A Reduced-Order Computational Model of a Semi-Active Variable-Stiffness Foot Prosthesis.

Michael A McGeehan1, Peter G Adamczyk2, Kieran M Nichols2, Michael E Hahn1.   

Abstract

Passive energy storage and return (ESR) feet are current performance standard in lower limb prostheses. A recently developed semi-active variable-stiffness foot (VSF) prosthesis balances the simplicity of a passive ESR device with the adaptability of a powered design. The purpose of this study was to model and simulate the ESR properties of the VSF prosthesis. The ESR properties of the VSF were modeled as a lumped parameter overhung beam. The overhung length is variable, allowing the model to exhibit variable ESR stiffness. Foot-ground contact was modeled using sphere-to-plane contact models. Contact parameters were optimized to represent the geometry and dynamics of the VSF and its foam base. Static compression tests and gait were simulated. Simulation outcomes were compared to corresponding experimental data. Stiffness of the model matched that of the physical VSF (R2: 0.98, root-mean-squared error (RMSE): 1.37 N/mm). Model-predicted resultant ground reaction force (GRFR) matched well under optimized parameter conditions (R2: 0.98, RMSE: 5.3% body weight,) and unoptimized parameter conditions (R2: 0.90, mean RMSE: 13% body weight). Anterior-posterior center of pressure matched well with R2 > 0.94 and RMSE < 9.5% foot length in all conditions. The ESR properties of the VSF were accurately simulated under benchtop testing and dynamic gait conditions. These methods may be useful for predicting GRFR arising from gait with novel prostheses. Such data are useful to optimize prosthesis design parameters on a user-specific basis.
Copyright © 2021 by ASME.

Entities:  

Mesh:

Year:  2021        PMID: 33704375      PMCID: PMC8086177          DOI: 10.1115/1.4050456

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   1.899


  1 in total

1.  A simulation-based analysis of the effects of variable prosthesis stiffness on interface dynamics between the prosthetic socket and residual limb.

Authors:  Michael A McGeehan; Peter G Adamczyk; Kieran M Nichols; Michael E Hahn
Journal:  J Rehabil Assist Technol Eng       Date:  2022-07-15
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.