| Literature DB >> 33687513 |
Marlies van Dijk1, Jorine E Hartman2, Sonja W S Augustijn2, Nick H T Ten Hacken2, Karin Klooster2, Dirk-Jan Slebos2.
Abstract
PURPOSE: For this study, we aimed to compare dynamic hyperinflation measured by cardiopulmonary exercise testing (CPET), a six-minute walking test (6-MWT), and a manually paced tachypnea test (MPT) in patients with severe emphysema who were treated with endobronchial coils. Additionally, we investigated whether dynamic hyperinflation changed after treatment with endobronchial coils.Entities:
Keywords: COPD; Dynamic hyperinflation; Emphysema; Endobronchial coils
Year: 2021 PMID: 33687513 PMCID: PMC8053153 DOI: 10.1007/s00408-021-00430-0
Source DB: PubMed Journal: Lung ISSN: 0341-2040 Impact factor: 2.584
Baseline and follow-up outcomes for manually paced tachypnea test (MPT), cardiopulmonary exercise test (CPET), and six-minute walk test (6-MWT)
| Baseline | Follow up | p value | |
|---|---|---|---|
| CPET | |||
| Dynamic hyperinflation (ml) | − 740 (− 1530 to − 170) | − 750 (− 1300 to − 320) | 0.93 |
| Breathing frequency (x/min) at maximum workload | 28 (16–55) | 26 (15–38) | 0.017 |
| Tidal volumes (ml) at maximum workload | 840 (480–1310) | 1010 (630–1780) | 0.001 |
| Maximum workload (Watt) | 27 (3–51) | 33 (16–61) | 0.001 |
| VO2 peak (ml/min) | 719 (508–1090) | 770 (546–1109) | 0.13 |
| MPT | |||
| Dynamic hyperinflation (ml) | − 900 (− 1470 to − 540) | − 990 (− 1690 to − 540) | 0.017 |
| Breathing frequency (x/min) | 40 (39–42) | 40 (39–41) | 0.53 |
| Tidal volumes (ml) | 589 (390–895) | 680 (460–975) | < 0.001 |
| 6-MWT | |||
| Dynamic hyperinflation (ml) | − 335 (− 690 to + 60) | − 480 (− 930 to − 110) | 0.11 |
| Distance | 321 (172–469) | 362 (160–469) | 0.004 |
VO2 = oxygen uptake
Fig. 1A Bland–Altman plot of dynamic hyperinflation measured by cardiopulmonary exercise testing (CPET) and a manually paced tachypnea test (MPT)