
due to intense discharges, as confirmed by the SEM images (see Figure c) showing the formation
of wider pores when AZ31 is anodized at this high current density.
Notably, the XRD patterns for the coating grown at 8 mA cm–2 confirmed the presence of crystalline Mg3(PO4)2 (according to ICSD card: 020796, see Figure d), whose reflections are missing
for the coating grown at 0.5 mA cm–2 probably due
to the formation of a thinner and/or less crystalline layer. Figure e shows the EIS spectra
recorded in HS at 37 °C after 1 h of immersion at theOCP. As
for all samples anodized for 30 min, the best EIS spectra fitting
was obtained by using the EEC that comprise the presence of an outer
porous layer in the structure of the coatings.
Figure 6
(a) Dependence of the
cell potential on time during AZ31 anodizing
at 160 °C for 30 min at different current densities. SEM micrographs
related to samples anodized at 160 °C for 30 min at (b) 0.5 and
(c) at 8 mA cm–2 with corresponding (d) XRD patterns.
(e) EIS spectra in the Nyquist representation recorded at OCP in HS
for anodized AZ31 samples at different current densities. Continuous
lines: fitting lines. Inset: Surface of the AZ31 coupons soon after
anodizing at different current densities.
(a) Dependence of the
cell potential on time during AZ31 anodizing
at 160 °C for 30 min at different current densities. SEM micrographs
related to samples anodized at 160 °C for 30 min at (b) 0.5 and
(c) at 8 mA cm–2 with corresponding (d) XRD patterns.
(e) EIS spectra in the Nyquist representation recorded at OCP in HS
for anodized AZ31 samples at different current densities. Continuous
lines: fitting lines. Inset: Surface of theAZ31 coupons soon after
anodizing at different current densities.The highest resistance corresponds to an anodizing current density
of 2 mA cm–2, while the worst result was shown by
theAZ31 sample anodized at 8 mA cm–2 (see best
fitting parameters reported in Table ).
Table 5
Fitting Parameters Related to EIS
Spectra Shown in Figure e
anodizing current density
Rel [Ω cm2]
Rp [kΩ cm2]
Qp [S sn cm–2]
n
Rb [kΩ cm2]
Qb [S sn cm–2]
n
χ2
0.5 mA cm–2
20
0.16
3.1 × 10–7
0.81
31
4.8 × 10–6
0.70
9.6 × 10–4
2 mA cm–2
15
0.6
4.9 × 10–7
0.76
64
2.5 × 10–6
0.71
7.4 × 10–4
8 mA cm–2
20
15
7.2 × 10–6
0.79
1.4
7.7 × 10–4
1
6.5 × 10–3
The surface of theAZ31 coupons soon after anodizing is uniformly
covered by a coating, whose color is a function of the anodizing current
density (see inset of Figure e). From the analysis reported in this section, best results,
in terms of corrosion resistance in HS, were shown by AZ31 samples
anodized at 2 mA cm–2 for 30 min with an anodizing
bath temperature of 160 °C.Figure shows the
SEM cross section of an AZ31 sample anodized in the optimized anodizing
conditions with theassociated EDS map of the elements contained in
the anodic film.
Figure 7
SEM cross section of the AZ31 sample anodized in the optimized
anodizing conditions with the associated EDS map of the elements contained
in the anodic coating.
SEM cross section of theAZ31 sample anodized in the optimized
anodizing conditions with theassociated EDS map of the elements contained
in the anodic coating.The coating thickness
resulted to be ∼12 μm. From
the elemental analysis (at. %), it is possible to assess the composition
of the coating on theAZ31 sample that is very close to that of Mg3(PO4)2, with P and O quite uniformly
distributed across the anodic layer.
Post
Anodizing Annealing Process
Once the anodizing process was
optimized in terms of current density,
anodizing time, and bath temperature, we studied the effect of a post
anodizing annealing process on the coatings’ morphology and
on their corrosion resistance. AZ31 coupons anodized at 2 mA cm–2 for 30 min at 160 °C were annealed under air
exposure at 350 °C for 24 h. The effect of the thermal treatment
on the coating microstructure was evaluated by SEM. Figure a,b shows the surface morphology
of theas-anodized and annealed coupons, respectively, while the insets
show the color of the coatings before and after the thermal treatment.
Moreover, the thermal treatment leaves thesample surface free of
the deposits present onto thesample soon after the anodizing process.
Figure 8
Scanning
electron micrographs showing the morphologies of anodic
coatings formed on AZ31 samples (a) before and (b) after thermal treatment.
Inset: Surface of the anodic coatings.
Scanning
electron micrographs showing the morphologies of anodic
coatings formed on AZ31 samples (a) before and (b) after thermal treatment.
Inset: Surface of the anodic coatings.Thermal treatment also had an effect on the corrosion resistance
of the alloy samples, as better evaluated by EIS. EIS spectrum of
the annealed coupon recorded in HS at 37 °C after 1 h of immersion
is shown in Figure and best fitting parameters (see Table ) were obtained by using the EEC, as shown
in Figure b.
Figure 9
EIS spectra
in the Nyquist representation recorded at OCP in HS
for anodized AZ31 samples before and after the thermal treatment.
Continuous lines: fitting lines.
Table 6
Fitting Parameters Related to EIS
Spectra Shown in Figure
sample
Rel [Ω cm2]
Rp [kΩ cm2]
Qp [S sn cm–2]
n
Rb [kΩ cm2]
Qb [S sn cm–2]
n
χ2
as-anodized
15
0.6
4.9 × 10–7
0.76
64
2.5 × 10–6
0.71
7.4 × 10–4
annealed
3.6
2.1
2.3 × 10–7
0.83
440
4.3 × 10–7
0.64
5.2 × 10–4
EIS spectra
in the Nyquist representation recorded at OCP in HS
for anodized AZ31 samples before and after the thermal treatment.
Continuous lines: fitting lines.The corrosion resistance was enhanced
by one order of magnitude
due to thermal treatment with respect to that estimated for AZ31 soon
after anodizing. A more compact morphology is representative of a
kind of sealed structure that diminishes the contact of the underneath
barrier layer, with highly aggressive HS leading to a reduction of
corrosion phenomena.We also evaluated the effect of longer
immersion time in HS on
the composition of the coating, especially in terms of at. % of Ca
and P, that is, the main elements of HAP compound. At. % obtained
by EDX analysis is reported in Table as a function of the immersion time from 3 to 14 days.
Table 7
At. % of Ca and P Components in the
Anodized AZ31 Sample after Different Immersion Times in HS
immersion time [days]
Ca [at. %]
P [at. %]
Ca/P ratio
3
6.04
15.33
0.39
7
9.04
14.92
0.61
14
13.02
16.98
0.77
The ratio Ca/P increases by increasing
the immersion time in HS,
from 0.39 to 0.77. These values are quite far from the ratio Ca/P
contained in HAP, that is, 1.67, but are affected by the high P signal
due to the P incorporated into coatings during the anodizing process.
The effect of immersion in HS was also assessed by looking at the
morphology of anodized AZ31 samples that is shown in Figure for the studied immersion
times.
Figure 10
Scanning electron micrographs showing the morphologies of anodic
coatings, after annealing, formed on AZ31 samples after (a) 3 days,
(b) 7 days, and (c) 14 days of immersion in HS. (d) Raman spectrum
related to the anodic coating after 14 days of immersion in HS.
Scanning electron micrographs showing the morphologies of anodic
coatings, after annealing, formed on AZ31 samples after (a) 3 days,
(b) 7 days, and (c) 14 days of immersion in HS. (d) Raman spectrum
related to the anodic coating after 14 days of immersion in HS.After 3 days of immersion (see Figure a), the surface looked still
compact as
that shown in Figure b, while a classical “mud” structure developed by increasing
the immersion time in HS (i.e., 7 and 14 days, see Figure b,c, respectively). Cracks
present on the surface of samples can be explained by taking into
account (i) the residual-stress release during the hard anodizing
process when the coating is immersed in HS, (ii) theHE due to the
corrosion of magnesium alloy,[34,35] and/or (iii) the penetration
of HS into anodic layer pores and the following formation of HAP with
higher volume.[35] The formation of HAP is
confirmed also by the Raman spectrum, as shown in Figure d, related to the anodized
AZ31 sample immersed for 14 days in HS. In particular, PO43– features in HAP are characterized by the ν1 symmetric stretching (P–O) mode at 960 cm–1 and by the ν2 bending (O–P–O)
mode at about 430–450 cm–1.[36,37]
Evaluation of H2 Evolution Rate
A significant issue about the usage of magnesium alloys in biomedical
applications is that theHE reaction (HER) is coupled to theMg corrosion
(dissolution) process according to the following reactionthat is,
water reduction reaction (WRR).
Actually, oxygen reduction reaction can also be considered coupled
to theMg corrosion process, but at the corrosion potential of Mg
alloys, WRR has to be considered as the dominant cathodic reaction.
Gaseous HE could be dangerous depending on its evolution rate. If
it is sufficiently high, hydrogen bubbles can accumulate in gas pockets
next to the implant, which can lead to a delay in thehealing of the
surgery region and/or even to thenecrosis of tissues.[38] In the worst case, hydrogen bubbles can also
be a serious issue for the blood circulating system. On the contrary,
if H2 generation is sufficiently slow, gas can be transported
away from the implant site avoiding any possible gas build-up. Another
issue related to reaction is the environment alkalization that, although could be beneficial
for thepassivation of Mg alloy surface, can also lead to cell death.[39]For these reasons, it is crucial to study
HER and to measure the
amount of hydrogen evolved. Typically, potentiodynamic measurements
are carried out in order to estimate the corrosion current density, icorr, and therefore, the amount of H2 generated in the condition of free corrosion in theHS electrolyte.
In Figure polarization
curves recorded for as-polished AZ31, anodized, and thermal treated
samples are shown.
Figure 11
Polarization curves, recorded at 1 mV s–1, for
as-polished, anodized, and after annealing AZ31 samples.
Polarization curves, recorded at 1 mV s–1, for
as-polished, anodized, and after annealing AZ31 samples.The thermal treated sample showed the lowest cathodic current
density,
the lowest passivity current density, and the highest pitting potential.
The highest icorr and, thus, the cathodic
current density was measured for theas-polished AZ31 sample. Actually,
the electrochemical estimate of icorr from
the polarization curves of Mg and Mg alloys suffers from the uncertainty
due to the negative difference effect or the anomalous HEphenomenon.[9,30,40,41] In fact, when Mg and Mg alloys are polarized anodically with respect
to their corrosion potential, higher dissolution rates are recorded,
but theHER rate increases by increasing the anodic polarization.
Therefore, a more accurate method to estimate theHE rate is mandatory.
In this regard, we used the method described by Strebl and Virtanen
that is based on the simple measurement of the hydrostatic force resulting
from the accumulation of hydrogen in a submerged container (see Section for experimental
details).[22] As expected, hydrogen evolved
during the measurement time led to a decrease in the recorded weight
by the instrument due to a buoyant force that, according to the Archimedes’
principle, is equal to the weight of the displaced fluid volume. The
amount (volume) of hydrogen evolved as function of time can be derived
according to the following equation[22]where Wbalance is the recorded weight, g is the gravitational
acceleration, and ρsolution is the density of theHS electrolyte. The volume of hydrogen evolved normalized for thesample area as a function of time is then reported in Figure a.
Figure 12
(a) Volume of hydrogen
evolved in 15 days for an anodized and thermal
treated AZ31 sample. (b) Volume of hydrogen evolved for an anodized
and thermal treated AZ31 sample after different immersion times in
HS.
(a) Volume of hydrogen
evolved in 15 days for an anodized and thermal
treated AZ31 sample. (b) Volume of hydrogen evolved for an anodized
and thermal treated AZ31 sample after different immersion times in
HS.As it is possible to see, the
volume of hydrogen evolved increases
during the first 50 h of measurements and then decreases (i.e., recorded
weight by the balance increases). This is probably due to the dissolution
of gaseous H2 into the solution, thus not exerting a buoyant
force anymore. Because of this phenomenon, this technique seems not
suitable to give reliable results about the amount of hydrogen evolved
for long-run measurements. To overcome this issue, a discontinuous
measurement of the volume of evolved H2 was carried out:
thesample of anodized and thermal treated AZ31 was immersed in HS
and, after a certain time, was removed from that solution and immersed
in HS for theHE rate evaluation setup. Then, the weight was recorded
for about 6 h before thesample was immersed again in another beaker
containing HS. In this way, we reached longer measurement times than
those of continuous measurement. Figure b shows the volume of hydrogen evolved for
the discontinuous measurement at different immersion times. The rate
of HE was 0.007 mL cm–2 day–1 soon
after the immersion in HS, then 0.011, 0.016, 0.010, and 0.008 mL
cm–2 day–1 after 3, 5, 7, and
14 days, respectively. These HE rates were lower or of the same order
of magnitude than 0.01 mL cm–2 day–1 that was reported as the tolerated rate to consider the treated
AZ31 sampleas the candidate biodegradable material for further tests
in thehuman body.[38] The highest recorded
HE rate resulted, in any case, lower or of the same order of magnitude
of other HE rates reported in the literature for Mg and Mg-based electrodes
for biomedical devices.[42−45]Finally, we investigated
thecytotoxicity of the anodized and thermal treated AZ31 samples
that are possible candidates for biomedical applications. Cytotoxicity
tests were carried out in vitro using the preosteoblastic
cell line, MC3T3-E1. Cell culture was carried out using media derived
from incubation with samples for 24 h at an established volume/surface
ratio. Cell viability assays were based on following the growth up
to 7 days of culture, whose results are shown in Figure .
Figure 13
Viability of MC3T3-E1
cells grown with not-treated medium (control)
and bare and coated AZ31 samples.
Viability of MC3T3-E1
cells grown with not-treated medium (control)
and bare and coated AZ31 samples.From the graph, it can be observed that in all three cases, the
number of cells increases by increasing the time [from 5 × 104 (seeding days) to about 3 × 105 (7 days)],
without statistically significant difference among thesamples at
each time point. Furthermore, it is important to underline that in
all cases, the viability of the cells grown with the incubated media
was abundantly over 70% with respect to the control viability, thus
confirming the noncytotoxicity of the tested materials, according
to the followed ISO standard.[46]
Conclusions
Anodizing of AZ31Mg alloys was performed
in HG electrolyte containing
K2HPO4 and K3PO4 salts
to grow a corrosion resistant and biocompatible coating that can be
used in biomedical devices. Anodic film onto themetallic substrate
had a two-layer structure, that is, an inner barrier layer and a thicker
outer porous layer. Theas-formed coatings were characterized in HS
in order to verify their compatibility with thehuman body environment.
The anodizing process was optimized in terms of a suitable current
density (2 mA cm–2), process time (30 min), and
bath temperature (160 °C) to maximize the corrosion resistance
of AZ31 coupons. A thermal treatment at 350 °C for 24 h was also
performed further enhancing thesamples’ corrosion resistance,
sealing the outer porous layer, making it more compact, and thus reducing
the contact time between the substrate and the aggressive environment,
that is, HS.The amount of hydrogen evolved from corrosion of
AZ31 samples was
measured with a simple gravimetric method rather than estimated from
electrochemical potentiodynamic measurements. H2 evolution
rate resulted to be lower than the tolerated rate to consider samples
as candidate biodegradable materials. Cytotoxicity tests also demonstrated
the biocompatibility of treated AZ31 specimens, showing that a physiological in vitro growth occurred in a 7 day interval.
Authors: X B Chen; D R Nisbet; R W Li; P N Smith; T B Abbott; M A Easton; D-H Zhang; N Birbilis Journal: Acta Biomater Date: 2013-11-27 Impact factor: 8.947